ITEM NO.15 – FLOOR MODEL ULTRACENTRIFUGE, QTY: 1 NO. (FILE REF:<br />

34/<strong>CCMB</strong>/APP/412/P12)<br />

SPECIFICATIONS OF FLOOR MODEL ULTRACENTRIFUGE:<br />

FLOOR MODEL ULTRACENTRIFUGE WITH ROTORS AND ACCESSORIES<br />

Applications: Separation of Subcellular Particles, Varus and Viral Particles, Rate-Zonal Separation Proteins in<br />

Sucrose Gradient, Separation of Lipoproteins, Pelleting RNA through a CsCl Gradient and Isopycnic Separation of<br />

Plasmid DNA.<br />

I<br />

SPECIFICATIONS:<br />

Maximum Speed : 100,000 rpm or more<br />

Maximum g Force (RCF) : 800,000 x g or more<br />

Control and Display : Microprocessor with Large touch-screen for user<br />

friendliness and convenience<br />

Speed Control Accuracy : +/- 10 rpm or less of set speed<br />

Refrigeration System : Thermoelectric cooling for lower power consumption<br />

Temperature Range : 0 to 40 C with 1C steps<br />

Temperature Control : +/- 0.5 C of set temperature<br />

Ambient temperature range of operation : 20C to 35C or more<br />

Sample imbalance tolerance : + 5 ml or 10% whichever is greater.<br />

Drive cooling : Air Cooled for lower power consumption and<br />

maintenance<br />

Accelerations/Decelerations profiles : 10 Acceleration/Deceleration or more profiles<br />

Sound level : < 55 dBa or less<br />

Heat Output : Should be less than 4000 BTU/hr<br />

Power : 220 Volts, 50Hz, (Single Phase) with wide<br />

tolerance for power fluctuations.<br />

II<br />

Essential Features:<br />

- System should automatically check for possible errors with rotor induced parameters and<br />

automatically make necessary correction.<br />

- Ultracentrifuge should be able to accept Fixed Angle Rotor, Vertical Rotor, Near Vertical Rotor or<br />

equivalent Swinging Bucket Rotors.<br />

- Samples volumes for centrifugation from 5 ml to 1500 ml.<br />

- Adaptors to accommodate small volume samples to large volumes should be available without<br />

sacrificing the maximum g force of rotors.<br />

- Ability to remove moisture with vacuum.<br />

- Automatic restart with set parameters after power interruptions.<br />

- Multi steps users programs.<br />

- Simulation software provision.<br />

- Speed / temperature vs. time plot.<br />

- All regulatory and safety standards to be complied.<br />

- Many user settable programs.<br />

- Provision for Bio-safety features/upgradation to be offered as option.<br />

III<br />

Rotors:<br />

Fixed Angle Rotor :<br />

1. Capacity : 8 x 38 ml or more<br />

Maximum Speed : 70,000 rpm 1 No.<br />

Maximum g force : 500,000 g or more<br />

K factor<br />

: 45 or less<br />

The rotor should be able to accommodate low volumes such as approximately 25ml<br />

or 15 ml with adapters and without compromising the Maximum rcf of 500,000 g or more.<br />

Near Vertical Rotor :<br />

BID DOCUMENT (YC)<br />