OPERATOR'S MANUAL BIOPROtect II JR Vent to ... - Baker Company
OPERATOR'S MANUAL BIOPROtect II JR Vent to ... - Baker Company
OPERATOR'S MANUAL BIOPROtect II JR Vent to ... - Baker Company
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Access <strong>to</strong> the Work Area<br />
Figure 1 Typical section with airflow visualization arrows<br />
The <strong>BIOPROtect</strong> <strong>II</strong> <strong>JR</strong> is provided with three modes of access <strong>to</strong> the work area; via the large hinged door<br />
that can be opened <strong>to</strong> access the entire work area, the hinged viewscreen (providing access <strong>to</strong> the work<br />
surface) or from the 8” x 42” access (sash) opening between the bot<strong>to</strong>m of the closed viewscreen and the air<br />
intake grate. The primary means of access must be through the 8” sash opening. This is the only means of<br />
access that will ensure the user protection from contaminants within the enclosure.<br />
The opera<strong>to</strong>r is only protected from airborne particulate when the large front door and the hinged viewscreen<br />
is closed and secured <strong>to</strong> provide an 8” x 42” access (sash) opening<br />
For easy set-up of equipment in the work area, the <strong>BIOPROtect</strong> <strong>II</strong> <strong>JR</strong> has one large door that exposes the<br />
entire work area from floor <strong>to</strong> ceiling <strong>to</strong> an area of 51.5” wide x 68” tall.<br />