ITEM NO.10 – SPECTRAL CONFOCAL MICROSCOPE, QTY: 1 NO. (FILE REF:<br />

34/<strong>CCMB</strong>/APP/407/P12)<br />

SPECTRAL CONFOCAL MICROSCOPE FOR HIGH SENSITIVE AND HIGH RESOLUTION IMAGING OF<br />

BIOLOGICAL SAMPLES:<br />

The system should be of latest state of the art technology capable of high sensitive spectral confocal imaging of fixed<br />

& Live Biological samples. The system should include multichannel Fluorescence imaging with Z-stack, time-lapse, co<br />

localization, FRET, FRAP, photon counting, Photo activation and conversion.<br />

The system should consist of:<br />

1. Fully Motorized Inverted Fluorescence Research Microscope for BF/DIC/FL with dedicated TFT/LCD display for<br />

full control of Microscope. Motorized XY Scanning stage with universal sample holders for slides, Petri dish and<br />

other live cell sample holders like 6 well plate, 96 well plate for tile, mosaic and multipoint imaging. Motorized 6<br />

position DIC nosepiece, motorized BF/PH/DIC Condenser, 6 position motorized FL turret. High precision built in Z-<br />

focus drive with step resolution of 15nm or better. Transmitted and Reflected Light illumination with 100W halogen<br />

and 120W metal halide with PC control and long lifetime of 2000 hours. High resolution confocal grade Plan<br />

apochromat objectives corrected for both UV & VIS lines 10x/0.45, 20x/0.70,40x/0.85,40x/1.30oil, 60/63x/1.40oil,<br />

100x/1.4 oil immersion along with DIC accessories for all objectives. Bandpass fluorescent filters for DAPI, GFP,<br />

FITC, Cy3 and Cy5.<br />

2. Anti vibration table.<br />

3. Completely motorized Point scanning laser based Confocal unit with built in Internal high sensitive spectral<br />

detectors (Ga AsP/HyD) based imaging for at least 2 channels in addition to the sta ndard 2 channel PMT<br />

detection with independent gain controls for all channels. All the FL detectors of the scan head should be filter free<br />

with freely selectable emission band width detection capability to suit to the emission spectra of the dyes. All<br />

detectors should be spectral. The detection unit should be inbuilt in the scanner unit. The system should include<br />

simultaneous detection and separation of at least 5 fluorophores based on spectral detectors. High sensitive<br />

motorized dichroic system with low angle of incidence with high efficient optics.<br />

The system should be capable of recording emission spectra with minimum spectral resolution 5nm or better.<br />

Computer controlled continuously variable confocal pinhole with software control. Scan resolution of 6Kx6K or<br />

above. Frame rates of 7 fps or higher with spectral detectors @512x512 pixels in spectral detection mode. Real<br />

ROI scan and bleach with various ROI shapes should be possible for FRAP experiments. The scan field diagonal<br />

should be 20 mm F.O.V. or more.<br />

4. PMT based transmitted light detector for DIC imaging.<br />

5. Laser unit with control electronics including multiline Ar 458/488/514nm, DPSS 561nm, HeNe 633nm & BDL<br />

405/408nm, all laser lines including 405nm should be controlled through 8 channel AOTF for fast laser switching<br />

and attenuation. Seperate laser ports for UV/405nm, Vis, IR and additional port for future upgrades is essential.<br />

System should have capability to upgrade for tunable laser with visible wavelength.<br />

6. Latest branded 64 bit control computer with Intel Xeon 6 Core Processor, DDR RAM 8 GB HDD: 1 TB SATA,<br />

DVD, Super Multi SATA +R/RW, Graphics : AT Fire GL V5200 256MB DH DVI, Gigabit Ethernet, Win 7 Ultimate<br />

64 bit OS , USB 2.0, Fire wire. Large 30” LCD/LED/ TFT monitor.<br />

7. Confocal system control software capable of controlling all motorized functions of microscope, scan head, lasers,<br />

AOTF including image acquisition & processing. Image acquisition for 3D, 4D, on-line spectral Imaging based on<br />

lambda stacks. Time series, FRAP, FRET, Photo activation and conversion. Imaging without bleed through and<br />

autofluorescence separation by online emission fingerprinting technique. Advanced multidimensional software for<br />

3D analysis with Deconvolution, multichannel volume rendering, reconstruction, measurements across z stack,<br />

movie, co-localization with histogram analysis, intensity profiles for quantification etc....<br />

Optional items:<br />

1) Additional HeNe 594nm laser should be offered.<br />

2) Upgradable to FCS/FCCS unit for Single Molecule Detection: The FCS unit with APD detectors should be capable<br />

of performing auto and cross correlation measurements in live cells and solutions for a wide range of dyes and proteins.<br />

3) The FCS unit should have facility for elimination/suppression of other excitation laser lines. All laser lines for Confocal<br />

Imaging should be capable of working in FCS/FCCS mode. Dedicated Plan Apochromat 40x/1.20 or 63x/1.2 N.A.<br />

water should be offered along with the system. FCS measurement software for Auto and Cross correlation<br />

capabilities to be quoted.<br />

4) Any patented detection system which is more sensitive and multi user friendly may also be offered.<br />

5) Additional GaAsP/HyD detector to be offered.<br />

BID DOCUMENT (YC)<br />