3. Commissioning<br />

3.1 Safety Instructions and Principles<br />

Plan the thermoregulation target and procedure.<br />

Determine the device configuration and system structure.<br />

Select an appropriate thermal fluid.<br />

Selection criteria for thermal fluids: Temperature range of the thermostat,<br />

application restrictions building safety regulations, your projected<br />

working temperature, viscosity, flash point.Position the thermostat<br />

and external devices on a stable and even surface. Ensure that the<br />

surface can safely hold the weight of the thermostat and external<br />

equipment.<br />

Caution! Potential risk of injury and material damage:<br />

Keep the device upright during transport.<br />

Place the device in an upright and stable position and make sure that<br />

it cannot tilt over.<br />

Keep the vicinity of the device clean: Prevent slip and tilting hazards.<br />

Lock the wheels of floor-mounted devices once you have positioned<br />

them as desired!<br />

Useful note: Place appropriately large collecting trays under the thermostat<br />

and the application.<br />

The use of the device is only allowed in normal environment conditions<br />

according to DIN EN 61010-1:2001:<br />

– Use only in the inside<br />

– Hight up to 2000m<br />

– Place on a dense, level, non-slip and non-flammable surface.<br />

– Place the device at a distance to walls and the ceiling that permits<br />

sufficient air circulation (heat dissipation, supply of fresh air for<br />

the thermostat and the workspace). A water-cooled unit requires<br />

a minimum of 10 cm and an air-cooled unit needs a minimum of<br />

20 cm.<br />

– Ambient temperature min- 5°C to max. 32°C<br />

– Short distances to the supplies<br />

– The device souldn’t be placed such that the access to the mains<br />

could be obstructed<br />

– Mains voltage fluctuation do not exceed + 10% of the stated input<br />

voltage<br />

– Transient overvoltage like they usualy appear in the mains supply<br />

– Applicable degree of pollution: 2<br />

– Class of overvoltage: II<br />

– Protection Classification: IP20<br />

The workspace of the thermostat must comply with local safety regulations.<br />

Please Note:<br />

All the safety instructions are vital and must be considered during<br />

the operation of the unit in compliance with the operating instructions.<br />