d. Remove the top cover of the chemical bath 1. Refresh the solution with 250 ml<br />
hydrogen peroxide (H2O2), unless the solution was made immediately prior to<br />
usage. Make sure the bath temperature is less than 30C.<br />
CAUTION: Do not overfill the bath, use the aspirator (figure 6) to remove some<br />
chemical prior to adding the hydrogen peroxide.<br />
Figure 6<br />
e. Verify it is at 90 o C prior to usage then immerse the cassette in Bath 1. (see figure<br />
7 below). Press Timer on the Bath 1 controller to activate the 10 minute timer.