Operating Displays<br />

CHART<br />

SCALE<br />

AUTO<br />

TUNE<br />

BASEFIELD<br />

5—7<br />

Entering 0 (zero) results in a reading<br />

interval equal to the reading time.<br />

The instrument goes to “sleep”<br />

between readings, for 4 seconds and<br />

up.<br />

The sensitivity for the analog output<br />

for the strip chart recorder.<br />

The following choices are available:<br />

1, 10, 100, 1000, 10000 nT.<br />

A “yes/no” toggle that controls the<br />

automatic tuning function.<br />

It is independent for the Total-field<br />

sensor and the Gradiometer sensor.<br />

The tuning value is updated after<br />

each reading in preparation for the<br />

next one.<br />

Auto tuning is particularly applicable<br />

if large variations of the ambient field<br />

are encountered over longer<br />

distances.<br />

Fixed tuning may be more<br />

appropriate in areas of large cultural<br />

electrical noise or very narrow, large<br />

amplitude anomalies. These may<br />

pull the tuning away from the desired<br />

frequency to that of the interfering<br />

frequency or to tuning field values<br />

greatly different from the background.<br />

As the setting of this parameter<br />

depends upon your particular survey<br />

conditions, it is not possible to<br />

suggest the best setting. The mode<br />

should therefore be established by<br />

experimenting. In general it is more<br />

applicable to use fixed tuning for the<br />

site characterization and drum<br />

location modes.<br />

The base field is used in conjunction<br />

with the base-station correction<br />

procedure.<br />

The base-station correction<br />

technique removes variation in the<br />

ambient field during the time that the<br />

base station is running. However,<br />

variations taking place from day to<br />

day are not corrected. The base field<br />

1<br />

No<br />

0<br />

P/N 788715 Rev. 0