°C, and <strong>the</strong>n ground to powder. After mixing with Niobium powder, <strong>the</strong> AgI samples were pressed into a<br />
copper holder and <strong>the</strong> measurement <strong>of</strong> 129 I/ 127 I ratios was carried out using <strong>the</strong> AMS system at <strong>the</strong> Tandem<br />
Laboratory, Uppsala University (5MV Pelletron, National Electrostatic Corporation, USA) at a terminal<br />
voltage <strong>of</strong> 3.5 MV. The 129 I NIST-SRM 4949C standard with 129 I/ 127 I ratio <strong>of</strong> 1.1×10 -11 was used (Alfimov<br />
2005). Blank samples were prepared using <strong>the</strong> same procedure as <strong>for</strong> samples <strong>for</strong> total iodine, iodide and<br />
iodate. The measured 129 I/ 127 I ratio in blank samples (1.5±0.5×10 -13 ) was two orders <strong>of</strong> magnitude lower than<br />
that in samples (0.5-15×10 -11 ), and was subtracted from <strong>the</strong> measured value in <strong>the</strong> samples. Instrumental<br />
background <strong>for</strong> <strong>the</strong> AMS system was controlled by measurement <strong>of</strong> natural AgI (iodargyrite), which gave a<br />
129 I/ 127 I ratio <strong>of</strong> 4×10 -14 .<br />
2.11 127 I preparation <strong>for</strong> ICP-MS measurements. For 127 I determination, around 2 mL <strong>of</strong> <strong>the</strong> solutions<br />
from various fractions was diluted to 20 mL with 1% NH3, and Cs (as CsCl) was added as internal standard.<br />
The concentration <strong>of</strong> 127 I was determined using an X Series II ICP MS (Thermo Electron Corporation).<br />