ITEM NO.15 – FLOOR MODEL ULTRACENTRIFUGE, QTY: 1 NO. (FILE REF:<br />
34/<strong>CCMB</strong>/APP/412/P12)<br />
SPECIFICATIONS OF FLOOR MODEL ULTRACENTRIFUGE:<br />
FLOOR MODEL ULTRACENTRIFUGE WITH ROTORS AND ACCESSORIES<br />
Applications: Separation of Subcellular Particles, Varus and Viral Particles, Rate-Zonal Separation Proteins in<br />
Sucrose Gradient, Separation of Lipoproteins, Pelleting RNA through a CsCl Gradient and Isopycnic Separation of<br />
Plasmid DNA.<br />
I<br />
SPECIFICATIONS:<br />
Maximum Speed : 100,000 rpm or more<br />
Maximum g Force (RCF) : 800,000 x g or more<br />
Control and Display : Microprocessor with Large touch-screen for user<br />
friendliness and convenience<br />
Speed Control Accuracy : +/- 10 rpm or less of set speed<br />
Refrigeration System : Thermoelectric cooling for lower power consumption<br />
Temperature Range : 0 to 40 C with 1C steps<br />
Temperature Control : +/- 0.5 C of set temperature<br />
Ambient temperature range of operation : 20C to 35C or more<br />
Sample imbalance tolerance : + 5 ml or 10% whichever is greater.<br />
Drive cooling : Air Cooled for lower power consumption and<br />
maintenance<br />
Accelerations/Decelerations profiles : 10 Acceleration/Deceleration or more profiles<br />
Sound level : < 55 dBa or less<br />
Heat Output : Should be less than 4000 BTU/hr<br />
Power : 220 Volts, 50Hz, (Single Phase) with wide<br />
tolerance for power fluctuations.<br />
II<br />
Essential Features:<br />
- System should automatically check for possible errors with rotor induced parameters and<br />
automatically make necessary correction.<br />
- Ultracentrifuge should be able to accept Fixed Angle Rotor, Vertical Rotor, Near Vertical Rotor or<br />
equivalent Swinging Bucket Rotors.<br />
- Samples volumes for centrifugation from 5 ml to 1500 ml.<br />
- Adaptors to accommodate small volume samples to large volumes should be available without<br />
sacrificing the maximum g force of rotors.<br />
- Ability to remove moisture with vacuum.<br />
- Automatic restart with set parameters after power interruptions.<br />
- Multi steps users programs.<br />
- Simulation software provision.<br />
- Speed / temperature vs. time plot.<br />
- All regulatory and safety standards to be complied.<br />
- Many user settable programs.<br />
- Provision for Bio-safety features/upgradation to be offered as option.<br />
III<br />
Rotors:<br />
Fixed Angle Rotor :<br />
1. Capacity : 8 x 38 ml or more<br />
Maximum Speed : 70,000 rpm 1 No.<br />
Maximum g force : 500,000 g or more<br />
K factor<br />
: 45 or less<br />
The rotor should be able to accommodate low volumes such as approximately 25ml<br />
or 15 ml with adapters and without compromising the Maximum rcf of 500,000 g or more.<br />
Near Vertical Rotor :<br />
BID DOCUMENT (YC)<br />
Page 54 of 63