Airflow Smoke Pattern Test<br />

Airflow smoke patterns have been shown <strong>to</strong> vary depending on the equipment that is used inside the<br />

cabinet. The <strong>BIOPROtect</strong> <strong>II</strong> <strong>JR</strong> has been tested with the cart only. A description of this testing is<br />

summarized in the test report. However, when the <strong>BIOPROtect</strong> <strong>II</strong> <strong>JR</strong> is set-up properly (i.e. your set-up<br />

should not block or restrict the grilles and diffusers provided) the appearance of the airflow directly at the 8”<br />

access openings must be consistent with that described in the test report.<br />

To check for the direction of air movement, use a smoke genera<strong>to</strong>r and trace along the front access<br />

opening on the inside of the cabinet. Observe that no smoke is escaping from the work area.<br />

In order <strong>to</strong> be sure that room air is not entering the work area, trace along the outside of the front access<br />

opening. Observe that no smoke penetrates <strong>to</strong> the work surface.<br />

Grounding Continuity Test<br />

Using an Ohmmeter, set it <strong>to</strong> read the low resistance scale. Touch the two leads <strong>to</strong>gether and see that the<br />

display reads 0.1Ω <strong>to</strong> 0.0Ω. Touch one lead <strong>to</strong> the ground lug on the cabinet power cord while <strong>to</strong>uching the<br />

other lead <strong>to</strong> bare metal on the unit where the user would be likely <strong>to</strong> <strong>to</strong>uch the cabinet. According <strong>to</strong> NSF/<br />

ANCI-standard 49-2202 the maximum allowable shall be 0.15Ω (Max allowable for UL is 0.50Ω ).<br />

http://solid/prereleased/prereleased/STAN. PRODUCT/313/DOCS/313D000.doc<br />

Page 19 of 59