Alarm<br />
Configu-<br />
ration<br />
Alarm<br />
Clear<br />
4. Thermoregulation<br />
via <strong>Control</strong>ler<br />
4.3 Editing Default Settings<br />
8. Alarm Configuration<br />
9. Alarm Clear<br />
Alarm: Read the alarm message - eliminate the cause - acknowledge<br />
the message (alarm clear).<br />
MAIN MENU:<br />
Alarm clear<br />
I-> Alarm config.<br />
Analog Interface<br />
ALARM CONFIG.:<br />
I-> Alarm mode<br />
Lower Alarm Limit<br />
Upper Alarm Limit<br />
Level alarm delay<br />
Exit<br />
ALARM MODE:<br />
Run Mode<br />
-> Stop Mode<br />
Please enter:<br />
Upper Alarm Limit<br />
40.0 C<br />
Please enter:<br />
Level alarm delay<br />
40<br />
MAIN MENU:<br />
2. setpoint<br />
I-> Alarm clear<br />
Alarm config.<br />
ALARM MODE:<br />
Please switch unit<br />
off and on<br />
ALARM CLEAR:<br />
I-> Restart<br />
Alarm clear<br />
Selection of the parameters that are to trigger an<br />
alarm monitored by the software.<br />
„Run Mode“ (ref. to „Alarm clear“)<br />
„Stop Mode“ (ref. to „Alarm clear“)<br />
Submenu selection:<br />
„Lower Alarm Limit“ / „Upper Alarm Limit“<br />
The lower and upper alarm limits define the<br />
temperatures that trigger an alarm and stop the<br />
thermoregulation process, depending on the<br />
Alarm Mode settings.<br />
Refer to 4.1 Safety Instructions!<br />
„Level Alarm Delay“<br />
The level alarm delay is defined by entering the<br />
delay time in seconds.<br />
At commissioning, a delay of 60 seconds is<br />
recommendable to provide for uninterrupted<br />
distribution.<br />
„Alarm Clear“ option<br />
Acknowledge the alarm.<br />
In the case of software-monitored alarm messages,<br />
the thermoregulation process is continued.<br />
The alarm message will persist until the cause<br />
auf the alarm has been eliminated and the alarm<br />
message has been acknowledged.<br />
„Alarm Mode“ when „stop mode“ has been previously<br />
set:<br />
The thermoregulation process will be stopped.<br />
Once the cause of the alarm has been eliminated<br />
and the alarm message has been acknowledged,<br />
it can be restarted manually.<br />
„Alarm Mode“ when „Run mode“ has been<br />
previously set after the alarm has been eliminated:<br />
„Restart“: The controller is restarted. The message<br />
disappears.<br />
„alam clear“: The message disappears.<br />
The following applies for each alarm mode:<br />
Faults caused by hardware failures (overtemperature protection / overheating<br />
protection, low fluid level protection, pump winding thermostat) result<br />
in cut-out of the thermoregulation process.<br />
36