Accucore HPLC Columns - Cromlab
Accucore HPLC Columns - Cromlab
Accucore HPLC Columns - Cromlab
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Accucore</strong> 150-Amide-HILIC<br />
Hydrophobicity<br />
Low<br />
pH Range<br />
0<br />
Pore Size<br />
2 to 8<br />
High<br />
14<br />
HILIC<br />
0 150 Å<br />
300<br />
Particle Size 2.6 µm<br />
• Amide phase bonded onto 150 Å pore diameter solid core particles<br />
• High retention of a broad range of hydrophilic analytes in HILIC mode<br />
• Recommended for hydrophilic biomolecules such as glycans<br />
<strong>Accucore</strong> 150-Amide-HILIC is designed for the separation of hydrophilic biomolecules in HILIC mode.<br />
The amide bonded phases provide strong hydrogen bonding interaction between the stationary phase and the<br />
analytes, resulting in unique selectivity compared to other HILIC phases. Combined with larger pore size of the solid<br />
core particles, <strong>Accucore</strong> 150-Amide-HILIC is well suited for separating a variety of hydrophilic molecules, including<br />
carbohydrates and peptides. As a result the <strong>Accucore</strong> 150-Amide-HILIC is an excellent choice for glycan separations.<br />
Glycan Ladder<br />
Counts<br />
1200000<br />
1100000<br />
1000000<br />
900000<br />
800000<br />
700000<br />
600000<br />
500000<br />
400000<br />
300000<br />
200000<br />
100000<br />
0<br />
-100000<br />
-200000<br />
1<br />
2<br />
3<br />
A<br />
(A) 2 µL injection of sample, where 11 glycans were separated.<br />
4<br />
30 35 40 45<br />
Minutes<br />
(B) 5 µL injection of sample, zoomed-in to the later part of the gradient rise.<br />
A further 10 glycans were detected.<br />
5<br />
350000<br />
300000<br />
250000<br />
200000<br />
150000<br />
100000<br />
50000<br />
6<br />
0<br />
10<br />
7<br />
5 10 15 20 25 30 35 40 45<br />
8<br />
Minutes<br />
9<br />
10 11<br />
B<br />
11<br />
12<br />
13<br />
14<br />
15<br />
16 17 18 19 20 21<br />
Column:<br />
Mobile phase A:<br />
<strong>Accucore</strong> 150-Amide-HILIC, 2.6 µm,<br />
100 x 2.1 mm<br />
acetonitrile<br />
Mobile phase B: 50 mM ammonium formate, pH 4.5<br />
Gradient:<br />
Flow:<br />
Backpressure at starting<br />
conditions:<br />
Run time:<br />
20–50 % B in 40.0 minutes<br />
50 % B for 5.0 minutes<br />
50–20 % B in 0.5 minutes<br />
50 % B for 4.5 minutes<br />
500 µL/min<br />
110 bar<br />
50 minutes<br />
Temperature: 60 °C<br />
Injection:<br />
Injection wash solvent:<br />
Fluorescence detector<br />
acquisition parameters:<br />
2 µL to 5 µL of sample.<br />
80:20 (v/v) acetonitrile:water.<br />
330 nm excitation wavelength;<br />
420 nm emission wavelength; acquisition start<br />
after 3 min from gradient start.<br />
35