<strong>SUMMARY</strong> REPORT<br />
WIDE FIELD FIBER-FED OPTICAL<br />
MULTI-OBJECT SPECTROMETER (WFMOS)<br />
Table 3.12-1: Spectrograph requirements based on the four surveys. The first three rows are a<br />
repeat of the science case. The other rows are derived requirements.<br />
Requirement DELZ DEHZ<br />
GALR<br />
Blue Red<br />
GAHR<br />
Min λ (nm) 630 420 480 818 480<br />
Max λ (nm) 970 670 544 885 680<br />
Resolving power 3500 1500 5000 5000 20000<br />
SRE width (Angstroms) * 2.82 3.63 1.03 1.70 0.29<br />
Number of SREs per spectrum 1488 688 680 412 6897<br />
Number of spectral orders per fiber 1 1 1 1 4<br />
Length of a single spectral order (mm) 80 37 37 22 93<br />
No. of spectrographs required 2.6 1.2 1.2 0.7 3.0<br />
% of detector area used 86.3% 39.9% 36.3% 22.8% 100%<br />
* SRE – Spectral Resolution Element<br />
Table 3.12-2: Basic derived spectrograph parameters. These are driven by the GAHR survey,<br />
which is the most demanding of the four surveys.<br />
Detector format for one spectrograph<br />
Format of a single CCD<br />
Number of spectrographs 3<br />
93 mm × 93 mm<br />
93 mm × 46.5 mm, buttable along the long side<br />
Collimator f-ratio<br />
Camera f-ratio<br />
Minimum beam size<br />
Fiber pitch at slit<br />
Projected fiber size at detector<br />
f/2.12<br />
f/1.07<br />
~300-mm diameter<br />
230 µm, center to center<br />
54-µm diameter<br />
SRE* width (µm at detector) 54.0 µm<br />
SRE height (µm at detector) = projected fiber pitch 116.25 µm<br />
* SRE – Spectral Resolution Element<br />
47