4.0 GENERAL<br />

Note: Follow all safety procedures for handling, pouring, and disposing of<br />

chemicals.<br />

Note: This procedure is to be conducted only within the exhausting <strong>PFC</strong> <strong>Sink</strong> fume<br />

hood.<br />

Materials: Use only the clean quartz constant temperature bath identified as Bath 1 or<br />

Bath 2. Use dedicated materials for any piece that comes in direct contact with the acid<br />

solutions. Keep the wafer carrier handles with the wafer carrier in the chemical baths.<br />

5.0 PROCESS PREPARATION<br />

5.1 Select bath and wafer cassettes<br />

1. There are two chemical baths located on the center-right side of the <strong>PFC</strong> sink (see figure<br />

1).<br />

CAUTION: Do not use any of the colored cassettes e.g., black or<br />

orange. (These are not Teflon)<br />

2. For 4 inch, use a full 4 inch cassette and they are located in the storage cart under the<br />

fumehood. (See Figure 3)<br />

Figure 3<br />

3. Single wafer Teflon carriers are available for single 2.5" or quartered wafers, or for<br />

irregularly shaped substrates.