ITEM NO.16 – ARABIDOPSIS PLANT GROWTH CHAMBER,<br />

34/<strong>CCMB</strong>/APP/413/P12)<br />

QTY: 1 NO. (FILE REF:<br />

SPECIFICATIONS, FEATURES AND CLAUSES FOR ARABIDOPSIS PLANT GROWTH CHAMBER:<br />

1. Temperature Range : 10 °C to 40 °C 1.0 °C (Lights On)<br />

4 °C to 40 °C 0.5 °C (Lights Off)<br />

2. Internal Volume / Capacity : Approx. 70 to 105 Cu ft.<br />

3. Growth Height : Approx. 15 inches per shelf<br />

4. Growth Area (floor area) : Approx. 30 Sq feet<br />

5. Racks/Shelves : Three Racks with 22 gauge stainless steel<br />

6. Light intensity (micromoles/m 2 /s) : 300 micromoles/m 2 /s or more<br />

2 level programming for fluorescent lighting<br />

1 level programming for incandescent lighting<br />

7. Operating Power Supply : 230 Volts, 50 Hz.<br />

8. Refrigeration System : Non CFC with outdoor Remote Air Cooled<br />

Condenser for all weather conditions with hot<br />

gas bypass system for continuous compressor operation<br />

9. Heat rejection to ambient : Less than 9000 BTU per hour.<br />

Essential features<br />

- Microprocessor controlled with digital display.<br />

- Programmable temperature and light settings.<br />

- High/Low temperature safety limits and alarms.<br />

- Upto 2 programmable light levels.<br />

- Bright digital LED or VFD display.<br />

- Air flow thru Plenum and Air diffusers.<br />

Optional :<br />

Humidification : Pan type with electronic RH sensor<br />

Important Clauses:<br />

1. The external dimensions of the chamber should not exceed 6 ½’ (length / width) x 38” (depth) x 7’ (height)<br />

which is the space available for the installation of the chamber.<br />

2. Cost should include a two year comprehensive warranty or more including parts and labour.<br />

3. All safety features incorporated for environmental conditions with certifications should be clearly stated.<br />

4. The list of users in India who are using similar Models of Plant Growth Chamber for growing Arabidopsis<br />

plants should be provided. A Certificate to the effect that the Chamber is functioning well and the period<br />

for which it has been in use should be included. This is necessary because the proper functioning of the<br />

Growth Chamber in promoting growth of Arabidopsis plants is critical for the success of our project.<br />

5. Detailed service manual with electronic circuit diagrams and engineering details should accompany the Plant<br />

growth chamber. Commitment for the same should be made in the offer.<br />

6. Please indicate local service support available and response time for attending a service call during and after<br />

warranty.<br />

7. Number of years for which the spare parts would be available should be indicated.<br />