30.10.2012 Views

EDAX EDS Manual

EDAX EDS Manual

EDAX EDS Manual

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>EDAX</strong> Phoenix Training Course –Introductory Diagrams - page 1


ENERGY DISPERSIVE X-RAY SPECTROMETRY--<br />

<strong>EDS</strong> INSTRUMENTATION & SIGNAL DETECTION<br />

1. X-Ray Detectors:<br />

ALAN SANDBORG<br />

<strong>EDAX</strong> INTERNATIONAL, INC.<br />

The <strong>EDS</strong> detector is a solid state device designed to detect x-rays and convert their<br />

energy into electrical charge. This charge becomes the signal which when processed<br />

then identifies the x-ray energy, and hence its elemental source.<br />

The X-ray in its interaction with solids, gives up its energy and produces electrical<br />

charge carriers in the solid. A solid state detector can collect this charge. One of the<br />

desirable properties of a semiconductor is that it can collect both the positive and<br />

negative charges produced in the detector. The figure below shows the detection<br />

process.<br />

Figure 1.<br />

There are two types of semiconductor material used in electron microscopy. They are<br />

silicon (Si) and germanium (Ge). In Si, it takes 3.8 eV of x-ray energy to produce a<br />

charge pair, and in Ge it takes only 2.96 eV of energy. The other properties of these two<br />

types will be discussed later in this section. The predominant type of detector used is<br />

the Si detector, so it will be favored in the discussions. With a Si detector, an O K x-ray<br />

whose energy is 525eV will produce 525/3.8= 138 charge pairs. A Fe K x-ray will<br />

produce 6400/3.8= 1684 charge pairs. So by collecting and measuring the charge, the<br />

<strong>EDAX</strong> Phoenix Training Course - <strong>EDS</strong> Instrumentation & Signal Detection - page 1


Important <strong>EDS</strong> Parameters<br />

Count Rate<br />

For a good quality spectrum (i.e. good resolution and fewest artifacts) you should use the<br />

50 or 100 us time constant (pulse processing time) with a deadtime of 20 to 40%, and 500<br />

to 2500 cps.These are good numbers if the sample consists largely of high energy peaks (><br />

1 keV), but if the spectrum is dominated by low energy peaks (< 1 keV) then a count rate of<br />

500 - 1000 cps is better and the 100 us time constant should be used.<br />

When maximum count throughput is required, such as when collecting fast x-ray maps, a<br />

faster time constant (2.5 to10 us) should be used with a count rate of 10,000 to 100,000<br />

cps. The deadtime should not exceed 50 to 67%. These conditions may not be optimum for<br />

low energy peaks in terms of their resolution and/or position. A lower count rate and slower<br />

time constant should be and/or it might be necessary to adjust the position of the ROI prior<br />

to collecting the map.<br />

Accelerating Voltage<br />

The overvoltage is a ratio of accelerating voltage used to the critical excitation energy of a<br />

given line for an element. Typically, the overvoltage should be at least 2 for the highest<br />

energy line and no more than 10 to 20 times the lowest energy line of interest. We use the<br />

number 10 for quantitative applications and the 20 for qualitative applications.<br />

For example, if you are interested in analysis of a phase containing Fe, Mg and Si and want<br />

to use the K lines for each, then 15 kV will probably work reasonably well. If, however, you<br />

need to analyze the same three elements plus oxygen as well, then you might use 5 to 10<br />

kV, but you might want to use the L line for the Fe.<br />

Why should the overvoltage be at least 2 for the highest energy element? Because at lower<br />

overvoltages the fraction of the interaction volume where the element can be excited<br />

becomes very small and you will not be able to generate very many x rays of that energy.<br />

Why should the overvoltage be less than 10 to 20 times the lowest energy peak? When the<br />

overvoltage number is excessive, the proportion of the interaction volume for which the low<br />

energy x rays can escape without being absorbed also becomes small. The result is a<br />

small peak and in the case of the quantification there will be a strong absorption correction<br />

which will magnify the statistical errors in our analysis.<br />

Take-Off Angle<br />

Typical take-off angles will range from 25 to 40 degrees. This angle is a combination of the<br />

detector angle, its position, sample working distance and sample tilt. The sensitivity for very<br />

low energy x rays and/or signals characterized by high absorption can be enhanced by<br />

increasing the take-off angle. Some inclined detectors (e.g. a detector angle of<br />

approximately 35 degrees above the horizontal) do not require sample tilt. Horizontal entry<br />

detectors require that the sample be tilted to achieve an optimum take-off angle.<br />

<strong>EDAX</strong> Phoenix Training Course - <strong>EDS</strong> Parameters - page 1


<strong>EDAX</strong> Detector Geometry<br />

The elevation angle (EA) is the angle<br />

between the horizontal and the detector<br />

normal. The intersection distance (ID) is the<br />

distance in mm between the pole piece<br />

where the electron beam intersects the<br />

detector normal. The azimuth angle can not<br />

be shown in a cross-sectional view as<br />

shown at the left, but it is the angle as<br />

viewed from above between the detector<br />

normal and normal to the tilt axis. The<br />

working distance (WD) is where the sample<br />

is in mm below the pole piece. The take-off<br />

angle (TOA) is the angle between the x-ray trajectory and the sample surface. If the sample is<br />

placed at the intersection distance and not tilted, the take-off angle will equal the elevation angle.<br />

If the working distance is shorter than<br />

the intersection distance, the take-off<br />

angle will be less than the elevation<br />

angle. This assumes that the sample is<br />

smooth and not tilted.<br />

If the working distance is longer than<br />

the intersection distance, the take-off<br />

angle will be more than the elevation<br />

angle. Again, this assumes that the<br />

sample is smooth and not tilted.<br />

If the sample is tilted toward the detector,<br />

the take-off angle will be greater<br />

than the elevation angle. If the sample<br />

were tilted away from the detector, the<br />

take-off angle would be less than the<br />

elevation angle. Note that the azimuth<br />

angle must be taken into account to<br />

determine the take-off angle.<br />

<strong>EDAX</strong> Phoenix Training Course - <strong>EDAX</strong> Detector Geometry - page 1


Dead Time & Time Constants<br />

In an <strong>EDS</strong> system the real time (or clock time) is divided into live time and dead time. The<br />

live time is the time when the detector is alive and able to receive an x-ray event (i.e. the<br />

time when it is doing nothing) and the dead time is when the detector or preamplifier is<br />

unable to accept a pulse because it is busy processing or rejecting an event(s). Basically,<br />

the charges from an x-ray photon can be collected in about 50 ns and in the end we will take<br />

roughly 50 us to process the filtered, amplified pulse (1000 times longer). Quite often, x-ray<br />

photons will come in too close to each other and we will reject the signals. That is why we<br />

can see the situation that we actually collect very few x-ray events at very high count rates<br />

and we actually process more counts when we use a lesser count rate.<br />

If we decide to take less time to process the pulse in the end (say, less than a 10 us time<br />

constant or pulse processing time) then we can process more counts. However, because<br />

we have taken less time, there is the possibility that we do not process the peak energy as<br />

accurately and the peaks will be broader; the resolution will be poorer or higher.<br />

The Phoenix system has 8 time constants or pulse processing times which will allow for<br />

optimum resolution or x-ray count throughput. Under most circumstances, we would like to<br />

have a dead time that is between 10 and 40% and perhaps between 20 and 30% if we would<br />

like to tighten the range of our analytical conditions. There are times when we are most<br />

concerned about the count throughput and are not concerned about resolution, sum peaks,<br />

etc., such as when we are collecting x-ray maps. Under these conditions a dead time as<br />

high as 50 to 60% is feasible. Under no circumstances should a dead time that is more than<br />

67% be used because we will actually get fewer counts processed.<br />

A series of plots are included on the following pages. These show: the fall-off in processed<br />

counts when the count rate is increased too high, a comparison in throughput between a fast<br />

time constant and a slower time constant, and a plot showing the dead time % for various<br />

count rates and time constants, and a plot of count rate versus dead time % for each of the 8<br />

time constants.<br />

<strong>EDAX</strong> Phoenix Training Course –Dead Time & Time Constants - page 1


EDX SPECTRUM INTERPRETATION AND ARTIFACTS<br />

Continuum X Rays<br />

As a result of inelastic scattering of the primary beam electrons in which the electrons<br />

are decelerated and lose energy without producing an ionization of the atoms in the<br />

sample, continuum x rays are formed. The continuum x rays are the background of our<br />

<strong>EDS</strong> spectrum and are sometimes referred to as the bremsstrahlung. In theory, the<br />

continuum can be expected to extend from the maximum energy of the primary beam<br />

electrons and increase exponentially to zero keV energy. In reality, the background<br />

goes to zero at the low end of the energy spectrum due to absorption by the detector<br />

window, the detector dead layer, and the gold layer. The intensity of the continuum is<br />

related to both the atomic number of the sample as well as the beam energy. The<br />

continuum intensity also increases with beam current.<br />

Characteristic X Rays<br />

Inelastic scattering events between the primary beam electrons and inner shell electrons<br />

which result in the ejection of the electron from the atom within the sample and may lead<br />

to the formation of a characteristic x ray. The ejection of the electron leaves the atom in<br />

an ionized, excited state and permits an outer shell electron to move to the inner shell.<br />

Because the energy levels of the various shells are related to the number of charges in<br />

the nucleus, the energy of the emitted x ray is “characteristic” of the element. The beam<br />

electron must have an energy greater that is just slightly greater than the energy of the<br />

shell electron (the critical ionization energy).<br />

Depth of Excitation<br />

Although electrons may penetrate to specific depths within a sample which can be<br />

illustrated with a variety of equations or with Monte Carlo programs, the electrons<br />

actually lose energy in steps as they go to greater depths in the sample. As a result, an<br />

electron may soon lose a sufficient amount of its energy such that it can no longer excite<br />

characteristic x rays. Typically, this occurs when its energy drops below the critical<br />

ionization energy of the elements in the sample. Each element within the sample will<br />

have its own critical ionization energy and its own excitation depth. The ratio of the<br />

primary beam energy to the excitation energy of the element is referred to as the<br />

<strong>EDAX</strong> Phoenix Training Course - Spectrum Interpretation & Artifacts - page 1


<strong>EDAX</strong> Phoenix Peak Identification<br />

The basic procedure for identifying peaks in the <strong>EDS</strong> spectrum is not always agreed upon.<br />

Some users start with the biggest peak, identify it and all associated peaks, and then<br />

progress to the smaller peaks in sequence until all are identified. When there are L- and Mseries<br />

peaks present, it may actually be better to identify the highest energy alpha peak first<br />

because this peak may be associated with several other higher energy peaks as well as one<br />

or more of the lower energy peaks. Then, the user can find the next lowest energy peak,<br />

etc., until all peaks have been accounted for. Another significant aid for peak ID is to click<br />

on the button for the automatic peak ID. The results of the automatic peak ID should be<br />

inspected and verified; it very often will provide the correct answer, but the user should be<br />

aware that its answer might be only a part of the solution and that more investigation may be<br />

required. In all cases, however, it is essential to look for escape peaks for the peaks of<br />

greatest height and for sum peaks for these same peaks when a high count rate has been<br />

used for spectrum acquisition.<br />

Typically, we examine that part of the spectrum that ranges in energy between 0 and 10 keV<br />

and we quickly become aware of characteristic features of K-, L- and M-shell x-ray peaks<br />

that aid us in our peak identification. In this energy range, we can see peaks with K-shell xrays<br />

that correspond to atomic numbers 4 through 32, L-shell peaks that range from roughly<br />

atomic numbers 22 to 79, and M-shell x rays ranging from 56 to the highest known and<br />

observable atomic numbers. For many elements it is possible to observe peaks from more<br />

than one shell in this energy range (it is often desirable to inspect the region between 10 and<br />

20 keV as well for confirmation of lower energy peaks). The resolution of the peak of one<br />

element from an adjacent atomic number is often better resolved by using higher energy<br />

peaks (see spectrum of Ni and Cu below).<br />

<strong>EDAX</strong> Phoenix Training Course - Peak ID - page 1


<strong>EDAX</strong> Peak ID Quiz –“sp_pkid” Spectra<br />

On the floppy disk provided with this notebook is a directory “sp_pkid”. This directory<br />

contains a series of spectra “pkidXX.spc” and “deconXX.spc”. The spectra pkid01 through<br />

pkid04 and pkid06 are the same spectra used for illustrations in the “Peak Identification”<br />

discussion. Other spectra on the disk can be used as a peak identification quiz and they<br />

range in difficulty from relatively easy to impossible. In the space below, identify the peak by<br />

element and shell (e.g. SiK, PbM, SnL, etc.).<br />

Quiz:<br />

PkID05.spc (easy). Between 1.5 and 12 keV, the peaks are:<br />

PkID05b.spc (easy). Between 3.7 and 7.5 keV, the peaks are:<br />

PkID07.spc (moderately hard). Between about 0.9 and 18 keV, the peaks are:<br />

PkID08.spc (moderate). Between 0.6 and 14 keV, the peaks are:<br />

PkID09.spc (hard). Between 0.55 and 15 keV, the peaks are:<br />

PkID10.spc (easy to moderate). Between 0.2 and 10 keV, the peaks are:<br />

Peak ID Quiz -- page 1


<strong>EDAX</strong> Peak ID Quiz –“sp_pkid” Spectra<br />

On the floppy disk provided with this notebook is a directory “sp_pkid”. This directory<br />

contains a series of spectra “pkidXX.spc” and “deconXX.spc”. The spectra pkid01 through<br />

pkid04 and pkid06 are the same spectra used for illustrations in the “Peak Identification”<br />

discussion. Other spectra on the disk can be used as a peak identification quiz and they<br />

range in difficulty from relatively easy to impossible.<br />

Quiz Answers:<br />

PkID05.spc. This spectrum shows a series of K-series peaks with atomic numbers 14, 16,<br />

18, 20, 22, 24, 26, 28, 30, and 32 (Si, S, Ar, Ca, Ti, Cr, Fe, Ni, Zn, Ge). There is good<br />

separation of the Ka peaks and the Kb peaks are easily seen. The L-series peaks are not so<br />

well resolved.<br />

PkID05b.spc. A similar spectrum to PkID05 except that this shows a K-series peak for every<br />

atomic number between 21 and 27 (Sc, Ti, V, Cr, Mn, Fe, and Co). In this area of the<br />

spectrum the Kb peak of a given atomic number is overlapped by the Ka of the next higher<br />

atomic number. Note the non-resolution of the L-series peaks for the same atomic numbers.<br />

PkID07.spc. This spectrum is an older, non-Sapphire spectrum; that is why you will get the<br />

dialog box asking if you want to “ignore”, “cancel” etc. From 1 keV to 18 keV, the peaks<br />

present are:<br />

ZnL, (NaK?), MgK, AlK + BrL, Si, NbL (probable), MoL, ClK, K K, TiK, V K, CrK, MnK,<br />

FeK, CoK, NiK, CuK, ZnK, BrK, MoK.<br />

PkID08.spc. Between 0.6 and 14 keV, the peaks are:<br />

NiL, SeL, ZrL, PdL, TeL, TiK, NiK, SeK.<br />

PkID09.spc. Between 0.55 and 15 keV, the peaks are:<br />

F K, TaM, ThM, TaL, ThL. (This spectrum might also have some Uranium and other<br />

radioactive elements).<br />

PkID10.spc. Between 0.2 and 10 keV, the peaks are:<br />

C K, CrL, NiL AlK, SiK, MoL, TiK, CrK, FeK, CoK, NiK. (There is probably also some<br />

CoL in the sample but it is not a resolved peak).<br />

PkID11.spc. Between 1 and 10 keV, the peaks are:<br />

P K, InL.<br />

PkID12.spc. Between 1 and 10 keV, the peaks are:<br />

S K, CdL.<br />

Decon01.spc. Between 1 and 10 keV, the peaks are (or might be):<br />

IrM, P K, ZrL, InL, IrL.<br />

Decon01b.spc. Between 1 and 10 keV, the peaks are (or might be):<br />

Y L (?), IrM, P K (?), ZrL (?), InL, SnL, IrL.<br />

<strong>EDAX</strong> Quiz on Peak ID - page 1


THE OPTIMIZATION OF X-RAY COUNT THROUGHPUT<br />

The x-ray count rate is certainly one our most fundamental parameters for determining<br />

optimal spectrum conditions. It is counter-intuitive for most users that increasing the<br />

count rate does not automatically result in a better spectrum or at a minimum in<br />

collecting more counts. The <strong>EDAX</strong> detector amplifier has two time constants (20 µs and<br />

40 µs) which control the amount of time taken to process an x-ray event. While the x<br />

ray is being processed, the detector is “dead” and will not process any additional events.<br />

For each time constant there is a maximum throughput. When it is desirable to have<br />

more counts in a spectrum or in a map, it is always possible to collect the signal for a<br />

longer interval of time, but it is also possible to use a shorter time constant. When the<br />

shorter time constant is employed, the spectrum resolution degrades but for many<br />

applications this is not a serious drawback.<br />

The figure below provides an example of how many counts can be processed in a fixed<br />

interval of time (20 clock seconds). With the longer time constant (“40 tc” or 40 µs),<br />

there are actually fewer counts processed when the count rate was increased from 2000<br />

to 10,000 cps. The same 10,000 cps count rate yielded 6 to 7 times more processed<br />

counts when the shorter time constant (“20 tc” or 20 µs). With the shorter time constant,<br />

doubling the count rate only increased the processed counts by about ¼.<br />

The resolution of the detector degrades noticeably when using the longer time constant<br />

(see figure on the next page), but resolution is not normally a fundamental concern<br />

when we do certain procedures such as collecting x-ray maps, or when the spectra<br />

collected do not have significant overlaps and we are not concerned with low-energy<br />

peaks or their analysis.<br />

<strong>EDAX</strong> Training Course - X-Ray Count Optimization - page 1


SEM QUANT ZAF<br />

Geometry<br />

� The EDX detector parameters are specific for each microscope. The<br />

elevation angle, intersection distance, azimuth angle, and scale<br />

setting must be known and understood for microscope/detector<br />

geometry. Before collecting a spectrum, an accurate accelerating<br />

voltage should be verified and the working distance and take-off<br />

angle must be accurate. Every microscope has an optimized geometry<br />

or an optimized range of conditions. If spectra are collected that<br />

are not correctly specified, then the subsequent quantification will<br />

not be accurate and comparisons to spectra collected under different<br />

conditions will be difficult at best. Also, spectra collected from<br />

a non-optimal geometry are subject to more stray radiation artifacts<br />

and possibly to a reduced count rate.<br />

� Every detector comes with a blue Detector Specification <strong>Manual</strong>.<br />

This manual contains blueprints for the detector and gives all the<br />

pertinent parameters (working distance, intersection distance, etc.)<br />

that need to be entered into the Geometry dialog in the EDAM panel.<br />

� Enter the kV reading from the microscope in the EDAM panel (button<br />

located in the upper right corner of the screen).<br />

� Select a time constant (Amp Time) in the EDAM panel. Longer time<br />

constants (100 usec, 50 usec, and 35 usec) are typical settings for<br />

spectrum collection.<br />

Setup Preset<br />

Predefined collection times can be created in the Setup: Preset<br />

dialog box found in the Setup pull-down menu.<br />

Spectrum Collect, Erase, Save<br />

� To start collecting a spectrum click on the stop watch button (far<br />

left).<br />

� To stop collection do the same (click on the stop watch button (far<br />

left).<br />

� To clear a spectrum click on the paint-roller button next to the<br />

stop watch.<br />

� To save the spectrum click on “File: Save As...”.<br />

� Follow the usual procedures for saving any file in MS Windows.<br />

The spectrum files will normally (default) be saved to the<br />

D:\edax32\eds\usr sub-directory. Another path can be chosen<br />

here.<br />

� The “Save As” dialog box will select the first set of characters<br />

from the spectrum label for the file name prefix. There is a<br />

128-character limitation to the filename as defined by Microsoft<br />

Windows ’98 and NT.<br />

� The default file type his as a “.SPC” file suffix, but the file<br />

type can also be changed to save the spectrum as a graphic file<br />

(.BMP or .TIF) to be inserted into MS Word or other report<br />

writing software.<br />

Labels and Add Text<br />

� Up to 216 characters can be saved as a spectrum label. The sample<br />

label is displayed at the top of the spectral window in the “A:”<br />

labeled box or, if it is an overlaid spectrum, in the “B:” labeled<br />

box. The sample label can be edited directly by clicking in the “A:”<br />

or “B:” labeled boxes above the spectrum or by selecting Label from<br />

the Edit pull-down menu.<br />

� Text can be added to the spectrum by selecting “Add Text” from the<br />

Edit pull-down menu.<br />

<strong>EDAX</strong> Basic Procedures- SEM Quant ZAF 1


Imaging/Mapping<br />

Image Collect/External XY<br />

To collect an image:<br />

1. Confirm that the kV and Magnification are correctly entered<br />

into the “IMG” panel. When Column and Stage control are not<br />

available these two parameters will have to be entered<br />

manually. Press the ENTER key after entering each parameter.<br />

2. Put the microscope into external scanning mode. Each<br />

microscope has its own procedures to allow the external<br />

control of its scan circuits. Depending upon the age of the<br />

microscope an enable button may need to be pressed on a third<br />

hardware box or the microscope may have to be put into an<br />

external mode in the microscope software. For newer<br />

microscopes the <strong>EDAX</strong> system will automatically take control of<br />

the beam when the “e-“ button is clicked. To confirm that<br />

Eexternal scan is activated the “Ext XY” button will turn<br />

yellow (active) in the Imaging/Mapping software and the<br />

microscope screen will blank or freeze with an image.<br />

3. Once these are correctly displayed in the image panel click on<br />

the “e - “ button or select “electron image” from the “Collect”<br />

pull-down menu.<br />

� To improve the image quality, click on the "IMG" control panel<br />

button and increase the matrix size to decrease pixelation effects.<br />

Re-collect the image by clicking on the "e-" button again. Typical<br />

images are collected at 512x400 or 1024x800 matrix sizes.<br />

� The “Strips” box denotes the number of times the screen will be<br />

updated during the image acquisition.<br />

Integrated Frames<br />

To decrease the "grain" or noisiness of the image, increase the<br />

number of integrated frames (IntF) by “wiping” or by clicking and<br />

dragging the mouse over the number in the “IntF” box and entering a<br />

new number. This is located in the lower half of the Img panel.<br />

Spectrum Collect and Quantification<br />

To collect a spectrum, the screen display should be set to display<br />

the spectrum and image (the button shows both and is found just to<br />

the left of the “e-“ button). A current image should be displayed.<br />

� Spot scan in the Imaging/Mapping software should be selected ("+"<br />

button found in the button bar at the top of the screen).<br />

� Clear any existing spectrum (Click on the paint roller button).<br />

� Position the cursor on the area of interest in the image<br />

displayed and click the mouse.<br />

� Start collecting the spectrum (Click on the stopwatch button).<br />

� Identify the peaks using the Peak ID control panel just as is<br />

done in the ZAF32 program.<br />

� When the peaks are of adequate size and quality, stop the<br />

spectrum collection (Click on the stop-watch button again).<br />

� To improve the spectrum display, you can click and drag on the<br />

spectrum to stretch it both horizontally and vertically. The<br />

buttons above the spectrum are used just as in the ZAF32 program and<br />

a text label can be entered just above and to the right of the<br />

spectrum (press the "enter" key following the label text entry.<br />

� To quantify a spectrum, click on the “Q” button in the button bar of<br />

the spectrum area. The quantification results appear in the upper<br />

right section where the histogram of the image is displayed. This<br />

quantification is a standardless quantification.<br />

<strong>EDAX</strong> Basic Procedures- Imaging/Mapping 1


Line Scan (option)<br />

The <strong>EDAX</strong> Line Scan option in the Imaging/<br />

Mapping software program allows the user to<br />

define a line for data collection, collects data<br />

along that line, then opens a Microsoft Excel<br />

spreadsheet containing the results (net<br />

intensities, weight percent concentrations, and<br />

atomic percent concentrations). This<br />

spreadsheet can be manipulated in Microsoft<br />

Excel to create line graphs. The following<br />

application note is designed to walk the user<br />

through these steps and provide a few tricks to<br />

make these line scans look professional.<br />

Collecting Line Scan Data<br />

To collect a line scan in the Imaging/Mapping<br />

program capture an image from the<br />

microscope. Collect a spectrum from the area<br />

of interest and identify the elements. Set up<br />

regions of interest from the element list.<br />

Once elements and ROI’s are defined the line<br />

scan can be collected.<br />

1. Open the Line scan panel and click on the<br />

line button .<br />

2. Draw a line across the area of interest on<br />

your image by clicking and dragging the<br />

mouse.<br />

3. Choose a dwell time that allows enough<br />

time to collect valid data (This is<br />

dependent upon count rate.)<br />

4. Select the number of collection points<br />

along the line.<br />

5. Click on the Line scan button, “L”, and<br />

enter a name for the data to be saved<br />

under.<br />

6. Click on OK and the line scan will begin.<br />

When the line scan is complete an Excel<br />

spreadsheet should be opened containing all<br />

of the acquired data.<br />

**Tip: If you stop a line scan during the<br />

collection be sure to close the Excel<br />

spreadsheet that is created before beginning<br />

the line scan again.<br />

Creating a Line Scan in Microsoft Excel<br />

Highlight the columns and rows of information<br />

you wish to graph.<br />

**Tip: Choose elements that are close in<br />

concentration so that the scaling allows for<br />

nicely displayed line graphs. This may require<br />

copying columns of data from the original<br />

spreadsheet and pasting into a new sheet.<br />

<strong>EDAX</strong> Basic Procedures- Line Scan<br />

Once the data is selected:<br />

1. Click on the Chart Wizard button in the<br />

toolbar or select “Chart” from the “Insert”<br />

pull- down menu.<br />

2. Choose line graph from the options of graph<br />

type.<br />

3. Setup procedures will prompt you through the<br />

initial creation of your line scan. During these<br />

steps you can label the line scan, define the<br />

X and Y axes, and create a legend.<br />

4. Once these parameters have been chosen<br />

click Finish for a look at the general format of<br />

your line scan.<br />

Wt %<br />

3<br />

2.5<br />

2<br />

1.5<br />

1<br />

0.5<br />

0<br />

Sample XYZ<br />

1<br />

66<br />

131<br />

196<br />

261<br />

326<br />

391<br />

456<br />

Advanced Line Scan Procedures<br />

The line scan above is a quick and easy creation<br />

using Microsoft Excel. However, correlating the<br />

line scan and image can be difficult without the<br />

image in the background of the line scan. The<br />

following procedures are provided in the Microsoft<br />

Excel Help section.<br />

Add a picture to a chart item<br />

Use this procedure to add a bitmap to the chart<br />

area, the plot area, and the legend in 2-D and 3-D<br />

charts.<br />

1. Double click on the plot area to open the<br />

Format Plot Area window.<br />

2. Click on the Fill Effects button, select the<br />

Picture tab, and click on Select Picture.<br />

3. Imaging/Mapping automatically saves the<br />

image with the line scan location to the folder<br />

you designated before starting the line scan.<br />

4. Click on OK to exit the Format window.<br />

Weight %<br />

2.5<br />

2<br />

1.5<br />

1<br />

0.5<br />

0<br />

1<br />

53<br />

105<br />

157<br />

Sample XYZ<br />

209<br />

**Tip: Line scans display best when collected<br />

from left edge to right edge.<br />

261<br />

313<br />

365<br />

417<br />

469<br />

AlK<br />

Y L<br />

TiK<br />

AlK<br />

SiK<br />

TiK


SEM Multi Point Analysis<br />

Spectrum Collect<br />

� Near the upper left of the screen are the “Start” and “Clear”<br />

buttons. Once the Start button has been clicked and the acquisition<br />

has begun the button becomes a “Stop” button.<br />

� The spectra can be saved under the “File” pull-down menu.<br />

Peak ID<br />

Click on “ID” from the menu bar will provide a peak ID panel (which<br />

can be removed by clicking on “R”) which will do an “Auto” peak ID.<br />

Elements can be highlighted from the “Elem List” and deleted<br />

(“Del”). The “Z-” and “Z+” keys will move up and down the periodic<br />

chart by atomic number. If the goal is to identify a specific peak<br />

in the spectrum, then click the right mouse button on the peak and<br />

look at the suggestions in the “Possible” list box.<br />

Standardless Quantification<br />

If the spectrum has been collected for a good length of time<br />

(approximately 100 Live seconds for statistical accuracy) and all<br />

elements are identified then by clicking on the “Conc” in the menu<br />

bar at the top of the screen will apply a standardless<br />

quantification on the spectrum. The results will appear in a<br />

spreadsheet dialog box. These results can be printed, translated<br />

into graphs, and save from this dialog box.<br />

Image Collect<br />

To collect an image click on “e-” (to indicate an electron image) in<br />

the button bar and it will collect from the SEM. This image is a<br />

frozen image and will not reflect any changes you have made to the<br />

SEM from the Microscope Control software unless you refresh the<br />

image by clicking on the “e-” button again. The image can be saved<br />

under “File”, “Save as...” and specifying “.BMP” from the list box<br />

and then giving a file name. During subsequent analysis there may<br />

be a variety of graphics superimposed on top of the image which will<br />

be saved with the image if it is saved at that point.<br />

Multiple Point Analysis<br />

In the menu bar, there is a choice “Loca - X” where the X can be<br />

either an S for spot, an L for a line or an M for a matrix. It is<br />

possible to select any of these 3 in the pull-down menu.<br />

� To select spots for analysis, the image must be displayed (see<br />

above), the menu bar should indicate “Loca-S”, and the mouse should<br />

be clicked on the image at the point of interest. Then the “S”<br />

button from the button bar should be clicked to save that location.<br />

More points can be added by clicking on the image and clicking on<br />

“S”.<br />

� Similarly, if lines have been selected and the menu bar now<br />

indicates “Loca-L”, you can click and drag on the image to define a<br />

linear array of points for analysis, then click on “S”.<br />

� For a matrix of points, click and drag a box to define the matrix<br />

and then click on “S” to save these locations.<br />

By default, the line will do 10 points per line and the matrix will<br />

be a 2 X 2 matrix of points. To change either of these, click on<br />

“Loca- “ and “Table...”, then highlight the number you want to<br />

change, and strike the enter key.<br />

� To display the locations on the image, click on “Loca - “ and click<br />

on the listing to display the locations or to display the locations<br />

with their numbers at the bottom of the pull-down menu.<br />

Before beginning the multi-point analysis, you should be sure that:<br />

<strong>EDAX</strong> Basic Procedures- Multi Point Analysis 1


Spectrum Mapping (option)<br />

Collecting Spectral Maps<br />

It is recommended that the following conditions be set up before<br />

collecting a spectrum map:<br />

� The detector geometry should be correct. (Setup: kV & Dist)<br />

� The count rate, dead time, and time constant should be setup for<br />

high throughput (counts per second), 20-40% dead time, and a time<br />

constant to allow for the other two parameters. Typically with a<br />

high count rate the time constant will be short, eg. 2.5, 4, 6.<br />

(Setup: EDAM & Calib)<br />

� An image and spectrum of the area to be mapped should be acquired<br />

and the elements of interest identified.<br />

With this complete the setup for Spectral Mapping can begin.<br />

� Click on SpcMap in the Multi-point Analysis menu bar. The Spectral<br />

Mapping dialog window will open.<br />

� Enter a dwell time. This is dependent upon the count rate. A dwell<br />

of 200 msec is acceptable for a count rate of 3000 cps. A shorter<br />

dwell time can be used if the count rate is higher. Likewise, a<br />

longer dwell time should be used with a lower count rate. The<br />

object is to dwell at each pixel long enough to collect sufficient<br />

data for analysis.<br />

� Select a matrix size. The resolution of your maps will have a<br />

significant effect on the total acquisition time.<br />

� Highlight the extension in the file name and enter four characters<br />

other than Abcd.SPZ. Press the Enter key on the computer keyboard<br />

when you finish typing in the new name.<br />

� Confirm that the SPZ and 10 keV boxes are checked. These are<br />

compression formats that decrease the size of the spectral map file.<br />

� Check the Estimated Time for the maps (displayed below the filename)<br />

to confirm an acceptable collection time.<br />

� Click on the START button.<br />

� The system will begin to acquire the data, dwelling at each pixel.<br />

When the collection is complete the system will beep.<br />

� To build the maps, click on the Build Map button. (Note that even<br />

though you have maps displayed at this point they are not saved or<br />

retrievable without being built.) A window will display an<br />

estimated time for the processing of the data if it is estimated to<br />

take more than two minutes. Choose YES.<br />

Displaying Maps<br />

To view the maps:<br />

� Click on the Show Map button.<br />

� Select the folder of interest and highlight one of the maps.<br />

� Click OPEN. All maps in the folder will be opened for viewing.<br />

� Toggle on the up and down arrows to flip through the maps in the<br />

folder.<br />

� Check the Color box to convert the image to a color scale. The warm<br />

colors indicate high intensity while the cool colors indicate lower<br />

intensities.<br />

To view multiple maps:<br />

� Click on the Adv button.<br />

� Highlight the maps you wish to view. To select more than one, hold<br />

the control button down and select maps with the cursor.<br />

� Click on the View Multiple Maps button. The maps selected are<br />

displayed to the right.<br />

<strong>EDAX</strong> Basic Procedures- Spectrum Mapping 1


Spectrum Utilities<br />

Introduction<br />

Very few customers may be aware of an<br />

extremely handy program that is part of your<br />

<strong>EDAX</strong> system. The Spectrum Utilities program<br />

can typically be found on the C drive, in the<br />

Utilities folder of the <strong>EDAX</strong> programs. Its<br />

executable file is abbreviated to SpecUtil.exe.<br />

SpecUtil was created to allow the user to print<br />

more than one spectrum on a page. In fact,<br />

SpecUtil is formatted to fit 10 spectra on a<br />

single 8.5’ x 11’ page. This feature is very<br />

helpful when a large volume of spectra are<br />

being compiled for a report or for archiving<br />

purposes. Programs that might benefit from<br />

this capability are particle analysis, GSR (gun<br />

shot residue), and multi-point analysis<br />

packages.<br />

Procedures for Printing Multiple Spectra<br />

1. Select File:Open from the menu bar and<br />

locate the file folder that contains the<br />

spectra you would like to print.<br />

2. Select any one of the spectra from the<br />

folder and click on the OPEN button. All<br />

the spectrum files in that folder will appear<br />

listed in the left column of the SpecUtil<br />

window.<br />

**Note This means all spectra meant to be<br />

printed on the same page must be located in<br />

the same folder.<br />

3. Hold down the control button and select<br />

the spectra to be printed.<br />

4. Click on SPC>>Printer in the menu bar.<br />

**Tip: The arrow buttons will manipulate the<br />

view of the spectrum. These should be used<br />

to prepare the spectra for printing. **Note:<br />

The view settings you chose will be applied to<br />

all of the spectra when they are printed.<br />

**Tip: The spectra can be printed in black and<br />

white, color outline, or color solid by toggling<br />

through the Print Display button to the left of<br />

the Printer button.<br />

Previewing Spectra<br />

To preview a spectrum double click on its name in<br />

the list.<br />

For viewing multiple spectra before printing begin<br />

with steps 1-3 of the printing procedures then do<br />

as follows:<br />

4. Turn off the printer option by clicking on the<br />

printer icon (a red X should appear over the<br />

icon).<br />

5. Select a viewing speed by adjusting the scroll<br />

bar.<br />

6. Click on the SCA>>Printer to begin the<br />

preview.<br />

Changing the Header<br />

The page header can be personalized by going to<br />

Help: Intro and entering a new label.<br />

Displaying Particle Images with Spectra<br />

When a CSV data file type is selected from<br />

File:Open Spectrum Utility has the added ability<br />

to read in a Particle & Phase Analysis datasets.<br />

Selecting a CSV dataset will display all the<br />

spectra from a single or multi-field run.<br />

With a particle data set active, other options<br />

become available at the bottom of the applications<br />

window.<br />

• Analysis location coordinates can be burned<br />

into new BMP (Loca*.Bmp) files.<br />

• Particle images can be inserted into the upper<br />

right hand corner of each corresponding<br />

spectrum area on printouts.<br />

<strong>EDAX</strong> Basic Procedures- Spectrum Utilities


Quantitative X-Ray Analysis using ZAF<br />

Spectra collected with the DX system can be analyzed to provide weight and atomic<br />

percent data. Standards can be used for this analysis or the sample may be analyzed<br />

without standards (i.e. “standardless”). In either case, the procedures are very similar<br />

and will be discussed below.<br />

Figure 1. Spectrum of an aluminum-chromium-nickel alloy collected at 15 kV.<br />

When many users attempt to interpret an <strong>EDS</strong> spectrum, there is often a tendency to<br />

infer weight percent from a visual estimate of the peak height and this usually leads to<br />

erroneous interpretations. For instance, in the spectrum in Fig. 1, the Cr peak is about 1<br />

½ times larger than the Ni peak which is about 2 ½ times larger than Al peak. Whereas,<br />

the correct analysis actually has the Ni being about 1 ½ times the Cr and the Ni is just<br />

about 20 times more abundant by weight percent than the Al. Even experienced microanalysts<br />

who have an extensive knowledge of overvoltage and absorption<br />

considerations would be reluctant to infer relative weight percents from a spectrum such<br />

as this.<br />

The correct interpretation of a spectrum would actually be to compare an element’s<br />

peak height to the height of the pure element collected under identical conditions.<br />

<strong>EDAX</strong> Training Course - Quantitative Analysis - page 1


INTRODUCTION<br />

--Quantitative Analysis with Compound Standards<br />

The primary purpose of this discussion is concerned with how to use standard(s) in quantitative<br />

analysis using the ZAF or eDX software. Even when standardless quantitative analysis is<br />

routinely performed, it is always a good idea to obtain a standard of a similar composition and<br />

see how close the standardless analysis is to the reported data. If it is desired or necessary to<br />

improve the accuracy of the analysis, then standards can be used to accomplish this.<br />

The examples discussed below utilize a sample set of 6 mineral spectra (a garnet standard; the<br />

spectra are labeled garnet01.spc to garnet06.spc). The garnet01.spc has been implemented<br />

as the standard to analyze the remaining 5 spectra which have been collected from different<br />

spots on the same standard. Quantitative results are summarized for analyses using the<br />

“Compound Standard” mode, and with the use of SEC factors (Standardless Element<br />

Coefficients).<br />

THE COMPOUND STANDARD<br />

The spectra to be analyzed will typically have been collected from a single session in sequence<br />

using the same element list, with the same accelerating voltage, and with the same beam<br />

current/spot size setting. It is possible to compensate for variations in beam current if a current<br />

meter and Faraday cup are available, but this is primarily of value if the user can not be certain<br />

that the analyzed elements should have a sum which equals 100% (or very near 100%). In this<br />

case, an assumption for the garnet standard of 100% for the reported elements (O, Al, Si, Ca,<br />

Mn and Fe) is certainly valid.<br />

Procedure. In the ZAF or eDX software, open the standard file spectrum (garnet01.spc; click<br />

on “File”, “Open” and specify the file name from the correct directory). The first step will be to<br />

provide the information necessary that will allow us to use this spectrum as a standard.<br />

Although, the sample is an oxide, oxygen will be measured directly as an element; and this<br />

sample was saved with oxygen in the element list and oxides were not checked as the sample<br />

type.<br />

In the space below, mouse clicks will be shown with brackets with the text enclosed within to<br />

designate the button or feature where the mouse is clicked. For example, “[OK]” indicates that<br />

the mouse is clicked on the “OK” button.<br />

-the “Quantify” control panel should be made active.<br />

-[Stds]<br />

-[Options]<br />

-[Compound]<br />

-[Setup]<br />

-Type the percentage for each element followed by an<br />

“enter” (38.78, 10.92, 17.08, 0.24, 15.35, and 17.64 for<br />

the O, Al, Si, Ca, Mn, and Fe, respectively)<br />

-[RZAF] (calculated pure element intensities will be<br />

displayed)<br />

<strong>EDAX</strong> Training Course - Analysis with Compound Standards - page 1


INTRODUCTION<br />

--Quantitative Analysis with Pure Element Standards<br />

The pure element standards method requires that a series of pure element standards be<br />

collected under constant conditions (same geometry, kV, and beam current into a Faraday<br />

cup). The constant conditions does not mean that they should have the same dead time or the<br />

same absorbed current while on the standards because both parameters will vary with the<br />

atomic number and possibly with overvoltage considerations. Probably the dead time should<br />

not be more than 30-40% on the sample or standard with the highest dead time.<br />

THE PURE ELEMENT STANDARDS SETUP<br />

Procedure. In the ZAF32 software, open the first standard file spectrum (”std15ala.spc”; click<br />

on “File”, “Open” and specify the file name from the correct directory). If there is a question<br />

about “save existing spectrum”, click on “No”.<br />

Before we start creating a pure element intensity table of elements based upon our direct<br />

measurements, it will be a good idea if we verify what parameters are associated with our table.<br />

In the Quant. control panel, click on “Stds” and then on “Pure”. Then you should click on<br />

“Factors” in the Quant. control panel and verify that the detector type is correct (probably<br />

“SUTW” and “Sapphire” for most of you) that the method is “ZAF”, and that the voltage is<br />

correct (15 kV in this case).<br />

In the space below, mouse clicks will be shown with brackets with the text enclosed within to<br />

designate the button or feature where the mouse is clicked. For example, “[OK]” indicates that<br />

the mouse is clicked on the “OK” button.<br />

-the “Quantify” control panel should be made active.<br />

-[Stds]<br />

-[Options]<br />

-[Pure]<br />

-[Setup], it will show the measured pure element intensity<br />

-[Save]<br />

-[Yes]<br />

-[OK]<br />

Then open the next spectrum (”std15cra.spc”) and repeat the steps above, followed by the last<br />

of the three pure element standards (”std15nia.spc”) and repeating the above steps again.<br />

When all are entered, then you are ready to begin processing your unknown samples. If you<br />

would like to confirm that everything is as it should be, click on “Factors” in the Quant. control<br />

panel and verify that Al, Cr, and Ni have values that are not 1. Also, the detector type should<br />

match your samples (“SUTW” and “Sapphire” in this case) and that the method is “ZAF”.<br />

If all standards have been processed, it would be a good idea to save a standards file. Click on<br />

“File” and “Save as…”, select the STD file type and give the file a name (e.g. “alcrni15.std”).<br />

The standards file saves a list of all elements, their shell (K, L, or M) and their pure element<br />

intensities. If an element does not have pure element data associated with it, it has a recorded<br />

<strong>EDAX</strong> Training Course --Analysis with Pure Element Standards - page 1


Quantitative Analysis with SEC factors --Hans Dijkstra<br />

For microanalysis the composition of a sample can be calculated from the measured X-ray<br />

intensities using:<br />

Meas<br />

W Z A F I<br />

% = • • •<br />

I<br />

Where Z, A and F are the matrix correction parameters, describing the atomic number effect<br />

(stopping power and backscatter effect), the absorption effect, and the fluorescence effect.<br />

Since for standardless analysis we have no Ι STD available, <strong>EDAX</strong> has chosen to use calculated<br />

standards, with in a simplified form:<br />

N<br />

I n p f<br />

A R<br />

Q E<br />

dE d s dE<br />

0 E j<br />

Std Ω<br />

j ( )<br />

Calculated = εεεε dωωωω j jl ( χχχχ)<br />

����<br />

4ππππ<br />

E / ( ρρρρ )<br />

0<br />

Where n is the number of electrons entering the sample, Ω/4π is the solid angle, ε is the detector<br />

efficiency, ω is the X-ray fluorescence yield, p is the relative probability for the transition involved,<br />

f(x) is the absorption correction, and the integral represents the cross-section of the ionization<br />

involved. Since n is unknown, and thus set to 1, the calculated intensity might be in a totally<br />

different order of magnitude as the measured intensity. Normalizing the W% to 100% solves this<br />

problem.<br />

This function seems to work rather accurate, but it is important to notice that some factors are left<br />

out of the calculations, like the solid angle, since this is a constant factor and this equation is only<br />

used for standardless analysis, i.e. the results are normalized to 100% anyway.<br />

One disadvantage of this equation is that εd, the detector efficiency, can not be predicted with<br />

sufficient accuracy for X-ray lines below 1 keV. Small variations in detector quality (Si dead layer,<br />

etc.) can cause variations in measured intensity. Therefore <strong>EDAX</strong> has introduced the SEC factors.<br />

The final equation now becomes:<br />

I<br />

W% = Z • A•F •<br />

SEC •<br />

I<br />

The SEC factors can simply be calculated by entering a compound standard, and calculate the SEC<br />

from the given W% (thus the ZAF factors and the standard intensity can be calculated) and the<br />

measured intensities. Also in this procedure calculated SEC factors may be off by an order of<br />

magnitude, and now this is solved by assuming one SEC factor to be identical to a default value<br />

(thus keeping it fixed), and scaling other SEC factors relative to the fixed one.<br />

________________________________________________________________________<br />

<strong>EDAX</strong> Training Course –Quantitative Analysis with SEC Factors - page 1<br />

Meas<br />

Std<br />

Std<br />

Calculated


Introduction to Digital Imaging<br />

Digital images are defined by the number of pixels (picture elements) in a line across the<br />

image, by the number of lines and the number of colors or gray levels in the image. The<br />

number of gray levels is usually expressed in bits (8, or 12 bits) which is the exponent used<br />

with the number 2. An 8 bit image has 256 levels (0 - 255), and a 12 bit image would have<br />

4096 (0 - 4095). At each a pixel a number is stored to represent or color or gray level. In<br />

most imaging programs on the PC, the ‘0’ value represents black, the highest value<br />

represents white and the intermediate values are shades of gray.<br />

As illustrated in the above figure, our work with images can be thought of as being image<br />

processing if we start with an image and if the result is an image. The image is processed<br />

or enhanced; perhaps sharpened, or the contrast is modified. Image analysis would involve<br />

any operation in which we start with an image and the results of the operation are numbers<br />

or data.<br />

A portion of a digital image is shown below. This is an 8 bit grayscale image which consists<br />

of 32 pixels per line and 32 lines of data. As with all digital images, the image can be<br />

zoomed to the point where the pixels become obvious. The four images at the below left<br />

show the same number of pixels shown with a different zoom factor.<br />

It is possible to view the actual numerical data for the image. A profile of pixel values are<br />

shown below for a horizontal line across the middle of the eye.<br />

52 36 26 25 71 114 126 128 149 176 182 90 27 45 51 62 50 68 48 40 39 29 99 126 68 36 35 30 39 45 57 77<br />

<strong>EDAX</strong> Phoenix Training Course - Introduction to Digital Imaging - page 1


<strong>EDAX</strong> Phoenix Procedures – PhotoImpact<br />

Introduction<br />

PhotoImpact is an image processing program that provides many additional features<br />

not available in the Phoenix imaging software. This program is comparable in many<br />

ways to Adobe Photoshop and has a manual, training software, as well as on-line<br />

help to describe many detailed procedures that can be implemented using this<br />

software. In x-ray microanalysis and electron microscopy, there is really only a small<br />

subset of its capabilities that most of us will ever need. It is intended that the<br />

following few pages will describe those procedures to enable the user to use this<br />

program without having to become an expert on image enhancement software.<br />

PhotoImpact actually consists of several programs. Four of the programs that will be<br />

mentioned at least briefly here are:<br />

(1) PhotoImpact, a basic image enhancement program that is most similar to<br />

Photoshop;<br />

(2) PhotoImpact Album, a program that allows the user to create thumbnail images,<br />

to print multiple images on a single page and to create a slide show of images;<br />

(3) PhotoImpact Capture, a screen capture program which is a fairly simple program<br />

that will not be described in detail here;<br />

(4) PhotoImpact Viewer, a simple image viewing program which can be set up in the<br />

Explorer to be the program opened when you double click on an image file –it is<br />

the program that the Album program opens when you double click on a thumbnail<br />

image.<br />

Most of the PhotoImpact programs will allow you to open multiple files. Once an<br />

image is opened it is possible to zoom it by clicking on the “+” or “-“ keys. Also, most<br />

operations can be undone (Ctrl-Z) and re-done (Ctrl-Y).<br />

Procedures<br />

In all procedures below the text shown in brackets represents a mouse click (e.g.<br />

“[OK]” would indicate a click on the “OK” button). Tabs are used to show the<br />

hierarchy of commands (i.e. a menu bar selection is shown to the left, with one<br />

additional tab to indicate a click from the pull-down menu, and an additional tab to<br />

show a click from the resulting dialog box, etc.).<br />

Most steps in image enhancement can be undone if it is decided that the result was<br />

not optimal. This is accomplished by clicking on “Edit” and “Undo”. The shortcut to<br />

undo an operation is Ctrl-Z and Ctrl-Y will re-do the same operation. There are<br />

multiple levels of undo for an image and the number of undo’s can be specified by<br />

clicking on File and Preferences. The higher the number of undo’s that you specify<br />

will be more memory intensive. It is possible, or a good idea to save a modified<br />

image with a new name to protect yourself from lost data when there are lengthy<br />

image processing protocols.<br />

To Open an Image<br />

<strong>EDAX</strong> Phoenix Training Course - PhotoImpact - page 1


Report Writing with Microsoft Office<br />

It is often desirable or a requirement to create reports that integrate text interpretation with data<br />

(spectra, quantitative results, images, etc.). MS Word and MS Excel are readily available word<br />

processing packages that can be used to prepare such a report.<br />

The spectral data in any <strong>EDAX</strong><br />

application is typically stored with<br />

the file extension .SPC. These files<br />

can be stored from, and recalled to<br />

<strong>EDAX</strong> applications only. When the<br />

goal is to store a spectrum that<br />

can be inserted into a Word<br />

report, the spectrum should be<br />

saved as a BMP or TIF file.<br />

** Note When saving the spectrum<br />

as an image it will be saved as it<br />

appears in the spectral window of<br />

the ZAF program. Therefore, if<br />

there is an area of interest the user<br />

should click and drag that area into<br />

view before saving. Figures 1-3<br />

demonstrate this feature.<br />

The spectrum, as an image file, can<br />

then be inserted into the Word<br />

document easily. Drawing a “text<br />

box” in MS Word and inserting the<br />

picture into the box will determine<br />

its size and location. The three<br />

figures to the left were inserted this<br />

way, as was this text on the right.<br />

The black outlines of the text boxes<br />

have been left visible to<br />

demonstrate this technique.<br />

• To insert text into these boxes<br />

simply click with the cursor<br />

inside of the box and start<br />

writing.<br />

• To insert an image in a text box<br />

position the cursor in the box<br />

and select Insert: Picture:<br />

From File.<br />

The images can be previewed<br />

in this dialog box, the desired<br />

image chosen, and inserted<br />

into the document.<br />

Figure 1 Entire spectrum saved as image file.<br />

Figure 2 Low end of spectrum saved as image.<br />

Figure 3 Peaks of interest saved as image file.<br />

<strong>EDAX</strong> Training Course - Report Writing with MS Word - page 1<br />

“Text boxes” can<br />

be directly added<br />

to the graphic.<br />

The text box color<br />

and border can be<br />

modified by<br />

double clicking<br />

within them.


Summary<br />

Parameters<br />

Typical parameters for good resolution (spectrum dominated by peaks with energies greater than<br />

1 keV) with the fewest artifacts:<br />

--kV > 2X highest energy peak<br />

--Time Constant = 50 or 100 us<br />

--Deadtime = 20 to 40 % (adjusted by changing count rate)<br />

--Take-off angle = 25 - 40 degrees<br />

--Working distance = intersection distance (for inclined detectors)<br />

When the spectrum has significant peaks with energies less than 1 keV, the parameters should be<br />

the same as above except that the count rate should be 500 to 1000 cps. When the sample is to<br />

be mapped, the same parameters should be used except that a count rate of 10,000 to 100,000<br />

cps (10 to 2.5 us time constant) should be used in order to improve the statistical quality of the<br />

map. This assumes that the sample will tolerate the heat caused by the higher beam currents,<br />

that the larger spot sizes do not degrade the resolution of the image or maps, and that the very<br />

low energy peaks are not of primary interest which usually require lower count rates and a longer<br />

time constant.<br />

Artifacts<br />

Escape Peaks = keV of parent peak - 1.74 keV (Silicon Ka energy).<br />

Sum Peaks = 2X energy of the dominant peak in a spectrum, or<br />

the energy of peak A + peak B when two peaks are dominant.<br />

Stray Radiation or “System Peaks” = Peaks derived from the pole piece, stage,<br />

sample holder or detector window support. These are usually more<br />

significant with horizontal <strong>EDS</strong> detectors.<br />

Peak ID<br />

Auto Peak ID will usually be adequate when the spectrum is dominated by K-series peaks. When<br />

the spectrum contains several L- or M-series peaks, the best strategy may be to manually identify<br />

the highest energy alpha peak and observe what other low-energy peaks are also associated with<br />

it. Then continue with the next highest energy alpha peak and continue until all are identified.<br />

Quantification<br />

Whether the quantification involves the use of one or more standards, or if it is a standardless<br />

analysis, the procedure is the same: the peak intensities are calculated apart from the<br />

background; the peak intensities are compared to intensities of the pure elements to calculate the<br />

k-ratio; and corrections are made for atomic number (Z), absorption (A) and fluorescence (F).<br />

When standards are used, the pure element intensities are actually measured, but these can be<br />

calculated when no standard is present. The ZAF corrections assume that the sample/detector<br />

geometry is well known, that the sample is smooth, and that it is homogeneous. If these<br />

assumptions are accurate, then the quantitative results will also be accurate.<br />

<strong>EDAX</strong> Phoenix Training Course - Summary - page 1


To Create and Save an Anaglyph Color Stereo Image with Photoshop<br />

This procedure assumes that you have created and saved the two images from which you want<br />

to create and save a color stereo image. Typically, these two images will differ in tilt by 5 to 10<br />

degrees and will have been collected at a eucentric working distance if one is available. Samples<br />

with a lot of surface topography (e.g. fibers or needle-shaped crystals) will require less tilt than<br />

relatively flat samples. The resulting stereo image will require a pair of red-cyan glasses to view<br />

the color anaglyph image. These glasses have a red lens for the left eye and a cyan lens for the<br />

right eye. Red-green glasses can also be used.<br />

Commonly, the stereo images will be collected with a minus 90 degree scan rotate. In order for<br />

the stereo pair to have the correct parallax, the tilt axis should be vertical on the display. The<br />

image collected at a lower tilt can be saved with a number after a file name indicating the tilt at<br />

the end of the file prefix (or an "r" or "red" to indicate which image this is in the pair), and the<br />

higher tilt image saved as well with the tilt indicated (or a "c" or "cyan" included in the file name).<br />

The two images that make up the stereo pair will be named "fracred.tif" and "fraccyan.tif" for<br />

instance.<br />

Ideally, the eucentric working distance, or the point where image registration occurs will be<br />

located at a mid-height position in the sample. This permits a maximum amount of topography to<br />

be portrayed in the resulting stereo image. Those portions of the sample that are at the eucentric<br />

position will show no color shift between the two images and will appear to be grey or have no<br />

color because the two images are in registration and the red and cyan sum to a neutral grey.<br />

Areas of the anaglyph image that appear to be "up" in the image will have the red image shown to<br />

the right of the cyan. Those areas of the stereo image that appear to be "down" will have the red<br />

image shifted to the left of the cyan image. When the amount of shift becomes too excessive, it<br />

becomes difficult for the brain to "fuse" the two images. This will occur if an excessive amount of<br />

tilt differential is applied to a sample with normal topography or surface roughness.<br />

Creation of the Color Anaglyph Image in Adobe Photoshop<br />

Two methods will be discussed for creating a stereo pair. Method II is probably the simpler of the<br />

two. These procedures have been used and tested in versions 3 and 4 of Photoshop.<br />

Method I:<br />

1. The two images that will make up the stereo pair should be recalled from disk using the<br />

normal procedure for recalling files ('File', 'Open', and then selecting the appropriate file type,<br />

directory and file name). In version 4 you can select both images by clicking on the first one<br />

and then a Ctrl-click on the second image from the file list.<br />

2. The cyan image should now be duplicated by clicking on the cyan image to make it the active<br />

image (i.e. the title bar above the image should be blue), followed by clicking on 'Image',<br />

'Duplicate...', and clicking 'OK' to accept the default image name for the duplicate.<br />

3. The next step below will require the use of the "Channels" palette or box. If this box should<br />

appear on the monitor, you should skip to step number 4. If this box can not be seen, click<br />

on 'Window', 'Palettes', and 'Show Channels'.<br />

4. In the Channels palette box there will be some text near the top of the box that reads<br />

"Layers", "Channels" and "Paths". If "Channels" is not already highlighted, you should click<br />

on it and then click on the right-facing triangle to its right (>); then click on 'Merge<br />

Channels...'. For "Mode" you should select 'RGB Color' with the channels equaling "3", then<br />

click on 'OK'. In the "Merge RGB Channels" dialog box you should assign the red or lower tilt


To Create a Stereo Image, page 2<br />

image to red and assign the cyan image and its duplicate to the green and blue channel, then<br />

click 'OK'.<br />

5. An untitled RGB color anaglyph stereo image should now appear on the monitor. Prior to<br />

saving the image, the contrast and brightness should be checked and adjusted if necessary.<br />

This can be done most easily with the "Adjust Levels" procedure which is implemented with a<br />

"Ctrl-L". This will bring up a dialog box that has a histogram and adjustors for black, the<br />

midtone and white. These adjustors are immediately below the histogram and they should be<br />

adjusted to provide an optimum image on the monitor. As a general rule, it probably makes<br />

sense to slide the black adjustor to the beginning of the left side of the histogram, and the<br />

white to end or to the right side of the histogram --this should be approximately equivalent to<br />

clicking on 'Auto'. The image displayed should be updated to reflect these changes --if it has<br />

not, this probably means that the "Preview" box has not been checked. If the resultant image<br />

appears to be too light, then you should move the midtone adjustor to the right. If the image<br />

is too dark, the midtone adjustor should be moved to the left. The midtone adjustment is<br />

most equivalent to the gamma function of the SEM. When you are safisfied with the<br />

appearance of the image, click on 'OK'.<br />

6. To check the 3D image, put the anaglyph glasses on (red for the left eye and cyan for the<br />

right eye) and see if the topography appears normal. If it does, proceed to step 7.<br />

Depending upon how the original images were collected or saved, it may be that the colors<br />

will have to be re-assigned or the glasses reversed. Also, if the tilt axis is not vertical in the<br />

two images as displayed, it may be necessary to do a 90 degree rotation of the image at this<br />

time (click on 'Image' and 'Rotate' and chose a clockwise or a counter-clockwise rotation of<br />

90 degrees.<br />

7. The color stereo image can now be saved by clicking on 'File' and 'Save As...', or by a "Ctrl-<br />

S". The file can be saved in a variety of formats, but the most common are probably the TIFF<br />

or JPEG. If saved as a TIFF, the resulting file will be approximately 1 MB or three times the<br />

size of the original standard definition XL SEM file. If saved as a JPEG with a "High" image<br />

quality setting, the file will probably be about 1/4 to 1/6 of the size of the TIFF depending on<br />

the level of fine detail in the image.<br />

Method II:<br />

1. The two images that will make up the stereo pair should be recalled from disk using the<br />

normal procedure for recalling files (“File”, “Open”, and then selecting the appropriate file<br />

type, directory and file name). In version 4 you can select both images by clicking on the first<br />

one and then a Ctrl - click on the second image from the file list.<br />

2. Convert the higher tilt image (the cyan image) to an RGB color image by clicking on “Image”,<br />

“Mode”, and “RGB Color”. Display the red channel by pressing a “Ctrl – 1”.<br />

3. Activate the lower tilt image by clicking on the image. This image should be converted to a<br />

grayscale image if it is not one already by clicking on “Image”, “Mode”, and “Grayscale”.<br />

4. Select the entire image with a Ctrl - a and then copy it to the clipboard with a Ctrl - c.<br />

5. Re-activate the higher tilt image with its red channel actively displayed, and then paste the<br />

image from the clipboard with a Ctrl - v.<br />

6. Display the entire color image with a Ctrl - ~. At this point you can modify the image and/or<br />

save it to disk just as in steps 5 – 7 in Method I above.<br />

Bob Anderhalt<br />

<strong>EDAX</strong> Applications Laboratory<br />

June 17, 1999


Another common file format supported by <strong>EDAX</strong> applications is the CSV format (comma<br />

separated values). Results files and summary tables are typically saved as CSV files. Spectra<br />

can also be saved in this format. This is actually a text format that is most commonly used to<br />

input data into a spreadsheet program from which it can be plotted or have additional calculations<br />

made. Portions of that spreadsheet file can be highlighted and copied to the clipboard and pasted<br />

into the Word report directly (see below) or into a text box.<br />

Inserting Quantitative Results into a<br />

Report<br />

1. In MS Excel<br />

• Click and drag to highlight the information<br />

to be copied.<br />

• Select Copy from the EDIT pull down<br />

menu.<br />

2. In MS Word<br />

• Draw a text box or position the cursor<br />

where the results should appear<br />

• Select Paste from the EDIT pull down<br />

menu.<br />

Advanced Options<br />

• Highlight the columns of the table<br />

• Select Table Autoformat from the Table<br />

pull down menu<br />

• Choose a style<br />

• Click OK<br />

Graphs and charts created in MS Excel can also<br />

be copy and pasted into an MS Word document.<br />

14%<br />

Stainless Steel Wt%<br />

62%<br />

1% 3%<br />

18%<br />

SiK<br />

MoL<br />

2%<br />

CrK<br />

MnK<br />

FeK<br />

NiK<br />

Element Wt % At %<br />

SiK 0.75 1.48<br />

MoL 2.58 1.5<br />

CrK 17.89 19.14<br />

MnK 1.72 1.74<br />

FeK 63.4 63.18<br />

NiK 13.67 12.96<br />

Total 100 100<br />

Element Wt % At %<br />

SiK 0.75 1.48<br />

MoL 2.58 1.5<br />

CrK 17.89 19.14<br />

MnK 1.72 1.74<br />

FeK 63.4 63.18<br />

NiK 13.67 12.96<br />

Total 100 100<br />

Element Wt % At %<br />

SiK 0.75 1.48<br />

MoL 2.58 1.5<br />

CrK 17.89 19.14<br />

MnK 1.72 1.74<br />

FeK 63.4 63.18<br />

NiK 13.67 12.96<br />

Total 100 100<br />

<strong>EDAX</strong> Training Course - Report Writing with MS Word - page 2


The report can also include the BMP or TIF files from x-ray maps or electron images. Just as with<br />

any BMP file, they can be inserted by clicking on “Insert”, “Picture” and the file names provided as<br />

usual.<br />

Fig. 1. BSE Image of an aluminum Fig. 2. Aluminum x-ray map.<br />

alloy fracture surface.<br />

After inserting the first image (Fig. 1), the cursor is shown at the lower right of the image. By<br />

pressing the space bar a few times before doing the next insert, the second image will appear to<br />

the right of the first but on the same line. After Figure 2 was inserted, the Enter key is pressed<br />

two times to provide some space between the first and a possible second row of images. This<br />

area can also be used for labeling the images.<br />

If the x-ray maps had been collected at lesser resolutions, it would have been possible to place<br />

more than two of them on a single line. Alternatively, it would also be possible to resize each<br />

image by clicking on it and adjusting one of the corners. For multiple images, the percentage of<br />

enlargement should be noted, so that it can be reproduced on each additional image of the series.<br />

a. BSE b. Phosphorus c. Silicon d. Titanium e. Carbon f. Oxygen<br />

When creating a template document where the images may not have been collected yet, or are<br />

being collected at the same time as the report is being generated, the text-box feature in MS<br />

Word is handy. The steps for inserting a text-box are the same described on the previous page.<br />

Inserting an image or pasting an image can be done right into the text-box. This is also useful<br />

when handling large images and maps as a standard image insert would occupy a large portion of<br />

the page. The images below have a 1024x800 resolution and have been reduced to fit the size of<br />

the text-box.<br />

<strong>EDAX</strong> Training Course - Report Writing with MS Word - page 3


The eDXAuto/Multi-Point Analysis software package provides integrated imaging with automated<br />

analysis including line scan and multi-point analysis features. Some of the graphic overlays on an<br />

image are shown in the four examples below. Each image is saved as a BMP file when the image<br />

area appears as shown. The image is refreshed (to Fig. I below) by clicking on “Image” and<br />

“Display Current”. Images are inserted in Word the same as described previously. It is possible<br />

to overlay more than one line scan on a single image, but the data are usually communicated best<br />

with a single line scan. There is a direct printout in which all line scans are printed in sequence on<br />

a single page.<br />

I. Sandstone image area (BSE). II. Image with 14 multi-point locations.<br />

III. Silicon linescan along blue line. IV. Calcium linescan.<br />

The results of a multi-point analysis are provided by eDXAuto as a summary table that can be<br />

highlighted and copied to the clipboard. It is then pasted into Word where it can be highlighted<br />

and a font size selected to fit the page. Important samples and analysis locations can be<br />

highlighted by underlining them or by making them boldface or a larger font. In the example on<br />

the following page, a table title and caption were added to the summary table.<br />

<strong>EDAX</strong> Training Course - Report Writing with MS Word - page 4


Table 1. Results of multi-point analyses from mineral sample. The sample<br />

locations with high iron concentrations are shown in boldfaced type and italics.<br />

DxAuto Results Summary : C:\DX4\AUTO_P\USR\SSA101.SMY 05-27-1997 17:07:44<br />

KV: 20.0 Tilt: 0.0 TKOff: 36.8<br />

Wt%: C K O K NaK MgK AlK SiK P K K K CaK TiK FeK<br />

Note<br />

ssa101 3.57 54.51 0.15 0.00 0.24 40.43 0.63 0.11 0.10 0.07 0.20<br />

ssa102 9.38 53.54 2.08 6.59 2.96 8.42 0.11 0.41 12.91 0.20 3.39<br />

ssa103 19.32 46.58 0.24 7.96 0.51 1.39 0.36 0.74 19.29 0.14 3.47<br />

ssa104 21.95 32.84 1.29 0.65 12.22 19.25 0.79 7.72 0.64 1.21 1.44<br />

ssa105 8.68 53.46 0.33 5.33 1.70 2.27 0.30 0.78 15.78 0.07 11.31<br />

ssa106 3.95<br />

ssa107 8.87<br />

ssa108 14.31<br />

ssa109 9.98<br />

ssa110 10.19<br />

ssa111 4.36<br />

ssa112 3.99<br />

48.55<br />

53.82<br />

46.41<br />

54.75<br />

46.35<br />

48.05<br />

48.25<br />

0.45<br />

0.37<br />

0.55<br />

0.38<br />

4.32<br />

0.38<br />

0.51<br />

0.17<br />

5.83<br />

5.41<br />

8.55<br />

0.72<br />

0.18<br />

0.21<br />

8.68<br />

0.17<br />

3.43<br />

0.49<br />

7.35<br />

8.69<br />

8.87<br />

26.93<br />

0.23<br />

3.97<br />

0.68<br />

21.43<br />

26.91<br />

26.78<br />

0.15<br />

0.14<br />

0.60<br />

0.21<br />

0.15<br />

0.18<br />

0.52<br />

10.34<br />

0.23<br />

1.12<br />

0.24<br />

0.09<br />

10.39<br />

10.29<br />

0.13<br />

19.73<br />

14.63<br />

18.45<br />

3.53<br />

0.16<br />

0.00<br />

0.19<br />

0.16<br />

0.00<br />

0.18<br />

0.00<br />

0.16<br />

0.11<br />

0.48<br />

**Note 10.46<br />

This 9.57table<br />

cannot<br />

be 6.09 changed using<br />

the 5.85Autoformat<br />

0.55 because it was<br />

0.46 pasted as text.<br />

ssa113 4.07 47.94 0.43 0.18 8.77 27.08 0.53 10.48 0.16 0.06 0.30<br />

ssa114 16.59 39.00 1.36 0.85 19.50 14.87 0.80 3.92 1.07 0.41 1.64<br />

ssa115 4.16 48.21 0.19 0.18 8.67 27.05 0.45 10.58 0.26 0.05 0.19<br />

The ability to interface with standard desktop publishing, spreadsheet, and image enhancement<br />

software is primarily permitted by the use of three universal file formats (BMP, TIF, and CSV).<br />

This ability is further enhanced with graphics overlays on the image and by cutting and pasting to<br />

the Windows clipboard. Thus providing an easy working environment for the user.<br />

<strong>EDAX</strong> Training Course - Report Writing with MS Word - page 5


Procedure<br />

-[File]<br />

-[Open]<br />

Select the correct drive and directory according to normal Windows conventions.<br />

Under File types, it is possible to select “All Files”, “BMP”, “TIFF”, or 25 other image<br />

file formats. It is possible to double click on a single file to view it in PhotoImpact, or<br />

to click on the first file of interest and then to Ctrl-click on additional images and then<br />

clicking on “Open” to activate multiple images. Similarly, clicking on the first file from<br />

a set of images, followed by a Shift-click on the last image from the set will allow you<br />

to “Open” all images from that set.<br />

Changing Brightness & Contrast<br />

Setup. An image should be open and active in PhotoImpact. To open an image,<br />

see procedure described above.<br />

Procedure<br />

-[Format]<br />

-[Brightness & Contrast]<br />

Click on the thumbnail with the most optimal brightness and contrast. Repeat if<br />

necessary, then [OK]. An alternative is shown below:<br />

-[Format]<br />

-[Tone Map]<br />

Click on the “Highlight Midtone Shadow” tab. Optimize the image with the sliders,<br />

then [OK].<br />

To Change Image Mode<br />

Images collected in the Phoenix imaging software will be “palette color” images (256<br />

colors) if it was an x-ray map or if it was an electron image. In order to print an image<br />

well or to perform an image processing operation, it is often necessary to change an<br />

image/map to either a gray-scale image or an RGB color image. With the image that<br />

you want to change open in PhotoImpact and active, [Format], [Data Type] and then<br />

click on the desired file type (typically, “Grayscale” or “True Color”. Note that the<br />

current image type will be gray-ed out and not selectable.<br />

To Create a 3D Anaglyph Image from a Stereo Pair<br />

Setup The 3D anaglyph image contains two images, one which is shown in red and<br />

the other is either green or blue-green (cyan). When viewed with the corresponding<br />

color glasses, each eye sees a different image where the differences in the image<br />

corresponds to a difference in tilt or perspective, and our brain reconstructs a 3dimensional<br />

image just like it normally does.<br />

<strong>EDAX</strong> Phoenix Training Course - PhotoImpact - page 2


It is assumed that you have a stereo pair on disk that it has been recalled to<br />

PhotoImpact. The two images should have the same pixel dimensions (e.g.<br />

512x400, or 1024x800) and they should differ only by their tilt. Also, the tilt axis<br />

should be vertical on the monitor with the SE detector on the left (on many SEMs this<br />

will require a -90 degree scan rotate). If the conditions for the two images are<br />

different than just described, you may need to depart from the procedure described<br />

below (i.e. the color designation may need to be changed, and/or the merged image<br />

may need to be rotated). In PhotoImpact the 3D image is created using a CMYK<br />

(cyan, magenta, yellow, black) merge function. One image will be assigned to cyan,<br />

the other to both magenta and yellow (this makes red) and a white image with no<br />

video information will need to be created and assigned to the black channel.<br />

It will be helpful when saving the two original images to denote the tilt of each image,<br />

or to append a letter for the color of each image (e.g. an “r” for the red image and a<br />

“c” for the cyan image).<br />

Procedure<br />

The two images should be converted to grayscale images; [Format], [Data Type] and<br />

[Grayscale]. You may need to keep a note as to which image is the lower tilt image<br />

because the conversion process will create a new image called “Untitled X” where “X”<br />

is a sequence number. When both images have been converted to grayscale, then it<br />

will be necessary to create a white image with the same definition (512x400 or<br />

1024x800, for instance) as the two images that differ in tilt. To create a new image<br />

that is completely white:<br />

-[File]<br />

-[New]<br />

A dialog box will appear, and you should ensure that the data type is grayscale (8bit),<br />

the image size matches the resolution of the two images (512x400 or 1024x800,<br />

for instance), then [OK]. If the image is a white image, you are ready to proceed. If it<br />

is not white that is most likely because the background color has been selected to be<br />

something different. In that case, click on the background box under the grayscale<br />

palette at the right side of the screen; then enter 255 for each of red, green and blue.<br />

When the two images have been converted to grayscale and the white image is<br />

available:<br />

-[Format]<br />

-[Data Type]<br />

-[Combine from CMYK]<br />

A dialog box appears. In PhotoImpact (different from several other image<br />

enhancement programs) you should assign the lower tilt image to cyan, the higher tilt<br />

image to both magenta and yellow, and the white image to the black channel, then<br />

[OK].<br />

Check the resultant image with your glasses (red for the left eye, cyan for the right<br />

eye). If the image does not work, see note at the end of the second paragraph of<br />

<strong>EDAX</strong> Phoenix Training Course - PhotoImpact - page 3


this section (i.e. the color designation may need to be changed, and/or the merged<br />

image may need to be rotated). To determine if the image needs to be rotated, turn<br />

your head sideways to see if the 3-dimensionality improves. If the image has what<br />

should be down as up (and vice-versa), then switching the glasses around will help to<br />

clarify this situation; this indicates that the original color assignments of the two<br />

images were incorrect.<br />

The anaglyph image can be saved as an RGB or true color TIFF or BMP file. It can<br />

also be saved as a JPG file. This is a compressed type of file that will take up less<br />

room on disk but can show some degradation in image quality. For stereo color<br />

images, saving a JPG file with a “Quality” setting of 75 to 80 (found from the “Save<br />

As…” dialog box under “Options…” if JPG is the file type) will provide an image<br />

without significant degradation and a file that is typically 1/10 the size of the original<br />

file. A series of these 3D images can be put into an album file (see a following<br />

section) and played as a slide show.<br />

To Overlay Text on the Image<br />

Procedure The image to be annotated should be displayed on the screen. The text<br />

tool should be activated by clicking on it (the “T” button next to the bottom of the lefthand<br />

tool bar). The bar above the image will show the font, its size, style, color, etc.<br />

By clicking on the active areas, it is possible to change the selection. To place text<br />

on the image:<br />

-With the text tool active, click on the image area where you would like the text<br />

to appear.<br />

-Type the text in the dialog box.<br />

-You may click on “Update” to see the text on the image or [OK].<br />

-If you would like to re-position the text, place the cursor on the text and drag<br />

the text to another location.<br />

-The font, size, color, etc. can be changed while the text is still active.<br />

-To deactivate the text, you may push the space bar or [Edit], [Object], and<br />

[Merge].<br />

You are able to type additional text entries by repeating the steps above<br />

To Overlay a Spectrum on the Image<br />

Procedure The image to be annotated should be displayed on the screen. The<br />

spectrum to be overlain should have been saved as a BMP file from the ZAF<br />

software and also recalled into PhotoImpact. Because the spectrum is a solid color<br />

(assuming that the spectrum “Page Setup” dialog box was not set to “Outline” mode)<br />

that is different from the spectrum background, it will be possible to cut and paste it<br />

into another image using what is commonly called the ‘magic wand’ tool. This tool is<br />

selected by first clicking on the third button from the top –it looks a little like a magic<br />

wand with an irregular dashed line around it. The bar above the image area will show<br />

some options for the magic wand tool and the most important may be the “similarity”<br />

number. It can probably any small number (


spectrum area is active. Next, the spectrum should be cut or copied to the clipboard<br />

and pasted into the image:<br />

-[Edit]<br />

-[Copy]<br />

Activate the image that you want to paste the spectrum into (i.e. click on it, or click on<br />

“Window” and select the image from the list at the bottom of the pull-down menu.<br />

Also, if you are going to paste the spectrum as a color image, then you should<br />

convert the gray-scale image to a true color image first; [Format], [Data Type], [True<br />

Color]. To copy the spectrum:<br />

-[Edit]<br />

-[Paste]<br />

Click and drag on the pasted object to position it where you would like it. When<br />

you are satisfied with its position, you should deactivate the text by pushing the space<br />

bar or [Edit], [Object], and [Merge].<br />

At this point you might want to use the text feature in PhotoImpact to add text to label<br />

the peaks. It would have been possible to copy the peak labels from the original<br />

spectrum file by using the magic wand tool to click on each letter, but this is tedious<br />

and time consuming. Also, when the spectrum is shown on the image, it has to stand<br />

out from the detail in the image. The file can be saved with the overlay.<br />

To Create Thumbnails of the Images and other Album Procedures<br />

Procedure<br />

-[File]<br />

-[New]<br />

A dialog box will appear that gives a choice of templates, but it will probably be best<br />

to use the “General Purpose” template. Type a title (e.g. “Maps”), then [OK]. From<br />

the “Insert” dialog box select the correct directory, then click, ctrl-click, shift-click or<br />

click and drag on the image files of interest, the [Insert]. You can chose an additional<br />

directory and repeat the insertion procedure from the previous sentence. To finish<br />

this session [Close].<br />

You can add additional thumbnails to an existing album file (or group of thumbnail<br />

images) by [Thumbnail] and [Insert]. The will bring up the same dialog box as in the<br />

previous paragraph.<br />

Other Album Procedures. If you double click on any thumbnail image it will open up<br />

the Viewer program and show you the image. The image can be zoomed by hitting<br />

the “+” or “-“ keys. The slide show program can be implemented by first selecting the<br />

thumbnails of interest (or Ctrl-A to select all), then [View], [Slide Show], then select<br />

the appropriate options and [Play]. The slide show can repeat continuously, or just<br />

play once and terminate, or it can be terminated with the escape key.<br />

<strong>EDAX</strong> Phoenix Training Course - PhotoImpact - page 5


<strong>EDAX</strong> Phoenix Training Course - PhotoImpact - page 6


The brightest part of the eye has a gray level of 182 and the darkest part of the eye has a<br />

gray level of 26. An image processing operation could be used to stretch or increase the<br />

contrast. If this were done, the range of gray levels could be increased to as much 0 to 255<br />

as shown at the bottom of the preceding page at the right.<br />

The previous image is an 8 bit, grayscale image. Other types of images can be only 1 bit<br />

(sometimes called a bitmap) which consists of black or white pixels, 256 color palette<br />

images (8 bits but it can contain a variable array of colors), 24 bit color images (sometimes<br />

called RGB or true-color images). The 24 bit color images contain 8 bits (256 levels) each<br />

of red, green and blue.<br />

Image files tend to be very large files. An image that is 1024x1024x8 bits is 1 megabyte in<br />

size, and 512x512x8 bits would be 250 kilobytes. If an image is stored as an RGB cololr<br />

image, it will take 3 times as much disk space. The <strong>EDAX</strong> digital image file sizes are 3.2<br />

megabytes (2048x1600x8 bits), 800 kilobytes (1024x800x8 bits), 200 kilobytes (512x400x8<br />

bits) and decrease by a factor of 4 for each smaller image size.<br />

<strong>EDAX</strong> Phoenix Training Course - Introduction to Digital Imaging - page 2


value of “1”. The standards file only contains a table of pure element intensities, it does not<br />

actually save a spectrum.<br />

PROCESSING THE UNKNOWN SPECTRA<br />

For the sake of consistency in this lab, click on “Results” and make sure that ”Intensity” and<br />

“%Conc” are both checked. Then click on “Options” and make sure that there are checks<br />

placed by “Methods”, “At%”, “K-Ratios”, “Wt%”, and “ZAF”, and that there is no check at this<br />

time next to “Multiple”. There are 7 spectra that were collected from an alloy consisting of<br />

aluminum (3.0 Wt%), chromium (38.5 Wt%), and nickel (58.5 Wt%). These spectra are labeled<br />

“alloy15a.spc” to ”alloy15d.spc”, and ”all15a.spc” to ”all15c.spc”.<br />

To quantify any of the alloy samples recall them from disk and click on “Quantify” from the<br />

Quant. control panel. The data can also be shown non-normalized by clicking on “Type”,<br />

“Option”, removing the check next to “Normalization”, “OK”, and clicking on “Quantify” again.<br />

At this point, each analysis is shown as a single table with the element, Wt%, At%, K-Ratio, Z,<br />

A, and F shown from left to right, and the elements shown from top to bottom by increasing<br />

energy. Another table will follow will give intensity data. If the goal is to accumulate multiple<br />

results (done by checking “Multiple” next to the “Quantify” button), then there will be several<br />

repetitions of these tables with the most recently acquired results at the bottom of the listing.<br />

It is possible to reformat the multiple results table by clicking on “Results” (near the top of the<br />

control panel), then on “Options”, and placing a check mark next to “Mult. Results”, then “OK”.<br />

Now, the multiple results table will be formatted with the elements shown from left to right and<br />

the sample number from top to bottom. There will be a series of tables for “Weight % by<br />

Element”, “Atomic % by Element”, “K-Ratio”, etc. If the results table is saved as a CSV file after<br />

multiple analyses have been performed, this file can be brought into MS Excel and additional<br />

mathematical operations can be performed (averages, standard deviations, and sums). Such a<br />

table has been copied to the clipboard and pasted into Word below.<br />

Given Weight % 3.0 38.5 58.5<br />

Weight % by Element<br />

Filename AlK CrK NiK Total<br />

ALL15A.SPC 3.05 37.82 58.04 98.91<br />

ALL15B.SPC 3.12 37.44 58.15 98.71<br />

ALL15C.SPC 3.40 37.96 57.97 99.33<br />

ALLOY15A.SPC 3.26 37.39 58.70 99.35<br />

ALLOY15B.SPC 3.07 38.24 57.90 99.21<br />

ALLOY15C.SPC 3.24 37.56 58.42 99.22<br />

ALLOY15D.SPC 3.18 38.86 57.21 99.25<br />

Mean 3.19 37.90 58.06 99.14<br />

Std. Dev. 0.11 0.52 0.47 0.22<br />

Saving a Compound Standard as a Pure Element Table<br />

<strong>EDAX</strong> Training Course --Analysis with Pure Element Standards - page 2


It is often desirable to use a compound standard together with several additional pure<br />

element standards. Of course, the same analytical conditions must apply to the pure element<br />

standards as well as to the compound standard (same analytical geometry, same beam<br />

current, voltage, etc.). In the example below, we will use one of the samples from the<br />

previous exercise (“alloy15a.spc”) as a compound standard. Just as in the case above, we<br />

should first verify that the parameters for our pure element intensity table is correct by<br />

clicking on “Stds” and “Pure”, then clicking on “Factors” and verify that the detector type,<br />

method and voltage are all correct.<br />

-the “Quantify” control panel should be made active if it is not already active.<br />

-[Stds]<br />

-[Options]<br />

-[Compound]<br />

-[Setup]<br />

-Type the percentage for each element followed by an<br />

“enter” (3.0, 38.5 and 58.5 for the Al, Cr, and Ni, respectively)<br />

-[RZAF] (calculated pure element intensities will be listed<br />

in the dialog box)<br />

-In this case, you should leave the “Use as compound”<br />

checkbox unchecked<br />

-[OK]<br />

-[Save]<br />

[Yes]<br />

-[OK]<br />

At this point, the calculated pure element intensities will be listed in the pure element table and<br />

the standards type will be listed as “pure”. This standards file could be saved as in the case<br />

above by clicking on “File” and “Save as…”, select the STD file type and give the file a name<br />

followed by an “Enter”. If another pure element standard were available (Mn or Fe for instance),<br />

it would be possible to add them to our table by following the procedure above ([Stds],<br />

[Options], [Pure], [Setup], [Save], [Yes], [OK]).<br />

Using this procedure it is possible to merge data from multiple pure element spectra and a pure<br />

element standard. It is even possible to use multiple compound standards. It may be a good<br />

ideal to take some notes of the pure element intensities calculated for each standard because it<br />

is likely that an element might be present in more than one standard. The table will record the<br />

last standard value that was saved and it might be best to use one of the previous values or an<br />

average of several standards. The pure element intensity values that can be viewed by clicking<br />

on “Factors” may also be edited to reflect your choice.<br />

<strong>EDAX</strong> Training Course --Analysis with Pure Element Standards - page 3


-[OK]<br />

-[Use as compound]<br />

-[OK]<br />

At the completion of this procedure, the compound standards button should be checked, and<br />

we are ready to re-call our spectra to be analyzed against this standard. You may wish to<br />

select “Multiple” results in order for you to more readily compare these data. Recall each<br />

spectrum ([File], [Open], specify “garnet02.spc”, [OK], and [No] to save the changes to the<br />

previous spectrum if you do not want to save it). At this point, you only have to [Quantify] and<br />

you will see your results. An additional spectrum can be recalled and the same procedure<br />

repeated. See Table 1 for a summary of the results for the analysis with standards as well as<br />

the standardless data.<br />

In order to use this compound standard in a subsequent session, you should save a<br />

quantitative methods file (QZF). Cick on “File”, “Save as…”, select the QZF file type and give<br />

the file a name followed by pressing the “Enter” key. There is a standards file type (STD) but<br />

this is reserved for the table of pure element intensities that result from the use of pure element<br />

standards.<br />

SEC FACTORS<br />

SEC factors are Standardless Element Coefficients. Even though it is said to be standardless,<br />

it is possible to use a standard to derive these factors and the resulting quantitative data are<br />

very nearly as good as data generated from analysis with standards. To maximize the quality<br />

of the results, the same accelerating voltage, the same matrix type (as close as possible)<br />

should be used, and the same set of elements should be present in standard and sample.<br />

Procedure. The “Quantify” control panel should be open for these steps and the sample to be<br />

used to develop the new set of SEC factors should be open (in our case, “garnet01.spc”).<br />

-[Stds}<br />

-[None] if this was not already checked.<br />

-[Options]<br />

-[Compound]<br />

-[Setup]<br />

-Type the percentage for each element followed by an “enter” (38.78,<br />

10.92, 17.08, 0.24, 15.35, and 17.64 for the O, Al, Si, Ca, Mn, and<br />

Fe, respectively)<br />

-[SEC]<br />

-[Update SEC Table]<br />

-[OK]<br />

-[OK]<br />

-[SEC]<br />

-[User]<br />

-[File] from the menu bar<br />

-[Save As…]<br />

-Change file type to .SEC and type “garnet” and hit the enter key<br />

(this will save the SEC factors for the next time similar samples<br />

are to analyzed)<br />

<strong>EDAX</strong> Training Course - Analysis with Compound Standards - page 2


At the completion of this procedure, these new SEC factors will be available for use. We are<br />

ready to re-call our spectra to be analyzed against these factors. You may wish to select<br />

“Multiple” results in order for you to more readily compare these data. Recall each spectrum<br />

([File], [Open], specify “garnet02.spc”, [OK], and [No] to save the changes to the previous<br />

spectrum if you do not want to save it). At this point, you only have to [Quantify] and you will<br />

see your results. An additional spectrum can be recalled and the same procedure repeated.<br />

See the SEC results at the bottom of Table 1. The error values are comparable to, and actually<br />

slightly better than the data for the analyses with the compound standard. It is expected,<br />

however, that the compound standard procedure should be a more robust procedure when we<br />

depart even slightly from the ideal conditions of this test because it is based more soundly upon<br />

a more rigorous, “first principles” approach.<br />

<strong>EDAX</strong> Training Course - Analysis with Compound Standards - page 3


Table 1. Garnet analyses for the same spectra when analyzed with no standards<br />

(“NoStd”), with a compound standard (“Stndrd”), and with SEC factors (“SEC”) derived<br />

from the same standard that was used for the compound standard.<br />

Garnet Analyses<br />

Carbon Coating Correction = 9<br />

NoStd Filename O Al Si Ca Mn Fe<br />

GARNET02.SPC 38.77 11.83 17.1 0.37 14.48 17.45<br />

GARNET03.SPC 38.39 12.00 17.36 0.30 14.51 17.44<br />

GARNET04.SPC 38.70 11.82 17.35 0.36 14.61 17.15<br />

GARNET05.SPC 38.40 11.89 17.40 0.35 14.50 17.46<br />

GARNET06.SPC 38.75 11.92 17.36 0.33 14.28 17.37<br />

Mean 38.602 11.892 17.314 0.342 14.476 17.374<br />

Std.Dev. 0.19 0.07 0.12 0.03 0.12 0.13<br />

%CV 0.49 0.59 0.69 8.77 0.83 0.75<br />

Reported 38.78 10.92 17.06 0.24 15.35 17.64<br />

%Error -0.46 8.90 1.49 42.50 -5.69 -1.51<br />

Stndrd Filename O Al Si Ca Mn Fe<br />

GARNET02.SPC 39.05 10.84 16.74 0.30 15.26 17.81<br />

GARNET03.SPC 39.07 10.88 16.91 0.30 15.19 17.65<br />

GARNET04.SPC 38.97 10.68 16.95 0.28 15.49 17.63<br />

GARNET05.SPC 38.74 10.73 16.97 0.35 15.34 17.87<br />

GARNET06.SPC 38.83 10.90 17.01 0.25 15.13 17.89<br />

Mean 38.93 10.81 16.92 0.30 15.28 17.77<br />

Std.Dev. 0.15 0.11 0.04 0.04 0.16 0.14<br />

%CV 0.38 1.01 0.25 14.20 1.06 0.78<br />

Reported 38.78 10.92 17.06 0.24 15.35 17.64<br />

%Error 0.39 -1.04 -0.84 23.33 -0.44 0.74<br />

SEC Filename O Al Si Ca Mn Fe<br />

GARNET02.SPC 38.95 10.85 16.77 0.30 15.29 17.84<br />

GARNET03.SPC 39.03 10.89 16.92 0.30 15.20 17.66<br />

GARNET04.SPC 38.95 10.68 16.95 0.28 15.50 17.63<br />

GARNET05.SPC 38.77 10.73 16.96 0.35 15.33 17.86<br />

GARNET06.SPC 38.87 10.89 17.00 0.25 15.12 17.88<br />

Mean 38.91 10.81 16.92 0.30 15.29 17.77<br />

Std.Dev. 0.11 0.11 0.03 0.04 0.17 0.13<br />

%CV 0.29 1.01 0.20 14.20 1.09 0.74<br />

Reported 38.78 10.92 17.06 0.24 15.35 17.64<br />

%Error 0.35 -1.03 -0.82 23.33 -0.40 0.76<br />

<strong>EDAX</strong> Training Course - Analysis with Compound Standards - page 4


Although this is commonly not an easy thing to do, the quantitative analysis actually<br />

does this comparison in its calculation (this is shown graphically in Fig. 2 below).<br />

Fig. 2. Same spectrum as in Fig. 1 (shown in solid black) compared to the outline of<br />

the pure element intensities for each element.<br />

The ratio of the element’s peak height in a sample to the pure element collected under<br />

the same conditions is actually used in the quantitative calculation as the k-ratio. All<br />

other steps in the quantitative analysis with and without standards are identical and the<br />

only difference is whether the pure elements are actually measured (with standards) or<br />

are calculated (standardless). Before the k-ratios are calculated, the peaks are<br />

identified, a background is fit, a peak-fitting routine is used and overlapped peaks are<br />

deconvoluted, and the net count intensities are measured. After the k-ratios are<br />

calculated, the ZAF corrections are applied, where: Z is an atomic number correction<br />

which takes into account differences in backscattered electron yield between the pure<br />

element and the sample (high atomic number pure elements will produce fewer x rays<br />

because some of the beam electrons leave the sample before losing all of their energy);<br />

A is the absorption correction which compensates for x rays generated in the sample but<br />

which are unable to escape when they are absorbed within the sample (low-energy x<br />

rays tend to be heavily absorbed); and F is the fluorescence correction which will correct<br />

for the generation of x rays by other x rays of higher energy. In order to calculate the<br />

ZAF factors, the composition must be known, but we need to know the ZAF factors to<br />

calculate the composition. This apparent contradiction is resolved by calculating ZAF<br />

<strong>EDAX</strong> Training Course - Quantitative Analysis - page 2


and the composition through a series of iterations until the last iteration makes no<br />

effective difference in the result.<br />

Table 1. Quantitative results for the spectra in Figs. 1 and 2.<br />

Element Wt % At % K-Ratio Z A F<br />

AlK 3.4 6.84 0.0159 1.0897 0.4277 1.0003<br />

CrK 37.96 39.6 0.3887 0.9866 0.9911 1.047<br />

NiK 57.97 53.56 0.5688 1.0014 0.9798 1<br />

Total 99.343 100<br />

The actual weight % is calculated by multiplying the k-ratio by 100% and dividing by the<br />

product of the Z, A and F factors (Table 1). The standardless results will always be<br />

normalized to 100%, while it is possible to provide a non-normalized result when<br />

standards are used.<br />

Standards can be pure element standards (one standard for each element) or a<br />

compound standard (consisting of multiple elements) can be used. The pure element<br />

standard will provide the pure element intensity directly. The spectrum from a<br />

compound standard has a known peak intensity and known compositions. The ZAF<br />

coefficients can be calculated from the known composition and the ZAF factors are<br />

applied in reverse (RZAF) to calculate the pure element intensities.<br />

The conditions used to collect the standard or standards and the unknown(s) must be<br />

constant. The detector-sample geometry must be the same, the accelerating voltage<br />

used must be the same, and the beam current should ideally be the same. The<br />

maintenance of the first two conditions as invariant is relatively simple. Slight variations<br />

of the beam current of a few percent will cause an inaccuracy of the final analysis when<br />

pure element standards are used. If a single compound standard is used and the<br />

results are normalized, the results will usually still be of good quality. The quality of the<br />

standardless analyses will not typically be affected by an unstable beam. When it is<br />

desirable or necessary to obtain non-normalized results, the beam current should be<br />

monitored with a specimen current meter and with a Faraday cup on the sample stage<br />

or in the microscope column. A Faraday cup can be constructed for the stage by drilling<br />

a hole into a carbon planchet or a sample stage and placing an aperture on top of the<br />

cavity. The depth of the cavity should be at least 6X the diameter of the aperture. If the<br />

beam is placed within the opening, then all of the primary beam is captured and<br />

conducted through the meter to ground and there will be no secondary or backscattered<br />

electrons leaving the sample.<br />

<strong>EDAX</strong> Training Course - Quantitative Analysis - page 3


Recalling Spectral Maps<br />

To open saved spectral maps for further analysis make the<br />

adjustments for the new analysis before opening the spectral map<br />

window:<br />

• To create new maps, update the element list<br />

• To create linescans, set the location marker to Loca-L.<br />

To recall spectrum map data without reprocessing the data:<br />

� Click on SpcMap in the menu bar.<br />

� Click on the FILENAME button.<br />

� Select one of the spectrum files (.SPZ) from the sample of interest.<br />

� Click on OPEN.<br />

� Click YES to load both the image and the spectra. This loads the<br />

original image and all of the spectra associated with it. The<br />

spectrum that appears on the screen with the image is the total of<br />

all spectra for that image.<br />

� To view the spectral maps, click on the SHOW MAP button.<br />

Creating Spectra<br />

� Load the stored data file of interest as shown above.<br />

� To build spectra from the image and hidden data activate the Add Up<br />

Spectra check box and choose Spectrum as the display type.<br />

� Click OK to exit the Spectral Mapping window. The sample image is<br />

shown with the sample spectrum. Clear the spectrum.<br />

� Select Loca-spot, line or matrix.<br />

� Draw on the image.<br />

• While in spot mode, click on a point. The spectrum for that pixel<br />

will be displayed. Clicking on another will add that spectrum to<br />

the first.<br />

• While in line mode, draw a line on the image. The spectrum that<br />

appears is a total of all the pixel spectra.<br />

• While in matrix mode, draw a box. The spectrum that appears is the<br />

sum of all the pixels included in that box.<br />

The operations for spectra act as they normally would. A spectrum<br />

can be erased and another created without reloading the spectral<br />

mapping file. This option is useful when overlaying spectra from<br />

different regions.<br />

Creating Linescans from Recalled Data<br />

� Open the stored data as shown in section 3.4 of this manual.<br />

� To build linescan from the image choose Line Scan as the display<br />

type.<br />

� Click OK to exit the Spectral Mapping window. The sample image is<br />

shown with the sample spectrum.<br />

� Define the location marker as line (Loca-L).<br />

� Draw a line on the on the image. The Line Scan Setup window will<br />

open.<br />

� Click on DISPLAY.<br />

Creating Quantitative Maps<br />

Once Spectral Data has been collected it is possible to create<br />

quantitative maps. A ZAF correction is performed on each of the<br />

spectra that were saved with the image.<br />

� If the original element list is not the list of elements desired for<br />

Quantitative maps (Elements need to be added or taken away.) adjust<br />

the element list now.<br />

� To create Quant Maps, open the Spectral Mapping file as shown in<br />

Section 3.4 of the manual.<br />

<strong>EDAX</strong> Basic Procedures- Spectrum Mapping 2


� Open the SpcMap window.<br />

� Click on the BUILD QMAP button.<br />

� To view the maps click on SHOW MAP. Select a map from the file.<br />

The Quant Maps are identified with a “Q” in their filename.<br />

� Click on the Color check box to change to map to a thermal color<br />

scale.<br />

� When clicking on a pixel in an element’s quant map the calculated<br />

concentration for that element at that pixel is displayed in the<br />

information area below the map.<br />

� For further viewing see Section 3.2 – Displaying Maps.<br />

Print<br />

There are three options for printing Spectral Maps, single map,<br />

image and map, and multi-color maps. Single X-Ray Map prints the<br />

active map in the left corner of the window. Image and Map prints<br />

the active map and the image. Multi-Color-Map prints all the maps<br />

highlighted.<br />

<strong>EDAX</strong> Basic Procedures- Spectrum Mapping 3


� You have a valid element list for the points of your analysis (see<br />

Peak ID)<br />

� You have a preset time for spectrum acquisition (listed under<br />

“Setup: Preset & Memory”)<br />

� Your “output” options are matched to your expectations and that you<br />

have specified three letters for a prefix for the spectra to be<br />

stored (under “Setup: Auto Output”).<br />

At this point you can click on “Auto” and “Start” and the procedure<br />

will begin. When it is finished (you will hear music) and the<br />

results can be examined by clicking on “Auto” and “Summary” and<br />

selecting your results file.<br />

Linescan (option)<br />

To start a linescan:<br />

� Make sure that you have a correct element list<br />

� Click on “Loca - “ and select “Line”.<br />

� Click and drag a line on a current image (you do not have to press<br />

“S” to save it).<br />

� Click on “Image” and “LScan conditions...”.<br />

� In this dialog box:<br />

� Click on “Read position”<br />

� Make sure that the number of points and the dwell time per point are<br />

suitable<br />

� Click on “Start Lscan”.<br />

You will know when the data acquisition is finished (again, you will<br />

hear a beep). From the same dialog box, click on “Display” and you<br />

will see the line scans for each element.<br />

In the menu bar and pull-down menus there are choices to select a<br />

single element, to overlay it on the image, etc.<br />

� To save the linescan, close the linescan dialog boxes by clicking on<br />

“OK”, etc., and click on “File”, “Save As...” and select linescan<br />

from the file type list box and give it a name and save the file.<br />

You may also wish to save the image as a BMP file if you have some<br />

graphics that are of interest.<br />

Print<br />

� “File: Print Output” opens a dialog box where the choice for<br />

printing the image and spectrum are available.<br />

� The spectrum can be printed as an outline or in solid.<br />

� The half page box should be checked when printing the spectrum<br />

and image. If only one is being printed then the full page can<br />

be selected.<br />

� To print click on the Print button.<br />

� To exit the dialog box without printing, click on OK.<br />

<strong>EDAX</strong> Basic Procedures- Multi Point Analysis 2


Mapping (standard, live, quantmaps)<br />

� The standard type of x-ray mapping requires a specified dwell time<br />

for each pixel and collects gross intensity information from regions<br />

of interest. Other options available include Live (this collects a<br />

number of very fast frames and allows the user to terminate the<br />

mapping session after any frame) and Quantitative maps (allows for<br />

the collection of de-convoluted and ZAF corrected maps).<br />

� To collect maps, the area of the sample to be mapped should be<br />

displayed on the microscope screen, an image collected and all<br />

elements should be identified (see above). A suitable count rate<br />

should be established with the optimum time constant (this will vary<br />

with the sample type, magnification selected, and if there are peak<br />

overlaps which will require the best detector resolution).<br />

� Open the ROI control panel and click on the “Auto” button. The<br />

ROI control panel icon is the second button in the cluster of<br />

panel buttons, next to the Peak ID panel button.<br />

� Open the MAP control panel:<br />

� For standard maps select a matrix and a dwell time that are<br />

consistent with the time available (Table 1), spatial<br />

resolution (Table 2) needed and signal-to-noise required.<br />

� For Live maps select a matrix size and check the check box<br />

next to “live” in the MAP panel. Be sure the number of<br />

frames is set to 1024 or so.<br />

� For Quantitative maps check the check box next to “quant”<br />

in the MAP panel. Select a matrix and dwell time that are<br />

consistent with the time available, spatial resolution<br />

(Table 2) needed and signal-to-noise required.<br />

The estimated time for map collection will be displayed in the<br />

status bar at the bottom of the screen. Use this when determining<br />

matrix and dwell times. Note* For live maps there is no estimated<br />

time as the maps will collect until the user stops the collection.<br />

� Click on the X button to begin the map collection.<br />

� Provide a prefix for the images to be stored, then click on OK.<br />

� Confirm the parameters for the map collection and click on OK.<br />

When the maps finish collecting they will automatically be saved.<br />

To recall maps use “File: Open”, choose to open an image group, and<br />

double click on any one of the maps saved with that filename. All<br />

maps and images with that base name will open.<br />

Table 1. Length of time to complete an x-ray map in minutes or<br />

hours (shown in underlined text).<br />

Dwell Times (ms)<br />

1 5 10 50 100<br />

Pixel Size<br />

64x50 0.1 0.3 0.6 2.7 5.4<br />

128x100 0.3 1.2 2.2 10.8 21.4<br />

256x200 1.2 4.6 8.9 43.0 1.4<br />

512x400 4.7 18.4 35.5 2.9 5.7<br />

1024x800 18.9 1.2 2.4 11.5 22.8<br />

2048x1600 1.3 4.9 9.5 45.9 91.4<br />

<strong>EDAX</strong> Basic Procedures- Imaging/Mapping 2


Table 2. Pixel size in micrometers at different magnifications and<br />

different horizontal image size (pixels).<br />

100x<br />

Magnification<br />

1000X 10,000X<br />

# of Pixels<br />

64 18.6 1.86 0.19<br />

128 9.3 0.9 0.09<br />

256 4.6 0.5 0.05<br />

512 2.3 0.2 0.02<br />

1024 1.2 0.1 0.01<br />

2048 0.6 0.06 0.006<br />

Table 3. Time to capture 512 x 400 image. **<br />

Int. F=1 Int. F=100 Int. F=500<br />

1 strip 2.67 sec 4.46 sec 12.74 sec<br />

10 strips 2.98 sec<br />

50 strips 6.7 sec<br />

** Time for 1024 x 800 image would be roughly 4 X these times.<br />

Overlay Maps<br />

This function creates an overlay image from up to six selected<br />

images and displays the result in the first empty window available.<br />

The element with the highest relative concentration in its map will<br />

be displayed at each pixel (substitution overlay).<br />

� Select View_16 screen mode from the View menu. (The maps must be in<br />

color for the overlays to make any sense. If they are not in color,<br />

Select “Color Palette” from the “Display” menu.)<br />

� Click on the images you want to overlay while holding the Ctrl Key<br />

pressed.<br />

� Select “Substitution Overlay” from the “Process: Color” menu.<br />

Reverse Palette<br />

The black background of maps can be swapped to white by selecting<br />

“Reverse Palette” from the “Display” pull-down menu. This is done<br />

mainly to save ink when printing. Maps should not be initially<br />

collected in this display mode.<br />

Zoom and Measure on Image<br />

To enable the zoom operation you have to be in View_4 or View_16<br />

screen mode.<br />

� Click the button (with a magnifying glass) from the toolbar or<br />

select “Zoom Enabled” from the “Display” menu.<br />

� Click and drag on the image to define the area to be zoomed. (You<br />

have to drag to the right and down to enable the zoom operation).<br />

� A popup window will show the zoomed area. You can move the zoom<br />

window around by clicking and dragging it to a new location on your<br />

screen.<br />

� To dispose of the zoom window double click on it.<br />

� To copy the zoomed area to the clipboard select “Edit: Zoom to<br />

Clipboard” from the “Edit” menu. From the clipboard you can paste it<br />

in an MS Word document or in a selected image view. To place the<br />

zoomed image into an image view, click in a free image view and<br />

select “Paste” from the “Edit” menu. The new image is available now<br />

for all image view operations (File, Print, Edit, Process or Zoom).<br />

<strong>EDAX</strong> Basic Procedures- Imaging/Mapping 3


� The line measurement tool button lies to the right of the zoom<br />

button. Clicking on the line measurement tool will move the<br />

selected image to the 1_view mode and the cursor will turn into<br />

crosshairs.<br />

� Click on the first point of the line to be measured and drag to the<br />

end point. The measurement will be given in blue text in the status<br />

bar at the bottom of the screen.<br />

� The measuring unit is micron.<br />

� To measure another line simply click on the measurement button<br />

again.<br />

Annotating the Image<br />

To display a label, micron marker, accelerating voltage, or<br />

magnification on the image, select “Annotate Image” from the “Edit”<br />

pull-down menu. The mag., kV, and label should reflect current<br />

parameters. By checking on and off appropriate boxes and clicking<br />

on the “Paste” button the chosen information will appear along the<br />

bottom of the image or map.<br />

� Selecting a font size, font type, and color (black, white, white<br />

on black, or black on white) are helpful features that emphasize<br />

the text annotation.<br />

� Activating the spot mode or crosshair allows you to position the<br />

text anywhere you would like before pasting it onto the image.<br />

To display a line and measurement:<br />

1. Open the “Annotate Image” dialog box open and click on the<br />

measurement button.<br />

2. Draw the line to be measured as instructed above. The length<br />

will be displayed in the label section of the Annotate Image<br />

dialog box.<br />

3. Check the line check box and click on “Paste”.<br />

4. Check off the line check box and check on the label check box.<br />

5. Turn on the spot mode (cross hair) and position the red crosshair<br />

where the measurement text should appear.<br />

6. Click on “Paste”. Use the “Undo” button to undo the last<br />

step.<br />

Any annotations will not be saved to the image unless the image is<br />

saved again.<br />

Printing<br />

In most instances, the spectrum will have been collected in spot<br />

mode and the red crosshair should still be shown on the image in the<br />

correct location.<br />

� “File: Page Setup” Ensure that both the image and spectrum check<br />

boxes have an "x" in them and the other parameters are as you want<br />

them to appear. Click on OK.<br />

� To print the spectrum and image choose Print while in the Spectrum<br />

view.<br />

� To print maps, select the maps while holding down the Ctrl key. The<br />

different views will print the maps in different sizes, fitting 1, 4<br />

or 16 maps to an 8”x11” page. Then select “File: Print” or click on<br />

the printer button.<br />

� To print the spectrum, image, and quantification results on a single<br />

page, have the quantification data showing (by clicking on the “Q”<br />

button), then click on the printer button or select “Print” from the<br />

“File” pull-down menu.<br />

<strong>EDAX</strong> Basic Procedures- Imaging/Mapping 4


(1) After this menu item has been selected, the cursor changes<br />

to an I-beam.<br />

(2) Click on the spectrum at the point where the text is to be<br />

inserted. An edit box appears and the comment is typed<br />

directly into the box. The text will scroll if not all text<br />

can be displayed in the edit box.<br />

(3) Once all text has been entered, press Enter to insert the<br />

text into the spectrum. Up to 32 characters can be entered<br />

for each text item and up to 10 comments can be added to a<br />

spectrum.<br />

� A comment can be changed or deleted from the spectrum by choosing<br />

the Select Text menu item, then clicking on the existing comment<br />

in the spectral window. The text will be displayed in the edit<br />

box and can then be changed. To completely delete the comment,<br />

delete all text in the box and press Enter.<br />

Peak ID<br />

� You should select the Peak ID control panel by clicking on the<br />

correct button (the left-most button out of the right-most group;<br />

the one that shows a spectrum with a “?”).<br />

� You can then click on “Auto” for an auto-peak ID (the same can be<br />

done by clicking on the “ID” button on the button bar). You may<br />

need to modify some of peaks identified. To remove an element,<br />

select the element from the “Saved Elem” list box and click the<br />

“Delete” button.<br />

� To add an element manually:<br />

(1) Type the four characters in “Element” box (e.g. SiKa,<br />

AgLa, or K Ka, etc.).<br />

(2) Pick an element that is near the element you want to<br />

add from the “Saved Elem” list and click “Z-” or “Z+”<br />

until the element of interest appears in the “Element”<br />

box, then click “Add”.<br />

(3) Click the cursor on the peak that you want to identify<br />

in the spectrum, then click on the choice from “Pos<br />

Elem” box and click “Add”.<br />

� After performing the peak identification procedure, there are<br />

commonly some vertical lines on the spectrum that indicate the<br />

position of the last peak or element requested during the peak ID<br />

step. If the spectrum is to be printed, these lines will sometimes<br />

print with the spectrum (in older versions of the software) and this<br />

is rarely the desired result. To remove the line(s), first select<br />

the “Peak ID” control panel, then click on “View” and “Element<br />

Marker”. A new method to remove these markers is to do a right or<br />

left mouse click in the spectrum area.<br />

HPD<br />

� This button (found in the Peak Identification panel) is used to<br />

calculate and display the background and deconvolution without<br />

displaying any quantification results. The user need not go to the<br />

Quantification Control Panel to calculate the background and<br />

deconvolution. The HPD (in light blue) relies on the element list<br />

and correct kV when modeling.<br />

� HPD is used for confirming the present element list. A user can<br />

click on the HPD button to see what the spectrum should look like<br />

with the elements identified. If the HPD does not match the<br />

spectrum then adjustments can be made to the element list and the<br />

HPD button clicked again. These steps can be repeated until the HPD<br />

<strong>EDAX</strong> Basic Procedures- SEM Quant ZAF 2


matches the spectrum and all elements have confidently been<br />

identified.<br />

� In the instance when two peaks overlap the HPD is most useful. A<br />

common example of this is the overlap of the Sulfur K series peaks<br />

and the Molybdenum L series or the Titanium L peaks and the Nitrogen<br />

K peak.<br />

� A spectrum, “Stnlss.spc”, provided in the D:Edax32:<strong>EDS</strong>:user<br />

directory has overlapping peaks of Fe, Cr, and Mn. Using the HPD<br />

with only Fe and Cr identified will show a mismatch where Mn should<br />

be. Adding Mn to the list and clicking on the HPD button then gives<br />

a good deconvolution model.<br />

Standardless Quantification<br />

� Before quantitative results can be reasonably accurate, the<br />

collection parameters must be correctly specified (oxides or<br />

elements, etc.) and the analysis should satisfy three fundamental<br />

assumptions of ZAF (Z - atomic number, A - absorption, and F -<br />

fluorescence) corrections:<br />

(1) the sample should be smooth and flat<br />

(2) the sample (as seen by the beam electrons) should be<br />

homogeneous<br />

(3) the sample should be infinitely thick to the<br />

electron beam.<br />

� If the sample is not smooth, then the geometry will be wrong and the<br />

take-off angle incorrect. If the sample contains different phases<br />

within the field of view examined, then the absorption corrections<br />

will not be correct. If the sample contains a thin film or films,<br />

then the results will differ according to the voltage selected with<br />

different penetration levels and a different proportion of substrate<br />

to the film will be sampled.<br />

� If the assumptions are reasonably well satisfied, then the<br />

quantitative result (even in standardless mode) will be reasonably<br />

accurate. (To test the degree of accuracy, standards can and should<br />

be used that match the types of samples you will analyze. Analyzing<br />

the standard as an unknown in standardless mode will provide data<br />

that you can compare to its published values.)<br />

� To quantify a spectrum select the quantify control panel using the<br />

button that looks like a balance weighing out “Fe” and “Cr”,<br />

followed by the “Quantify” button, or by clicking one of the middle<br />

buttons (the one between “ID” and “AUTO”) in the center of the main<br />

button bar.<br />

� When the multiple results box is checked (in the Quant panel) new<br />

results are appended to the previous quantification thus providing<br />

one print out or file with results from many spectra. **The results<br />

from the last spectrum quantified are displayed at the end of the<br />

results.<br />

Saving Quantification Data<br />

� To save quantification results click on the “Save” button in the<br />

Quantification panel, then follow standard Windows procedures for<br />

saving a file.<br />

� The results are saved as .CSV (comma separated values) files<br />

which can be opened in any spread sheet program, typically<br />

Microsoft Excel.<br />

Printing<br />

� To print the spectrum, first check the “Page Setup” dialog box found<br />

under the “File” pull-down menu. There are several choices here for<br />

how the spectrum and/or its quantitative results will appear.<br />

<strong>EDAX</strong> Basic Procedures- SEM Quant ZAF 3


� Common choices are to select black and white (if using a Laser<br />

Printer or if your color cartridge is bad) or to select the<br />

spectrum print out in outline mode --this will use less ink of<br />

one specific color and will probably mean that the printout will<br />

photocopy better.<br />

� Other options allow you to print overlain spectra and<br />

quantitative results.<br />

� When the multiple results box is checked in the “Page Setup”<br />

dialog box, the results from a series of spectra print organized<br />

in order of element and parameter as opposed to in order of<br />

sample.<br />

� The printing of the spectrum is accomplished by clicking on “File”<br />

and “Print” as is normally required in a Windows application.<br />

Spectrum Overlay<br />

� Under “View” on the menu bar you will find several choices on the<br />

pull-down menu. The procedure used most often is the overlaying of<br />

one spectrum on another for purposes of comparison.<br />

� To overlay a spectrum onto another one, click on “View”,<br />

“Compare...”, then check the “Overlay” box and then “Open” a<br />

spectrum from disk to be used as memory B for the overlay. (This<br />

requires that you save the spectrum first before it can be used for<br />

the overlay.) When the spectra are being shown in overlay mode,<br />

clicking the overlay off and back on again with the overlay button<br />

on the button bar (the one to the right of the Home/House button)<br />

becomes available.<br />

Calibration<br />

Select the calibration control panel by clicking on the button from<br />

the right-most group that shows a spectrum with a caliper measuring<br />

a peak (third from the right usually). By default, this control<br />

panel assumes that you have aluminum and copper in the field of<br />

view. Typical SEM parameters are 25 kV and a spot size to assure a<br />

30-35% deadtime for a given time constant, and a field of view with<br />

approximately 7/8 of it occupied by copper (1/8 being aluminum).<br />

This should give a spectrum with the Al Ka peak (1.486 keV) and the<br />

Cu Ka peak (8.04 keV) about the same height. If they are not equal<br />

then you should move the stage slightly until the peaks are about<br />

equal. Typically, the maximum number of counts collected should be<br />

at least 5000. Clicking on “Start” will begin the calibration<br />

procedure. The time constant or pulse processing time can be<br />

changed in the EDAM control panel and the calibration procedure will<br />

need to be run for each time constant.<br />

<strong>EDAX</strong> Basic Procedures- SEM Quant ZAF 4


.<br />

<strong>EDAX</strong> Training Course - X-Ray Count Optimization - page 2


Decon02.spc. Between 0.5 and 15 keV, the peaks are:<br />

RhL, PdL, AgL.<br />

SpecUtil Printouts of the K-, L- and M-Series Spectra<br />

There are 10 spectra for the K-series peaks, 10 for the L, and 7 for the M-series peaks. The<br />

peak of interest is the largest peak in all cases or the peak with a dot over the peak. You will<br />

notice that KS08 and LS02 are both the same spectra (Fe).<br />

K-Series Spectra. At energies below about 2.5 keV, the alpha and beta peaks are not<br />

resolved or not present. The distance between the two peaks becomes greater at higher<br />

atomic numbers.<br />

L-Series Spectra. At energies greater than about 3 keV, the L-series peaks consists of 6<br />

(or 7) peaks depending on the peak position and peak intensity: Ll, (Ln), La, Lb1, Lb2, Lg1,<br />

and Lg2 (from lower to higher energy). At energies below 3 keV there are typically only two<br />

peaks that are resolved: the Ll and the La (the other peaks may be present but not resolved,<br />

and may make the La appear asymmetric). At low energies (below 1 keV), the Ll becomes<br />

larger, and at very low energies (below about 0.6 keV) the Ll becomes larger than the La.<br />

M-Series Spectra. At first glance, the M-series spectra are similar to the L-series spectra. At<br />

energies below 3 keV, there are only two significant peaks visible: the Mz and the Ma. It<br />

might be possible to also see a Mg and a M2N4 peak but these are very small. At low<br />

energies the Mz becomes larger and becomes the dominant peak at very low energies.<br />

Above an energy of about 3.0 keV, it becomes possible to resolve additional peaks, but at<br />

this point we are at the highest atomic number portion of the periodic chart. There are no Ma<br />

peaks above 4 keV, and between 3 and 4 keV the only Ma peaks are for radioactive<br />

elements such as Th, Pu and U.<br />

Number K-Series L-Series M-Series<br />

01 Be Ti<br />

02 C Fe W<br />

03 Al Ni Re<br />

04 Si Zn Os<br />

05 Sc Ru Pt<br />

06 V Pd Au<br />

07 Cr Ag Bi<br />

08 Fe Sn Th<br />

09 Ni Sb<br />

10 Zn Te<br />

<strong>EDAX</strong> Quiz on Peak ID - page 2


PkID11.spc (easy). Between 1 and 10 keV, the peaks are:<br />

PkID12.spc (easy). Between 1 and 10 keV, the peaks are:<br />

Decon01.spc (very hard, maybe impossible). Between 1 and 10 keV, the peaks are (or<br />

might be):<br />

Decon01b.spc (very hard, maybe impossible). Between 1 and 10 keV, the peaks are (or<br />

might be):<br />

Decon02.spc (moderate to hard). Between 0.5 and 15 keV, the peaks are:<br />

Peak ID Quiz -- page 2


Characteristics of the K - Series Peaks<br />

K-series peaks are typically fairly simple and consist of a Ka and a Kb peak. The Kb peak<br />

occurs at a higher energy and starts to become resolvable from the Ka for atomic numbers<br />

greater than sulfur (Z = 16). The Kb peak is usually 1/8 to 1/6 the size of the Ka peak as can<br />

be seen in the spectrum below for titanium.<br />

Characteristics of the L - Series Peaks<br />

The L-series peaks are much more complex than the K-series peak(s), and may consist of<br />

as many as 6 peaks: an La, Lb1, Lb2, Lg1, Lg2 and Ll. The peaks were arranged (in the<br />

latter sentence) from highest energy to lowest energy with the exception of the Ll peak which<br />

occurs before (to the left of, or at a lower energy than) the La peak (see spectrum for tin on<br />

the following page). Aside from the Ll peak which is a very small peak, the peaks are<br />

progressively smaller with energy. For L-series peaks from elements less than the atomic<br />

number of silver (Z=47), this peak series is not resolvable. One of the common mistakes in<br />

peak identification is to be unaware of the Ll peak and to hypothesize the existence of a<br />

minor or trace amount of an element that does not exist in the sample.<br />

<strong>EDAX</strong> Phoenix Training Course - Peak ID - page 2


Characteristics of the M - Series Peaks<br />

Most M-series peaks will show only two resolvable peaks, the Ma peak which also includes<br />

the Mb (not resolved) and the Mz peak. Just as with the Ll line in the L-series peaks, the Mz<br />

occurs at a lower energy than the Ma peak and there have been many instances of<br />

identifying another element by the analyst who did not recognize the Mz peak, or who was<br />

using a peak identification method that did not include the Mz peaks. For instance (see<br />

following page), the Mz for osmium is perfect for being a small amount of Al or Br. The Mz<br />

for gold is a good fit for the hafnium Ma peak, etc. M series peaks tend to be markedly<br />

asymmetric due to the mixing together of the Ma and Mb (the Mb being a smaller peak at a<br />

higher energy). To perform a manual peak ID for the M series peak, it is often necessary to<br />

click the mouse cursor slightly to the left of the peak “centroid” because of this. When the<br />

peak for the Ma is selected, the lines for both the Ma and Mb are shown. There may also be<br />

an Mg peak which may show up as low, high-energy shoulder for the large M peak, and a<br />

very small “M2N4” peak (see spectrum on the following page)<br />

Auto Peak ID<br />

Auto peak ID is fast, easy and will be reasonably accurate with K-series peaks that do not<br />

have serious overlaps. Such an overlap would consist of a Ka from a trace element that<br />

happens to be near a major peaks Kb (i.e. a trace of Mn with Cr, or V with Ti, etc.). When<br />

the overlap consists of M, L and K series peaks, the auto peak ID routine is not well suited to<br />

<strong>EDAX</strong> Phoenix Training Course - Peak ID - page 3


get all the elements correctly. In this case, the use of the peak deconvolution software is<br />

recommended (see following discussion).<br />

The auto peak ID routine makes use of an “omit” list, which is a list of elements that the auto<br />

peak ID will not find (see figure below). This list can be reached from the Peak ID control<br />

panel by clicking on “EPIC” (Edax Peak Identification Chart), the “Auto ID”. It is possible to<br />

remove elements from this list or to add elements. <strong>Manual</strong> ID will still list an element that is<br />

in this omit list.<br />

<strong>EDAX</strong> Phoenix Training Course - Peak ID - page 4


Deconvolution and Peak ID<br />

Deconvolution is used in the quantification routine to estimate the size and position of all<br />

peaks associated with each assigned element. The fit of all the peaks calculated is shown<br />

as a lined spectrum on top of the background, so that any serious gaps or misfits can be<br />

observed by the user. The deconvolved spectrum can be observed by clicking on the<br />

“Quant” button in the button bar and “OK” to remove the quantitative results from the display.<br />

It may be a good idea that your “results” mode should not be in “Multiple” mode for this<br />

action, since this may require several iterations and the intermediate results are seldom of<br />

interest.<br />

If the spectrum has more counts as a cluster than the deconvolved spectrum, then the<br />

manual peak ID can be used to find a peak that can account for those peaks. In the case<br />

below, there are some counts that are not accounted for by the deconvolution in the area of<br />

the Cr Kb. This is common in a stainless steel and indicates the presence of the Mn Ka<br />

peak. It is also possible to subtract the deconvolution from the spectrum (click on “Proc” and<br />

“Subtract deconvolution”). This will make it more clear where a peak’s centroid should be.<br />

To undo the subtraction, click on “Edit” and “Undo”.<br />

The peak-fitting routine used is most accurate for the K-series peaks. The complexity of the<br />

L-series peaks does not permit as accurate of a fit, but in most cases this fit will be more<br />

than adequate to identify the peaks.<br />

<strong>EDAX</strong> Phoenix Training Course - Peak ID - page 5


“overvoltage”. Typically, it is thought that the overvoltage should be two or greater for<br />

<strong>EDS</strong> analysis.<br />

X-Ray Absorption<br />

A consideration of the depth of excitation for low-energy x rays would lead us to believe<br />

that the low energy x rays may be produced nearly within the entire depth of penetration<br />

of the electron beam. In actuality, even though this may be true for very low energy x<br />

rays, these x rays are also very easily absorbed by the sample and relatively few of<br />

them may actually escape from the sample except for those that were actually<br />

generated near the sample surface. The loss of these absorbed electrons must be<br />

corrected for in the quantitative analysis. The ratio of absorbed to emitted x rays<br />

increases with the accelerating voltage. As a result, absorption considerations also<br />

places an upper limit on our overvoltage --no more than 20 for qualitative work and<br />

ideally less than 10 for quantitative analysis.<br />

X-Ray Fluorescence<br />

Characteristic x rays can be produced by other x rays as well as by high energy<br />

electrons. This is typically referred to as x-ray-induced fluorescence or x-ray<br />

fluorescence. For instance, nickel K-alpha x rays have an energy near the critical<br />

ionization energy of iron K-alpha and will readily fluoresce iron x rays. This leads to an<br />

increase in the iron peak in the spectrum and a decrease in the nickel peak beyond<br />

what would be expected given the abundance of the these two elements. The<br />

fluorescence correction would correct for this effect by effectively removing some of the<br />

iron x-ray counts and placing them with the nickel x-ray counts.<br />

Detector Efficiency<br />

The efficiency of the <strong>EDS</strong> detector is controlled by the window type (if any), the gold<br />

contact layer and the silicon dead layer. The super-ultra thin windows (SUTW) will<br />

typically permit the detection of beryllium and boron, but not as efficiently as higher<br />

atomic number elements. The beryllium window is relatively good for K-radiation of<br />

elements with atomic number greater than silicon (Z=14) but drops to what is basically<br />

zero for elements less than sodium (Z=11) in atomic number. Although not much of an<br />

issue in the SEM, there can also be inefficient detection for very high energy x rays<br />

because these x rays may pass through the entire detector thickness and create no<br />

detectable signal.<br />

X-Ray Artifacts<br />

Peak Broadening. The resolution of the EDX detector is typically specified as the<br />

FWHM at manganese (Z = 25) and is on the order of 140 eV. There is a predictable<br />

relationship that peaks of a lower atomic number will have a lesser FWHM (full width at<br />

the half maximum height of a peak) than peaks at higher atomic numbers. When we<br />

calibrate the spectrometer, the Al Kα and Cu Kα are used but the resolution at Mn Kα is<br />

calculated or interpolated. The Al peak will typically have a much smaller FWHM than<br />

the Cu peak.<br />

<strong>EDAX</strong> Phoenix Training Course - Spectrum Interpretation & Artifacts - page 2


Peak Distortion / Asymmetry. Peaks in the spectrum will typically show an asymmetry in<br />

which the peak is usually sharper at the high end and this is partly due to the absorption<br />

edge of the element. The tailing on the low-energy side of the peak is due in part to<br />

incomplete charge collection (not all electron-hole pairs are collected –this is thought to<br />

be a problem for x rays that strike the detector near its edges).<br />

Escape Peaks. Occasionally an x ray striking the detector will cause the fluorescence of<br />

a silicon x ray. This may result in two x rays, one with the energy of silicon (1.74 keV)<br />

and one with the original x-ray energy minus the silicon energy. If both x rays remain in<br />

the detector, the two x rays are summed and the correct energy assigned in the multichannel<br />

analyzer. If the silicon x ray escapes from the detector, then what remains in<br />

the detector is called an escape peak which has an energy that is 1.74 keV less than the<br />

original x ray. Only x rays with energies greater than the absorption edge of silicon<br />

(1.84 keV) can cause the fluorescence of silicon, so we can expect to see the escape<br />

peaks associated with K radiation for phosphorus and up in atomic number. The size of<br />

the escape peak relative to its parent peak (typically no more than a percent or two)<br />

actually diminishes at higher atomic numbers as a result of the higher energy x rays<br />

<strong>EDAX</strong> Phoenix Training Course - Spectrum Interpretation & Artifacts - page 3


tend to deposit their energy deeper in the detector where the silicon x rays (if generated)<br />

have more difficulty escaping from the detector.<br />

Absorption Edges. The background of our spectrum will typically show a drop or an<br />

edge at the energy associated with the absorption edge of silicon and may show<br />

another near the absorption edge of the gold M line. These result from the passage<br />

through the gold layer and silicon dead layer with a concomitant absorption of the x-ray<br />

continuum.<br />

Silicon Internal Fluorescence Peak. It is possible that the silicon dead layer of our<br />

detector may also be fluoresced by an incoming x ray and the resultant silicon x ray may<br />

then travel towards the detector and produce a very small, but recognizable peak. It has<br />

been estimated that this peak can be as high as 0.2 weight percent of some samples.<br />

This is difficult to verify or confirm, because just having the background correction not<br />

quite correctly specified, could easily produce a silicon content this high. Also, given the<br />

poor quality of such a small peak, it could be easily confused with the silicon absorption<br />

edge.<br />

Sum Peaks. The sampling interval during which x rays are detected is a few tens<br />

nanoseconds. It is possible that two x rays of the same energy will enter the detector<br />

within this time period and be counted as a single x ray of twice the energy.<br />

Compounds can give rise to sum peaks that are the sum of two different elemental<br />

energies. Aluminum alloys will commonly have sum peaks (2.98 keV) that are equal to<br />

the position of other elements, such as argon or silver. Sum peaks are typically<br />

problems at higher count rates and when a phase is dominated by a single element. To<br />

determine if a small peak may be a sum peak (i.e. a detection artifact, not<br />

representative of the composition of the sample), you can select sum peaks from the<br />

Peak ID control panel and it will show a marker for the most probable sum peak. To<br />

confirm that the peak in question is really a sum peak, save the peak and use this peak<br />

as an overlay. Then, lower the count rate (perhaps to ¼ of what it was) and collect a<br />

new spectrum. If the major peak is of a comparable size in both spectra and the signalto-noise<br />

appears similar in both cases, but the peak in question is no longer visible or<br />

greatly diminished, then the peak in the first spectrum was a sum peak. If the peak is<br />

undiminished, then it represents a real part of the sample.<br />

Stray Radiation. Stray x rays are x rays that originate anywhere other than where the<br />

primary beam interacts with the specimen and may be produced by a variety of<br />

processes. Most occur as a result of x rays created by high-energy backscattered<br />

electrons striking the pole piece, stage, chamber or another part of the sample outside<br />

of the imaged area. While it is possible for these x rays to enter the detector, careful<br />

collimation of the detector can eliminate most of this problem. Probably more serious<br />

are backscattered electrons that strike the pole piece or stage and are exchanged for<br />

another backscattered electron that returns to strike the sample at some distance from<br />

the point of interaction with the primary electron beam. It is expected that the amount of<br />

stray radiation would be greatest when the sample is a rough surface or is near an<br />

edge. Backscattered electrons would be able to directly strike other surfaces of the<br />

sample and create x rays from these surfaces which may be well outside of the imaged<br />

area. To minimize stray radiation, the sample should be smooth, and located at the<br />

working distance that the detector is pointed at. It also helps to put the backscattered<br />

<strong>EDAX</strong> Phoenix Training Course - Spectrum Interpretation & Artifacts - page 4


electron detector in place because this acts as a sink for BSE and will protect the pole<br />

piece from being struck by BSE.<br />

A Warming Detector. The current generation of <strong>EDAX</strong> detectors have a temperature<br />

sensor which disables the detector when it runs out of liquid nitrogen. Just prior to the<br />

point when the detector will be disabled, you may notice an unusual, asymmetric “lowend<br />

noise peak”. At this time, the detector should have liquid nitrogen added to it. The<br />

system can still be used for qualitative work provided that very low-energy peaks are not<br />

the subject of the investigation. It should take approximately a half hour for the detector<br />

to regain the ideal cooled condition. As mentioned, in the cooling period it can still be<br />

used for qualitative work, but the system should not be calibrated or used for any work<br />

where the very best resolution is required.<br />

<strong>EDAX</strong> Phoenix Training Course - Spectrum Interpretation & Artifacts - page 5


energy of the detected x-ray can be determined. The charge is collected to the terminals<br />

electro-statically by a bias voltage applied across the detector. This voltage is 500 to<br />

1000 volts for most detectors. The transit time for the charge is ~50 nanoseconds.<br />

Since the desired signal is made up of moving charge, any stray moving charges<br />

(currents) must be minimized. A semiconductor produces a thermally generated current<br />

which must be reduced by lowering the temperature of the detector to eliminate this<br />

current as a noise compared to the signal. This involves cooling the detector with liquid<br />

nitrogen (77 degrees K). Si detectors need only to be cooled to ~90 degrees K, so some<br />

thermal losses can be tolerated. The Ge detector produces more current because of its<br />

smaller band gap (which also causes the signal to be larger), so the cooling requirement<br />

is more critical. Cooling the detectors with mechanical devices is not usually done when<br />

high image resolution is required because the vibration they produce cannot be<br />

tolerated. Liquid nitrogen cooling is almost universally used for under these conditions.<br />

In summary, the function of the detector is to convert x-ray energy into electrical charge.<br />

Later the processing of this signal will be discussed.<br />

2. The detector efficiency.<br />

Note in Figure 1 that the detector is made up of layers, some integral to the detector,<br />

and some external to it. An x-ray which strikes the detectors active area must be<br />

absorbed by it in order for the charge signal to be created. The absorption of x-rays in a<br />

layer of thickness t is given by Beer's Law,<br />

− ( µρt<br />

)<br />

I / I 0 = e<br />

Where :<br />

I = Final Intensity<br />

I 0 = Initial Intensity<br />

µ = mass absorption coefficient<br />

ρ = density<br />

t = thickness<br />

With this equation the efficiency of the detector can be calculated, taking into account<br />

the absorbing layers in front of the detector (Be or Polymer window, Au metalization,<br />

and dead layer) as well as the thickness and material (Si or Ge)of the active region of<br />

the detector. Figure 2 shows a set of Si detector curves for a various detector<br />

thicknesses and an 8 micron Be window. The 3 mm thickness would be most typical for<br />

commercially available detectors.<br />

The window material has the most dramatic effect on the low energy efficiency of either<br />

type of detector. Be windows have been used since the very first <strong>EDS</strong> detectors, but<br />

after the early 1980's, ultra thin windows made of polymers have also become<br />

available, and in the last few years have become widely used. The Be window is<br />

typically between 7 and 12 microns in thickness, with the very thinnest available about 5<br />

microns thick. These only allow practical detection of the x-rays of Na (Z=11) and higher<br />

atomic numbered elements. The polymer type of super ultra thin windows (SUTW)<br />

<strong>EDAX</strong> Phoenix Training Course - <strong>EDS</strong> Instrumentation & Signal Detection - page 2


allow detection of possibly Be (Z=4), or certainly B (Z=5) and above. Table 1 shows a<br />

comparison of the efficiency of an ultra thin window to a Be window.<br />

Table 1. Transmission of K x-rays through various windows<br />

Window<br />

Type<br />

B C N O F<br />

8 micron Be 0% 0% 0% 0% 5%<br />

Ultra Thin<br />

Polymer<br />

25% 85% 42% 60% 70%<br />

If there are contamination layers present on either the window or detector, then the low<br />

energy efficiency will be adversely affected. These layers could be due to oil<br />

contamination on the detector window or ice on the detector itself. (p. 14 of Ref 1) All<br />

precautions to prevent their formation should be taken. Each manufacturer will have a<br />

procedure to remove them if they are formed.<br />

3. The geometrical efficiency.<br />

The collection of x-rays by the detector is affected by the solid angle that the detector<br />

area intercepts of the isotropically emitted x-rays from the sample. The largest possible<br />

solid angle is desirable, especially when small regions of the sample are being<br />

investigated. The solid angle (Ω) in steradians is given by:<br />

2<br />

Ω= A/d<br />

Where:<br />

2<br />

A= detector area, mm<br />

d = the sample to detector distance<br />

The area is most commonly 10 or 30 mm 2 and the sample to detector distance varies<br />

from 10mm to 70mm, depending on the specific design of the microscope/detector<br />

interface. Solid angles of ~0.3 to 0.02 steradian are most common.<br />

4. Signal Processing:<br />

The charge from each x-ray event entering the detector must be processed and stored<br />

in a memory in order to make up a spectrum from an sample. The processing is outlined<br />

below:<br />

a. preamplifier:<br />

The preamplifier has the function of amplifying the charge signal and converting it to a<br />

voltage signal. A schematic is shown on the following page.<br />

<strong>EDAX</strong> Phoenix Training Course - <strong>EDS</strong> Instrumentation & Signal Detection - page 3


Detector<br />

FET<br />

Reset<br />

The preamplifier.<br />

<strong>EDAX</strong> Phoenix Training Course - <strong>EDS</strong> Instrumentation & Signal Detection - page 4<br />

C<br />

Output<br />

The FET is mounted adjacent to the detector, and is cooled to ~ 120 degrees K to<br />

reduce its electrical noise. The configuration shown is called a charge sensitive<br />

amplifier, and it converts charge at its input to a voltage output. C is a feedback<br />

capacitor. The FET must be reset periodically. This is accomplished by various<br />

methods, the most common being the pulsed optical feedback method.<br />

Voltage<br />

(mv)<br />

v<br />

Time ------><br />

Output signal of an x-ray event<br />

The output signal of an x-ray is shown above. The signal is small, and has noise form<br />

the FET added to it. The voltage step v, contains the size of the x-ray event, and it must<br />

be processed further to amplify its size, and increase its signal to noise ratio.


. The Signal processor.<br />

The processor is usually an analog amplifier system. Digital signal processors are not<br />

yet in widespread use, so will not be considered here. The amplifier system has several<br />

functions:<br />

1. Amplify the signal to 0-10 volt range.<br />

2. Filter the signal to optimize the signal to noise ratio (optimize the energy resolution).<br />

3. Detect x-ray events that overlap or "pileup' on one another, creating erroneous<br />

information.<br />

The filtering can be done with various kinds of networks, producing pulses that are<br />

gaussian shaped, triangularly shaped or trapezoidally shaped pulses. The networks can<br />

have several different time constants available. The effect is to produce pulses that are<br />

about 100 microseconds wide in order to achieve the best energy resolution. This<br />

increases the probability of pulse pileup and the need to correct for it.<br />

In order to detect the presence of pulse pileup, a second amplifier system is used just to<br />

detect x-ray events. This is often called an inspector channel. This inspector channel<br />

senses when two events will pileup, and then rejects these events. The quality of the<br />

spectrum is maintained, but dead time is introduced, which increases at higher count<br />

rates.<br />

For each time constant of the amplifier, a maximum throughput exists, which should not<br />

be exceeded, but which can be approached. If a higher count rate is desired, then a<br />

smaller time constant can be used, which will have a higher throughput, but a poorer<br />

resolution.<br />

Throughput curves for two time constants are shown below.<br />

<strong>EDAX</strong> Phoenix Training Course - <strong>EDS</strong> Instrumentation & Signal Detection - page 5


Stored CPS<br />

8000<br />

7000<br />

6000<br />

5000<br />

4000<br />

3000<br />

2000<br />

1000<br />

Thruput Curves<br />

0<br />

0 5000 10000 15000<br />

Input CPS<br />

20000 25000 30000<br />

c. The Multichannel analyzer.<br />

50usec<br />

100usec<br />

The processed pulses are digitized using an Analog to Digital Converter. This device<br />

measures the voltage height of each event and sorts the events into a multichannel<br />

memory. This memory is organized such that each channel represents 10ev of energy.<br />

From this digitized spectrum, the x-ray intensities from each element can be obtained. A<br />

spectrum is shown below.<br />

<strong>EDAX</strong> Phoenix Training Course - <strong>EDS</strong> Instrumentation & Signal Detection - page 6


5. Energy Resolution.<br />

The resolution of the peaks observed generally follows the following relation:<br />

2 2<br />

FWHM= SQRT[(FWHM) noise<br />

Where:<br />

F= f ano f actor= 0.11<br />

E= energy of the x-ray, ev<br />

+( 235 . F E ε)]<br />

ε=<br />

3.8e v/charge pair (Si), 2.96e v/charge pair (Ge)<br />

FWHM is the full width of the peak at half its maximum in ev. The noise term is due to<br />

the electronic noise, primarily from the FET after signal processing has been applied. Its<br />

value can vary from 40 eV to 100 eV, depending on the detector material, size, and age<br />

of system. A typical detector may give resolutions as shown below. Older detectors may<br />

show poorer resolution below 2 KeV because of dead layer effects. Newer detectors<br />

show performance closer to the theoretical values due to improved manufacturing<br />

techniques.<br />

6. Collimation:<br />

FWHM, ev<br />

250.00<br />

200.00<br />

150.00<br />

100.00<br />

50.00<br />

Resolution vs Energy for 70ev noise<br />

0.00<br />

0 5 10 15 20<br />

Ene r gy, Ke v<br />

FWHM<br />

The detector must be shielded from x-rays produced from locations at other than the<br />

sample. X-rays are being produced wherever scattered electrons are striking portions of<br />

the column. The electron microscope itself must be designed to minimize such x-rays<br />

from reaching the detector through the sides of the detector. Also a collimator is<br />

designed to prevent the detector from "seeing" materials around the sample which may<br />

be exposed to scattered electrons or x-rays from apertures in the column. In special<br />

cases in the transmission electron microscope, a specially designed low background<br />

sample holder is used which uses Be or another low atomic number alloy.<br />

<strong>EDAX</strong> Phoenix Training Course - <strong>EDS</strong> Instrumentation & Signal Detection - page 7


References:<br />

1. X-Ray Spectrometry in Electron Beam Instruments ( D. Williams, J. Goldstein, and D.<br />

Newbury, eds.) Plenum Press, New York (1995)<br />

2. Principles of Analytical Electron Microscopy ( D. C. Joy, A. Romig, Jr., and J.<br />

Goldstein, eds. ) Plenum Press, New York (1989)<br />

<strong>EDAX</strong> Phoenix Training Course - <strong>EDS</strong> Instrumentation & Signal Detection - page 8


X-RAY SIGNAL GENERATION<br />

Signal Origin<br />

The interaction of the electron beam with specimen produces a variety of signals, but<br />

the most useful to electron microscopists are these: secondary electrons (SE),<br />

backscattered electrons (BSE) and x rays. The SE signal is the most commonly used<br />

imaging mode and derives its contrast primarily from the topography of the sample. For<br />

the most part, areas facing the detector tend to be slightly brighter than the areas facing<br />

away from the SE detector, and holes or depressions tend to be very dark while edges<br />

and highly tilted surfaces are bright. These electrons are of a very low energy and very<br />

easily influenced by voltage fields.<br />

The BSE signal is caused by the elastic collision of a primary beam electron with a<br />

nucleus within the sample. Because these collisions are more likely when the nuclei are<br />

large (i.e. when the atomic number is large), the BSE signal is said to display atomic<br />

number contrast or “phase” contrast. Higher atomic number phases produce more<br />

backscattering and are correspondingly brighter when viewed with the BSE detector.<br />

X-ray signals are typically produced when a primary beam electron causes the ejection<br />

of an inner shell electron from the sample. An outer shell electron takes its place but<br />

gives off an x ray whose energy can be related to its nuclear mass and the difference in<br />

energies of the electron orbitals involved. The Kα x ray results from a K shell electron<br />

being ejected and an L shell electron moving into its position. A Kβ x ray occurs when<br />

an M shell electron moves to the K shell. The Kβ will always have a slightly higher<br />

energy than the Kα and is always much smaller. Similarly, an Lα x ray results from an<br />

M shell electron moving to the L shell to fill a vacancy (see Figure below). The<br />

occurrence of an Lβ x ray means that an N shell electron made the transition from the N<br />

shell to the L shell. The Lβ is always smaller and at a slightly higher energy than the Lα.<br />

The L-shell x rays are always found at lower energies than the K lines. Because the<br />

structure of the electron orbitals is considerably more complex than is shown below,<br />

there are actually many more L-shell x-ray lines that can be present (it is not uncommon<br />

to see as many as 5 or 6). M-shell x-ray peaks, if present will always be at lower<br />

energies than either the L or K series.<br />

<strong>EDAX</strong> Phoenix Training Course - X-ray Signal Generation - page 1


In energy-dispersive spectroscopy (<strong>EDS</strong>), the x rays are arranged in a spectrum by their<br />

energy and (most commonly) from low atomic number (low energy) to high atomic<br />

energy (higher energy). Typically, the energies from 0 to 10 keV will be displayed and<br />

will allow the user to view: the K-lines from Be (Z = 4) to Ga (Z = 31); L-lines Ca (Z =<br />

20) to Au (Z = 79), and M-lines from Nb (Z = 41) to the highest occurring atomic<br />

numbers. From the interpretation of the x-ray signal, we derive qualitative and<br />

quantitative information about the chemical composition of the sample at the<br />

microscopic scale.<br />

Spatial Resolution<br />

The spatial resolution of these signals are significantly different from each other. The<br />

figure below is a representation of the depth of penetration of the electron beam into a<br />

sample. No scale has been placed on this image, but the depth of penetration<br />

increases as the accelerating voltage of the primary beam is increased. It will also be<br />

deeper when the sample composition is of a lower density and/or is of a relatively low<br />

average atomic number. All three of the signals discussed above are produced<br />

throughout this interaction volume provided the beam electrons still have enough energy<br />

to generate it. However, some electrons or x rays may be of lesser energy and<br />

generated at a considerable depth in the sample. Thus, they may be absorbed and not<br />

generate a signal that can escape from the sample.<br />

The SE signal is one that is readily absorbed and therefore we are only able to detect<br />

the SE signal that originates relatively close to the surface (i.e. less than 10 nm). The<br />

BSE signal is of higher energy and is able to escape from a more moderate depth within<br />

the sample as shown. The x-ray signal can escape from a greater depth, although the<br />

x-ray signal absorption is actually variable depending upon its energy. For example,<br />

oxygen is of relatively low energy and can only escape from the near-surface region of<br />

the sample, while iron is of significantly higher energy and would escape from a greater<br />

depth. In quantitative x-ray analysis, it is possible to compensate for these effects with<br />

the absorption correction.<br />

<strong>EDAX</strong> Phoenix Training Course - X-ray Signal Generation - page 2


Although the discussion thus far has only mentioned the depths from which the signal<br />

can emerge, the width of the signal is proportional to its depth and provides an estimate<br />

of the signal resolution. Because the SE signal that is generated at even relatively<br />

shallow depths is absorbed, its resolution depends primarily on the position of the<br />

entering electron beam which has a spread related to the electron probe diameter or the<br />

spot size of the electron beam. The x-ray signal can emerge from greater depths<br />

(especially high-energy x rays) and the lateral spread of the primary beam electron can<br />

be quite large relative to the beam diameter. The only effective way to improve the<br />

resolution of these signals is to decrease the accelerating voltage which will decrease<br />

the beam penetration. The BSE signal is in between these two extremes and its<br />

resolution can be improved by decreasing the spot size to some extent, but the<br />

relationship between spot size and resolution is not as direct for the BSE signal as for<br />

the SE signal. The resolution of the BSE signal can also be improved by lowering the<br />

accelerating voltage, although this usually means having to increase the gain of the BSE<br />

detector and may result in a degradation of the signal-to-noise ratio of the image.<br />

Directionality of Signals<br />

All of the signals that emerge from the sample can be considered to be directional to at<br />

least some extent. A directional signal can be recognized in a photomicrograph of a<br />

sample that displays topography because there will be a very harsh contrast such that<br />

surfaces that face the detector will be bright and surfaces that face away from the<br />

detector will be dark. If the trajectory of the signal can be altered to favor detection, or if<br />

a symmetric array of detectors is employed, the effect of directionality is minimized.<br />

The trajectories of the SE signal are influenced by a positive voltage on a wire mesh<br />

network in front of the detector which attracts the SE from the sample, even from<br />

surfaces that face away from the detector. The BSE detector is typically arranged in an<br />

array such that they collect signals from a large, symmetrically arranged area. Some<br />

BSE detectors consist of two or more segments and the appearance of illumination<br />

direction in the imaged area changes drastically depending on which detector or<br />

segment is used for imaging. When all segments are selected, the result is a balanced,<br />

symmetric image that does not show an apparent directionality in its illumination.<br />

The x-ray signal is effectively the most directional of all the signals because there is only<br />

one detector and it is usually at a 35 degree angle to the surface of the sample. There<br />

is certainly no simple way to influence the trajectory of x rays to increase the efficiency<br />

<strong>EDAX</strong> Phoenix Training Course - X-ray Signal Generation - page 3


of the detector. As a result, if one is trying to collect the x-ray signal from a surface that<br />

slopes away from the detector, the x-ray count rate will be greatly diminished. If a<br />

topographic high area is between the imaged area and the x-ray path to the detector,<br />

there will also be few detected x rays. The effects of directionality for the x-ray signal<br />

are greatly diminished when working with polished samples rather than samples with a<br />

large topographic variation.<br />

The Analysis of Rough Surfaces or Particles<br />

There are difficulties associated with trying to analyze samples with anything other than<br />

a smooth surface. The figure on the next page shows a cross-sectional configuration<br />

with a detector at right and a small sample or portion of a sample at the left. The<br />

interaction volume is shown when the beam is at three different locations (‘A’, ‘B’ and<br />

‘C’) and the shaded region represents the area from which a low-energy signal may<br />

escape from the sample without being absorbed. The arrows represent the flux of these<br />

low-energy signals which will be highest at ‘A’ and at ‘C’ but relatively low at ‘B’.<br />

If the low-energy signal is the secondary electron signal and the detector at the right is<br />

the secondary electron detector (with a positive grid voltage), then the edges near ‘A’<br />

and ‘C’ will appear brighter than the center of the sample at ‘B’. Alternatively, if the lowenergy<br />

signal is some relatively low energy x ray (such as aluminum in a nickel alloy, or<br />

oxygen in a mineral, or the copper ‘L’ x-ray signal as compared to the copper ‘K’ x rays),<br />

and the detector is the <strong>EDS</strong> detector, then the height of the low energy peak would be<br />

highest at ‘C’ and lowest at ‘A’. Note that even though there would be a high flux of<br />

low-energy x rays leaving the sample at ‘A’, that these will not be detected because<br />

there is no detector positioned so as to intercept the signal.<br />

Another difficulty in analyzing particles or rough surfaces is when the surface of the<br />

particle has a slope which is not parallel to the surface of the stage. In the figure shown<br />

above, the upper surface is shown parallel to the stage and our consideration was really<br />

centered on what might be regarded as “edge” effects. However, when the sample<br />

surface is inclined toward the detector at a greater angle than the stage surface, then<br />

the take-off angle will be greater than that of a parallel surface. If our sample consists<br />

of a single, non-parallel sloping surface, then its take-off angle could be determined and<br />

an accurate analysis performed, provided that we have some way of knowing the local<br />

surface tilt of the sample. If the sample is rough and consists of many surfaces of<br />

variable orientation, then it is unlikely that a reliable analysis can be performed.<br />

<strong>EDAX</strong> Phoenix Training Course - X-ray Signal Generation - page 4


<strong>EDAX</strong> Phoenix Training Course - X-ray Signal Generation - page 5

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!