Options 90, 95 &99<br />

37 Rev. J<br />

OPT.<br />

NO.<br />

90<br />

95<br />

99<br />

OPTION NAME INSTRUCTIONS DISPLAY<br />

DELAYED START<br />

(i.e.: “DLAY STRT)<br />

(Not available on all<br />

models)<br />

CLEAR ALL<br />

PROGRAM<br />

MODES<br />

CLEAR ALL 6 hr<br />

MEMORY<br />

If Primary Mode was not selected above, -Enter option number<br />

for the Option Mode desired to run following the delay.<br />

Go through mode set-up.<br />

If an infusion mode is set up prior to entering this option, that<br />

mode will become the delayed mode <strong>and</strong> you will not be asked<br />

during the Delayed Start setup to select a mode. -Push RUN.<br />

SET / OPT CODE-<br />

OPT 60 / AUTO PROG / PUSH ENTR<br />

(example, Option 60 entered)<br />

DLAY STRT / OPT 60 / AUTO PROG /STEP<br />

1 / PUSH RUN<br />

RATE KVO / DLAY STRT / 1.5 hrs / OPT 60 /<br />

If a KVO delay start selected (pump runs at KVO rate<br />

AUTO PROG<br />

If not………………………………………………... DLAY STRT /1.5 hrs / OPT 60 / AUTO PROG<br />

Note: Pushing STOP at any time during a delayed start or a<br />

KVO alarm delay pauses the timer. To continue the delay,<br />

push RUN.<br />

Stop the pump. IN STOP/PUSH RUN-<br />

Push OPTIONS. SET/OPT CODE-<br />

Select 95. OPT 95/CLR PRGS/PUSH ENTR-<br />

Push ENTER PRGS CLRD/SET RATE<br />

Note:<br />

Use of this option enables the operator to clear all pump memory, except the following items which must be<br />

cleared individually<br />

• Keylock<br />

• Alarm Volume<br />

• Occlusion Pressure Options<br />

• Maximum RATE or VOL. LIMIT Settings<br />

• Biomedical Mode Entries<br />

Programming Modes<br />

Stop the pump. IN STOP/PUSH RUN<br />

Push OPTIONS. SET/OPT/CODE<br />

Select 99. OPT 99/CLR ALL/PUSH ENTR<br />

Push ENTER ALL CLR/SET RATE