<strong>Glacier</strong> <strong>Transportable</strong> <strong>Sampler</strong><br />

Section 2 Installation<br />

TAKING 200 ml<br />

CALIBRATE SAMPLE<br />

The <strong>Glacier</strong> displays the screen at left and goes through a complete<br />

sample collection cycle. The <strong>Glacier</strong> deposits the sample in<br />

the graduated cylinder.<br />

Note<br />

The volume delivered during calibration is the programmed<br />

Sample Volume.<br />

2.10.6 Calibration Step 6 The <strong>Glacier</strong> displays the amount of liquid it has deposited into<br />

the graduated cylinder.<br />

VOLUME DELIVERED<br />

200 ml<br />

Measure the volume and compare the numbers. If they match,<br />

press ENTER and the <strong>Glacier</strong> will return to the Standby state.<br />

Note<br />

Place the graduated cylinder on a level surface. Read the liquid<br />

volume from the bottom of the meniscus, not the top of the<br />

curved edges where the liquid wets the sides of the cylinder.<br />

If the amounts differ, enter the actual volume delivered. To enter<br />

the actual volume:<br />

1. Press the appropriate NUMBER buttons on the keypad.<br />

Note<br />

If you enter an incorrect value with the number buttons, press<br />

the STOP button. The <strong>Glacier</strong> restores the original value and<br />

waits for a new value.<br />

___ ml! ARE YOU<br />

SURE? YES NO<br />

2. Press the Enter button to accept the value. The <strong>Glacier</strong><br />

updates its internal pump tables. If there is a significant<br />

difference between the old and new values, the <strong>Glacier</strong> displays<br />

the screen at left.<br />

3. Compare this value with the measured volume.<br />

If they match, select YES using the ARROW button and<br />

then press ENTER.<br />

If they do not match, select NO using the ARROW button<br />

and then press ENTER. The <strong>Glacier</strong> returns to the Volume<br />

Delivered screen (step 1, above).<br />

When through with Calibration Step 6, the <strong>Glacier</strong> logs the calibration<br />

time in the event log and returns to the Standby state. Be<br />

sure to reconnect the pump tube end to the bulkhead fitting.<br />

2.11 Start the Program The <strong>Glacier</strong> has two programs—the current program and the<br />

stored program. To run the current program press the GO<br />

button.<br />

1 2 3<br />

You can also quickly recall the stored program and start it. To do<br />

so, press the PROGRAM button twice, followed by the GO button.<br />

This action will load the stored program settings as the current<br />

settings and run the program.<br />