Connection<br />
Relay outputs<br />
Analogue outputs<br />
Analogue inputs<br />
Digital input<br />
Digital interfaces (optional)<br />
3 potential free dry contacts, freely configurable, 6A, 250 V, max. 550 A<br />
2x (built in casing) or 3x (wall-mounted casing) 0/4 – 20 mA, electrically<br />
isolated, maximum 500 Ohm load<br />
4 measurement inputs, for chlorine, pH, ORP and temperature<br />
external controller stop or water deficiency sensor<br />
RS485, baud rate 9600, data format 8Bit, 1 start and 1 stop bit, no even<br />
parity. It is possible to activate the <strong>TOPAX</strong> <strong>DE</strong> with the software TopView<br />
(Windows, A/N 78349).<br />
1.14 Different types of <strong>TOPAX</strong> <strong>DE</strong> and its function<br />
Function <strong>DE</strong> 1 <strong>DE</strong> 2 <strong>DE</strong> 3 <strong>DE</strong> 4 <strong>DE</strong> 5 <strong>DE</strong> 6<br />
Free chlorine<br />
chlorine dioxide<br />
pH-value Metering Metering &<br />
Controlling<br />
Temperature<br />
Redox<br />
Flow input/<br />
Disturbance<br />
variable<br />
RS 485<br />
(optional)<br />
Metering & Controlling<br />
X<br />
Metering<br />
Metering<br />
Metering<br />
Metering & Controlling<br />
X X X X X X<br />
X<br />
NOTE!<br />
NOTE!<br />
The disturbance variable (<strong>DE</strong> 3/<strong>DE</strong> 6) only works in connection with 0/4-20 mA<br />
stepping motor.<br />
| Operating Manual <strong>TOPAX</strong> <strong>DE</strong> | Overview of Device <strong>Operation</strong> Functions