Servicing<br />

1. Caution: Electric shock hazard. There are no user servicable parts.<br />

Do not remove the case. Refer servicing to qualified service<br />

personnel.<br />

2. Refer to the Service Manual for pump inspection <strong>and</strong> preventative<br />

maintenance procedures.<br />

3. When pump return to <strong>SIGMA</strong> is desired:<br />

a. Phone 1-800-356-3454 for repair authorization (RA)<br />

number. A P.O. number for non-warranty repairs is<br />

required.<br />

b. Ship pumps to <strong>SIGMA</strong> International at:<br />

711 Park Avenue, Medina, NY 14103<br />

(Return pumps in original boxes, with original inserts to<br />

prevent damage during shipment).<br />

c. Include a problem description, a contact person, a phone<br />

number, a return address, <strong>and</strong> label the shipping box with<br />

the RA number.<br />

4. Required Maintenance <strong>and</strong> Frequency:<br />

a. Maintenance consists of routine cleaning <strong>and</strong> operations<br />

evaluation in accordance with the procedures described in<br />

the service manual. It is recommended this evaluation be<br />

performed at least once every twelve (12) months.<br />

5. Battery Use <strong>and</strong> Maintenance:<br />

a. If a LOW BATT display is seen upon start, the battery is<br />

insufficiently charged. To recharge the battery, plug the<br />

power cord into an outlet for 18 - 24 hours. Recharging is<br />

automatic <strong>and</strong> occurs whenever the pump is plugged in,<br />

whether stopped or operating.<br />

b. If left idle for long periods of time (3 months or more) the<br />

pump should be plugged into an AC outlet to recharge the<br />

battery.<br />

c. Batteries are maintenance free <strong>and</strong> should last 200 full<br />

discharge/charge cycles.<br />

6. Protective Grounding:<br />

a. Do not operate the pump if the integrity of the protective<br />

earth conductor is in doubt. Have the pump serviced.<br />

43 Rev. J