Analog<br />

Interface<br />

4. Thermoregulation<br />

via <strong>Control</strong>ler<br />

4.3 Editing Default Settings<br />

15. Analog Interface - Parameter Output<br />

(Applies to all thermostats with <strong>Compatible</strong> <strong>Control</strong> CC3)<br />

MAIN MENU:<br />

Alarm mode<br />

I-> Analog Interface<br />

Display<br />

ANALOG INTERFACE:<br />

Temp. T1 (Zero)<br />

Temp. T2 (Span)<br />

Conf. Input<br />

I-> Conf. Output<br />

Exit<br />

ANALOG OUTPUT:<br />

I-> Output value<br />

Current at T1<br />

Current at T2<br />

Exit<br />

SOURCE ANALOG OUT:<br />

No output<br />

Internal temp.<br />

External temp.<br />

Programmer<br />

Setpoint<br />

I-> Manual value<br />

Exit<br />

ANALOG OUTPUT:<br />

Output value<br />

I-> Current at T1<br />

Current at T2<br />

Exit<br />

Please enter:<br />

DAC-Output at T1<br />

591<br />

ANALOG OUTPUT:<br />

Output value<br />

Current at T1<br />

I-> Current at T2<br />

Exit<br />

Please enter:<br />

DAC-Output at T2<br />

4104<br />

Select „Analog Interface“ (AIF): The thermoregulation<br />

process is controlled via an analog<br />

signal (currents from 4 to 20 mA), the<br />

strength of which represents the setpoint.<br />

The temperature range can be set by the<br />

user.<br />

Submenu selection: „Output Value“<br />

Definition of the measuring point for temperature<br />

output values.<br />

„No Output“ Constant output of 4 mA as a<br />

power supply, e.g. for an external thermometer.<br />

„Internal Temp.“: Measurement using an<br />

internal sensor in the thermostat. In the case<br />

of a bath thermometer, the sensor is inside<br />

the bath.<br />

„External Temp.“: Measurement with an<br />

external sensor, e.g. in an external bath.<br />

„Programmer“: In a thermoregulation program,<br />

each segment can be assigned an<br />

output current, e.g. for adapting the speed<br />

of an external pump to individual program<br />

segments.<br />

„Manual Value“ Any percentage from 0<br />

%...100 % analog to 4...20 mA with encoder.<br />

Submenu selection: „Current at T1“<br />

Presetting of a converter output value for T1<br />

(Zero): The encoder is used to set a value<br />

that results in the desired current value for<br />

T1, e.g. 4 mA at the analog setpoint encoder.<br />

Submenu selection: „Current at T2“<br />

Presetting of a converter output value for T2<br />

(Span): The encoder is used to set a value<br />

that results in the desired current value for<br />

T2, e.g. 20 mA at the analog setpoint encoder.<br />

Back to main menu<br />

Caution: The controller module may be destroyed if currents<br />

exceeding 20 mA are used and/or if the polarity is confused!<br />

40