07.03.2014 Views

Glacier Transportable Sampler - Isco

Glacier Transportable Sampler - Isco

Glacier Transportable Sampler - Isco

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Glacier</strong> <strong>Transportable</strong> <strong>Sampler</strong><br />

Section 2 Installation<br />

To set the suction line size:<br />

1. Determine the size of suction line in use. Compare this to<br />

the blinking selection.<br />

2. If the selection is incorrect, press the ARROW button. This<br />

will change the blinking selection.<br />

3. Press the ENTER button to accept the selection. The <strong>Glacier</strong><br />

loads the size into the current program settings and<br />

advances to the next step.<br />

Note<br />

Selecting 1 /4-inch (6-mm) suction line disables the Bottle Full<br />

detection.<br />

2.10.2 Calibration Step 2 The <strong>Glacier</strong> advances to the Suction Line Length display.<br />

SUCTION LINE LENGTH:<br />

7.6 m (0.9-30.2 m)<br />

The Suction Line Length is the measured length of tubing in use.<br />

The length is measured from end-to-end, without the strainer or<br />

tube coupling. When using English units of measure, the suction<br />

line tubing must be cut in whole foot lengths. When using metric<br />

units of measure, the tubing must be cut in decimeter lengths.<br />

To enter the Suction Line Length:<br />

1. Measure the length of tubing. Cut the tubing if necessary,<br />

to the nearest whole foot or decimeter.<br />

2. Enter the tubing length. Press the appropriate NUMBER<br />

buttons on the keypad.<br />

Note<br />

If you enter an incorrect value with the NUMBER buttons,<br />

press the STOP button. The <strong>Glacier</strong> restores the original value<br />

and waits for a new value.<br />

3. Press the ENTER button to accept the value. The <strong>Glacier</strong><br />

loads the length into the current program settings and<br />

advances to the next step.<br />

2.10.3 Calibration Step 3 The <strong>Glacier</strong> advances to the Suction Head display.<br />

Note<br />

This step only appears when the liquid detector is disabled<br />

(see section 3.8.4).<br />

SUCTION HEAD:<br />

3.1 m (0-7.6 m)<br />

Suction Head is defined as the vertical distance from the surface<br />

of the liquid to the inlet of the pump.<br />

Since the <strong>Glacier</strong> will be unable to calculate the suction head as<br />

it collects each sample, the <strong>Glacier</strong> will ask for a manual or<br />

“fixed” suction head. The <strong>Glacier</strong> skips this step when the Liquid<br />

Detector is enabled.<br />

2-16

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!