Equivalent Performance to Sub-2 μm<br />
with Lower Pressure<br />
With solid core design, tight particle size distribution and uniform packed bed<br />
<strong>Accucore</strong> <strong>HPLC</strong> columns have broadly equivalent performance to sub-2 μm columns<br />
and yet generate only a fraction of the backpressure.<br />
mAU<br />
fully porous < 2 μm<br />
100 x 2.1 mm<br />
<strong>Accucore</strong> RP-MS 2.6 µm<br />
100 x 2.1 mm<br />
0 0.5 1.0 1.5 2.0 2.5<br />
Minutes<br />
Mobile phase A: water<br />
Mobile phase B: acetonitrile<br />
Gradient:<br />
Flow:<br />
Temperature: 40 °C<br />
Injection: 1 µL<br />
Detection:<br />
Analytes:<br />
65–95 % B in 1.7 minutes<br />
95 % B for 0.3 minutes<br />
500 µL/min<br />
UV at 247 nm (0.1s rise time, 20 Hz)<br />
1. Acetophenone<br />
2. Propiophenone<br />
3. Butyrophenone<br />
4. Valerophenone<br />
5. Hexanophenone<br />
6. Heptanophenone<br />
7. Octanophenone<br />
Pressure<br />
<strong>Accucore</strong> RP-MS<br />
2.6 µm, 100 x 2.1 mm<br />
Fully Porous<br />
< 2 µm, 100 x 2.1 mm<br />
Resolution (critical pair) 3.72 4.20<br />
Run time (min) 3.50 3.50<br />
Maximum pressure (bar) 171 338<br />
Lower backpressure eliminates the requirement for U<strong>HPLC</strong><br />
systems with maximum pressure ratings >600 bar. If a U<strong>HPLC</strong><br />
system is used then the lower backpressure reduces wear<br />
on the instrument.<br />
9