Model <strong>7800</strong> <strong>User</strong>s <strong>Manual</strong><br />

Revision E<br />

Figure 2.6<br />

2.2.5 LINEARIZATION MENU<br />

The linearization menu allows each channel to have its own linearization curve stored in<br />

the controller’s non-volatile memory. Input versus output points must be entered in<br />

percent of full scale values. This means if the range is 0-200 ppmH2S then 100 ppm is<br />

50% of full scale. Zero input will provide a zero output and 100% input a 100% output.<br />

Nine intermediate points may be entered to define the curve.<br />

Ch a n n e l XX<br />

Measur ement Name<br />

Al ar m 1<br />

Al ar m 2<br />

Al ar m 3<br />

Da t a F r o m<br />

Li near i ze<br />

Conf i gur e<br />

Ca l i b r a t e<br />

Entry menu<br />

EDIT<br />

NEXT<br />

Figure 2.7<br />

Ch a n n e l XX<br />

Li near i zat i on<br />

%I n p u t %Out put<br />

10 . 0 0<br />

20.00<br />

30.00<br />

40.00<br />

50.00<br />

60.00<br />

70.00<br />

80.00<br />

90.00<br />

10 . 0 0<br />

20.00<br />

30.00<br />

40.00<br />

50.00<br />

60.00<br />

70.00<br />

80.00<br />

90.00<br />

Linearization menu<br />

2.2.6 CONFIGURE MENU<br />

From the entry level setup menu in Figure 2.8 the CONFIGURE menu may be entered<br />

for setting variables defining how the controller presents monitored data to the various<br />

graphic displays.<br />

Ch a n n e l XX<br />

Me a s u r e me n t Na me<br />

Al ar m 1<br />

Al ar m 2<br />

Al ar m 3<br />

Da t a F r o m<br />

Li near i ze<br />

Co n f i g u r e<br />

Ca l i b r a t e<br />

Entry menu<br />

EDIT<br />

NEXT<br />

Figure 2.8<br />

Channel XX<br />

CONFI GURATION<br />

Measur ement Name<br />

Eu ni t s<br />

Zer o<br />

Sp an<br />

De c i ma l Po i n t s<br />

Ch a n n e l On ?<br />

Co p y T o<br />

Co p y No w?<br />

PCTLEL<br />

0.000<br />

10 0 . 0<br />

0<br />

YES<br />

XX<br />

Input / Copy To: menu<br />

<strong>Scott</strong> <strong>Safety</strong><br />

12<br />

087-0021 Rev F