SERVICINGCAUTION Electric shock hazard.There are no user serviceable parts. Do not remove the case. Refer servicing to SIGMA trainedand qualified service personnel. Refer to the Service <strong>Manual</strong> for inspection and maintenanceprocedures.To Return <strong>Pump</strong>s to SIGMA- Phone 1-800-356-3454 for a repair authorization (RA) number. A P.O. number fornon-warranty repairs is also required.- Ship pumps to SIGMA 711 Park Avenue, Medina, NY 14103- Include a problem description, contact person, phone number and return address.Label the shipping box with the RA number. Return pumps in original boxes, withoriginal inserts to prevent damage during shipment.Required Maintenance and Frequency- Maintenance consists of routine cleaning and annual performance evaluations asdescribed in the service manual.- <strong>Pump</strong>s suspected of being damaged must be tested for proper performance beforebeing returned to patient use. This includes pumps that have been physicallydamaged, dropped or those that have fluid intrusion.- 30 -