<strong>Glacier</strong> <strong>Transportable</strong> <strong>Sampler</strong><br />

Section 1 Introduction<br />

1.3 Technical<br />

Specifications<br />

Table 1-1 lists the technical specifications for the <strong>Glacier</strong> <strong>Transportable</strong><br />

<strong>Sampler</strong>.<br />

Table 1-1 <strong>Glacier</strong> <strong>Transportable</strong> <strong>Sampler</strong> Specifications<br />

Size (H W D) 25 15 24 in. 63 38 60 cm<br />

(Does not include optional mobility kit)<br />

Weight 60 lb. 28 kg<br />

(Includes refrigerator, controller, empty bottle, and bottle retainer. Does not<br />

include battery or optional mobility kit)<br />

Power<br />

The mains line cord is the<br />

disconnect device.<br />

Installation Category<br />

Pollution Degree 4<br />

Maximum Altitude 1<br />

2000 meters<br />

Pump<br />

Intake Purge<br />

Tubing Life Indicator<br />

Intake Suction Line:<br />

Length<br />

Material<br />

Inside Dimension<br />

Pump Tubing Life<br />

Maximum Suction Lift<br />

Typical Repeatability<br />

Typical Line Velocity:<br />

3 ft (1 m)<br />

10 ft (3.1 m)<br />

15 ft (4.6 m)<br />

Liquid Presence Detector<br />

Controller<br />

DC Power Configuration: 12 VDC, 6 amperes<br />

Dual-power Configuration: 12 VDC, 6 amperes, or<br />

87 to 264 VAC, 47 to 63 Hz, 2 amperes<br />

II<br />

Before and after each sample collection<br />

Provides message to change tubing<br />

3 to 99 feet 1 to 30 m<br />

Vinyl or PTFE line<br />

0.375 in. or 0.250 in. 0,952 or 0,635 cm<br />

2,000 samples typical<br />

26 feet<br />

± 10 ml<br />

2.9 ft/sec 0,88 m/sec<br />

2.5 ft/sec 0,76 m/sec<br />

1.9 ft/sec 0,58 m/sec<br />

Non-wetted, non-conductive sensor for automatic detection of liquid<br />

Weight 8 pounds 3.6 kg<br />

Dimensions 10 x 12.5 x 10 inches 26 x 32 x 25 cm<br />

Operational Temperature 32° to 120° F 0° to 49° C<br />

Enclosure Rating NEMA 4x, 6 IP 67<br />

Program Memory<br />

Flow Meter Signal Requirements<br />

Real Time Clock Accuracy<br />

Refrigerator<br />

Non-volatile ROM<br />

5 to 15 volt DC pulse or 25 millisecond or longer isolated contact closure<br />

1 minute per month typical<br />

Sample maintenance temperature 37.4 ± 1.8° F 3.0 ± 1.0° C<br />

Software<br />

Sample Frequency Selection<br />

1 minute to 9,999 minutes in 1 minute increments 1 to 9,999 flow pulses<br />

1-5