6. SET DOSE / 0 ugkm displays. Enter a dose rate.<br />

- The exceedable Soft Limits for Dopamine are 3-50 ugkm. If 51<br />

ugkm were entered DOSE HIGH/DOSE LIMIT/50<br />

ugkm/OVER RIDE/DOSE LIMIT/50 ugkm/YES? NO?<br />

displays. Push YES to enter 51 ugkm. The pump will infuse<br />

with an override message displayed (MAX DOSE/OVER<br />

RIDE).<br />

- The unexceedable Hard Limits for Dopamine are 1-75 ugkm.<br />

If 76 ugkm were entered, it would not be accepted <strong>and</strong> DOSE<br />

HIGH/DOSE LIMIT/50 ugkm/SET DOSE/0 ugkm displays.<br />

7. Enter all other values the display asks for <strong>and</strong> push RUN to start the<br />

infusion.<br />

8. To titrate a dose rate without stopping the pump, push ENTER, select<br />

the new dose rate <strong>and</strong> push ENTER again.<br />

Drug Label Mode<br />

1. Drug labeling works in all modes of the pump.<br />

2. The Drug Label Mode applies to drugs <strong>and</strong> fluids where no rate limits<br />

have been set up in the pump’s Drug Library.<br />

3. To label a drug or fluid, Push YES when DRUG LABL/YES? NO?<br />

Displays <strong>and</strong> select the drug using the keypad. The drug name <strong>and</strong><br />

LABL ONLY displays.<br />

4. Use OPTION 99 to erase an errant drug label entry.<br />

9 Rev.J