Analog<br />
Interface<br />
4. Thermoregulation<br />
via <strong>Control</strong>ler<br />
4.3 Editing Default Settings<br />
15. Analog Interface - Parameter Output<br />
(Applies to all thermostats with <strong>Compatible</strong> <strong>Control</strong> CC3)<br />
MAIN MENU:<br />
Alarm mode<br />
I-> Analog Interface<br />
Display<br />
ANALOG INTERFACE:<br />
Temp. T1 (Zero)<br />
Temp. T2 (Span)<br />
Conf. Input<br />
I-> Conf. Output<br />
Exit<br />
ANALOG OUTPUT:<br />
I-> Output value<br />
Current at T1<br />
Current at T2<br />
Exit<br />
SOURCE ANALOG OUT:<br />
No output<br />
Internal temp.<br />
External temp.<br />
Programmer<br />
Setpoint<br />
I-> Manual value<br />
Exit<br />
ANALOG OUTPUT:<br />
Output value<br />
I-> Current at T1<br />
Current at T2<br />
Exit<br />
Please enter:<br />
DAC-Output at T1<br />
591<br />
ANALOG OUTPUT:<br />
Output value<br />
Current at T1<br />
I-> Current at T2<br />
Exit<br />
Please enter:<br />
DAC-Output at T2<br />
4104<br />
Select „Analog Interface“ (AIF): The thermoregulation<br />
process is controlled via an analog<br />
signal (currents from 4 to 20 mA), the<br />
strength of which represents the setpoint.<br />
The temperature range can be set by the<br />
user.<br />
Submenu selection: „Output Value“<br />
Definition of the measuring point for temperature<br />
output values.<br />
„No Output“ Constant output of 4 mA as a<br />
power supply, e.g. for an external thermometer.<br />
„Internal Temp.“: Measurement using an<br />
internal sensor in the thermostat. In the case<br />
of a bath thermometer, the sensor is inside<br />
the bath.<br />
„External Temp.“: Measurement with an<br />
external sensor, e.g. in an external bath.<br />
„Programmer“: In a thermoregulation program,<br />
each segment can be assigned an<br />
output current, e.g. for adapting the speed<br />
of an external pump to individual program<br />
segments.<br />
„Manual Value“ Any percentage from 0<br />
%...100 % analog to 4...20 mA with encoder.<br />
Submenu selection: „Current at T1“<br />
Presetting of a converter output value for T1<br />
(Zero): The encoder is used to set a value<br />
that results in the desired current value for<br />
T1, e.g. 4 mA at the analog setpoint encoder.<br />
Submenu selection: „Current at T2“<br />
Presetting of a converter output value for T2<br />
(Span): The encoder is used to set a value<br />
that results in the desired current value for<br />
T2, e.g. 20 mA at the analog setpoint encoder.<br />
Back to main menu<br />
Caution: The controller module may be destroyed if currents<br />
exceeding 20 mA are used and/or if the polarity is confused!<br />