Options 20<br />

21 Rev. J<br />

OPT.<br />

NO<br />

20<br />

OPTION<br />

NAME<br />

PIGGYBACK<br />

- TIME (MIN.)<br />

AND VOL.<br />

LIMIT<br />

INSTRUCTIONS DISPLAY<br />

Stop the pump. IN STOP/PUSH RUN<br />

Push OPTIONS. SET/OPT CODE<br />

Select 20. OPT 20/PIG MODE/PUSH ENTR<br />

Push ENTER. SET/PIG TIME/0 min<br />

Select the piggyback TIME in minutes (from 1-1440<br />

min). For example, select 30 min.<br />

PIG TIME/30 min/PUSH ENTR<br />

Push ENTER. SET/PIG VOL/VOL 0<br />

Select the piggyback VOLUME LIMIT (from 0.1-<br />

9999 ml). For example, select 50 ml.<br />

VOL 50/PIG VOL//PUSH ENTR<br />

Push ENTER CHEK/PIG CLMP/PUSH ENTR<br />

Ensure that the slide clamp on the piggyback line is<br />

open, press ENTER<br />

OPT 20/PIG MODE/IN STOP/PUSH RUN<br />

Open all clamps <strong>and</strong> push RUN to start the infusion. PIG/RATE XXX/PIG TIME/XXXXMINS alternately<br />

display<br />

Or to clear all PIG MODE values push CLEAR when<br />

PUSH RUN or OPT 20 displays.<br />

Note: The piggyback instructions under OPT. 21 below apply to OPT. 20 also.