Acquire CD spectra in continuous temperature ramp mode for determining thermodynamic data<br />
(simultaneous analysis of T melt curves at each wavelength, to observe the secondary/tertiary structural<br />
changes at each thermal transition, record Tmelt curves at fewer wavelengths in a single temperature<br />
ramp experiment.<br />
Acquire CD spectra as a function of temperature in a stepped temperature ramp mode i.e CD vs Wavelength<br />
vs Temperature data.<br />
Acquiring CD signal in a continuous temperature ramp mode for determining thermodynamic data at<br />
single wavelength (Tmelt technique for determining protein stability)<br />
Post acquisition smoothing of the CD spectra would be preferred to smoothing during acquisition so that raw data is<br />
never lost.<br />
Should be able to export data to other file formats like text files etc.<br />
System should include a standalone PC with the latest Operating system and standard communication interface and<br />
control.<br />
The offered system should include the following software along with the required licenses (min imum of 5 user<br />
licenses)<br />
i. The basic control & analysis software should be capable of full system control monochromator settings, post<br />
acquisition smoothing, export to spreadsheet and other file formats, kinetic data acquisition, multiwavelength<br />
variable temperature programming, denatured protein analysis, curve fitting etc.<br />
ii.<br />
iii.<br />
Secondary structure analysis program to be included<br />
Program for analysis of multiwavelength temperature scan data capable of calculating midpoint temperatures,<br />
Van’t Hoff enthalpies of thermal transitions, calculate concentrations against temperature, providing<br />
thermodynamic and structural <strong>information</strong> etc.<br />
System should be flexible and allow field upgrades for different applications and measurement modes like Peltier<br />
temperature control, Simultaneous fluorescence detection, Stopped flow CD etc<br />
Suitable Nitrogen gas flow meter for purging the optics and the sample compartment should be included with the<br />
system.<br />
3) Accessories to be offered optionally<br />
i. Peltier temperature control accessory for CD.<br />
ii. Peltier temperature control accessory for CD, simultaneous Fluorescence and fluorescence detected CD)<br />
Items {3 (i) and 3 (ii)} should have the following specifications and features:<br />
Temperature range: -10 deg to 100deg C<br />
Temperature accuracy: ± 0.05 deg C or better<br />
Should have cuvette temperature probe for monitoring sample temperature<br />
Stirring facility to be included<br />
Should be able to accommodate both the rectangular and cylindrical cuvettes of different pathlengths.<br />
Hardware control of the Peltier should be through the control software of the CD instrument.<br />
Step wise and continuous temperature ramping facility.<br />
Additional detectors, suitable recirculating cooler (water bath) and other accessories needed for the above<br />
Peltier units {Items 3 (i) and 3 (ii)} to be included in the offer.<br />
iii.<br />
iv.<br />
Accessory for stopped flow with peltier temperature control, Four cell auto changer with Peltier temperature<br />
control, Fluorescence polarization and anisotropy accessory, Diffuse reflectance accessory to be offered.<br />
Cells of different pathlength 0.01mm, 0.02 mm, 0.05 mm (rectangular cells); 0.1 mm,0.2mm, 0.5mm, 1mm,<br />
2mm, 5mm & 10 mm ( rectangular & cylindrical cells) -- 4 each<br />
v. Suitable adapters / cell holders for the above to be offered.<br />
vi. Spare lamps (2 nos )<br />