Calibration procedure for Modus Airflow alarm(Digital Differential Pressure Moni<strong>to</strong>r).Note: This procedure MUST be followed every time the damper width has been changed.Initial Set Point SettingsBalance the cabinet air flows <strong>to</strong> nominal operating set point per manufacturer instruction.( Seeinstruction label on front of cabinet).Record voltage <strong>to</strong> mo<strong>to</strong>r/blower (M1, M2 on speed control) and measure and record damper/ slotwidthMake certain that the Modus displays 0.000 when zero pressure is applied(This can be easilyaccomplished by removing the Tygon tubing from the hose barb fitting). Use the “zero” function ifnecessary.Re-connect the tygon tubing and make sure that the value displayed is negative(-) when the cabinet isrunning.Low Flow Alarm SettingPlace a DIM airflow hood on the front of the cabinet or use the 3 inch method for measuring inflow.Adjust cabinet <strong>to</strong> a 20% reduction in airflow by decreasing the voltage <strong>to</strong> mo<strong>to</strong>r. DO NOT adjust thedamperRecord the exhaust/ damper pressure on Modus (indication of exhaust plenum pressure at20% reduction) and voltage <strong>to</strong> mo<strong>to</strong>r/ blower (across M1, M2 on speed control)Set the Modus High Alarm set point <strong>to</strong> the 20% reduction in pressure recorded in step 14 (per Modusprogramming instructions below)Set the Modus dead band <strong>to</strong> .002 (per Modus programming instructions below).Return cabinet <strong>to</strong> Nominal Set Point voltage, recorded in step 2Check the CalibrationSet Point CheckWhile measuring the intake airflow with a DIM, reduce the airflow by decreasing the voltage <strong>to</strong>mo<strong>to</strong>r/ blower until a 20% reduction in flow is measured. Check that the low set point alarms at thisvalue. If not, reset Modus low alarm set point <strong>to</strong> alarm as desired, per instructions. Return cabinet<strong>to</strong> Set Point Voltage.Page 20 of 54

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!