ALARMSAir-in-Line- Push OK and then push RUN to advance small bubbles past the air detector. Eachpush of RUN advances approximately 0.1 mL. Use a syringe to aspirate air from thelower Y site or re-prime the set.Audio- May be silenced for 2 minutes by depressing any key.- Low, medium and high volume levels may be selected in the CONFIG screen.Depleted Battery- The battery is fully depleted and unable to run the pump. To continue the infusionand recharge the battery by plugging the pump’s AC power adaptor into an AC outlet.Confirm that the adaptor’s power cord connector is attached to the pump. Fullcharging requires a minimum of 8 hours for the Standard Battery and 12 hours for theWireless Battery Module.Door Not Fully Closed / Set Outside Channel- The pump’s door has not closed and latched correctly. Ensure the slide clamp isclosed, open the door using the slide clamp and re-load the IV set. Close the pump’sdoor ensuring both door latches shut securely.Door Open- The slide clamp has been closed and inserted in the keyhole when the pump wasrunning. The pump is stopped. Close the door, open the slide clamp, remove it fromthe keyhole and push RUN to restart the infusion OR open the door and unload theIV set.Downstream Occlusion- Eliminate a closed clamp, kinked tube, positional catheter, clotted catheter, cloggedIV filter or other sources of occlusion below the pump and the pump will restartautomatically.<strong>Infusion</strong> Complete- The VTBI (volume to be infused) has counted down to zero and has been delivered.The pump is running at a rate of 1.0mL/hr (KVO rate) – keep vein open (or the actualinfusion rate, whichever is lower). Push STOP to halt the KVO rate and return to theSetup Screen. Select a new VTBI value and push RUN.Inactive Alarm- The pump has been inactive for 2 minutes and no action has been taken. Follow theprompted action and resume or restart the pump by pushing Run.In Stop – Load Set- Load the IV set and push RUN.In Stop – Open Slide Clamp- Open the slide clamp, remove it from the keyhole and push RUN.In Stop – Push Run- Push RUN to begin the infusion.- 21 -

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!