4. Thermoregulation<br />

via <strong>Control</strong>ler<br />

4.3 Editing Default Settings<br />

8. Alarm Configuration<br />

9. Alarm Clear<br />

Alarm: Read the alarm message - eliminate the cause - acknowledge<br />

the message (alarm clear).<br />

MAIN MENU:<br />

Alarm clear<br />

I-> Alarm config.<br />

Analog Interface<br />

ALARM CONFIG.:<br />

I-> Alarm mode<br />

Lower Alarm Limit<br />

Upper Alarm Limit<br />

Level alarm delay<br />

Exit<br />

ALARM MODE:<br />

Run Mode<br />

-> Stop Mode<br />

Please enter:<br />

Upper Alarm Limit<br />

40.0 C<br />

Please enter:<br />

Level alarm delay<br />

40<br />

MAIN MENU:<br />

2. setpoint<br />

I-> Alarm clear<br />

Alarm config.<br />

ALARM MODE:<br />

Please switch unit<br />

off and on<br />

ALARM CLEAR:<br />

I-> Restart<br />

Alarm clear<br />

Selection of the parameters that are to trigger an<br />

alarm monitored by the software.<br />

„Run Mode“ (ref. to „Alarm clear“)<br />

„Stop Mode“ (ref. to „Alarm clear“)<br />

Submenu selection:<br />

„Lower Alarm Limit“ / „Upper Alarm Limit“<br />

The lower and upper alarm limits define the<br />

temperatures that trigger an alarm and stop the<br />

thermoregulation process, depending on the<br />

Alarm Mode settings.<br />

Refer to 4.1 Safety Instructions!<br />

„Level Alarm Delay“<br />

The level alarm delay is defined by entering the<br />

delay time in seconds.<br />

At commissioning, a delay of 60 seconds is<br />

recommendable to provide for uninterrupted<br />

distribution.<br />

„Alarm Clear“ option<br />

Acknowledge the alarm.<br />

In the case of software-monitored alarm messages,<br />

the thermoregulation process is continued.<br />

The alarm message will persist until the cause<br />

auf the alarm has been eliminated and the alarm<br />

message has been acknowledged.<br />

„Alarm Mode“ when „stop mode“ has been previously<br />

set:<br />

The thermoregulation process will be stopped.<br />

Once the cause of the alarm has been eliminated<br />

and the alarm message has been acknowledged,<br />

it can be restarted manually.<br />

„Alarm Mode“ when „Run mode“ has been<br />

previously set after the alarm has been eliminated:<br />

„Restart“: The controller is restarted. The message<br />

disappears.<br />

„alam clear“: The message disappears.<br />

The following applies for each alarm mode:<br />

Faults caused by hardware failures (overtemperature protection / overheating<br />

protection, low fluid level protection, pump winding thermostat) result<br />

in cut-out of the thermoregulation process.<br />

36