Changing the Supply FilterThe <strong>BIOPROtect</strong> <strong>II</strong> requires two supply filters measuring 32” x 45” x 3” deep (consult test report forpart numbers and specifications). These filters are arranged so as <strong>to</strong> be serviced from inside the<strong>BIOPROtect</strong> <strong>II</strong> work area. Consequently you will have <strong>to</strong> remove the cart and any equipment from insidethe cabinet prior <strong>to</strong> changing the filters.To service the supply filters first swing the diffusers down allowing them <strong>to</strong> rest on the rear wall byremoving the ¼-20 screws at the front edge of the diffuser panel. Next using a 3/8” driver remove the ¼-20 hardware that secures the frame <strong>to</strong> the filter. This frame supports the filter.Prepare the new filters by putting a light coat of silicone grease on the face of the gasket. Clean thesealing flange thoroughly. Place the new filter in the center of the frame and install the frame in<strong>to</strong> thecabinet using the ¼-20 hardware.Conduct a filter leak check as described in the earlier chapter.Page 23 of 54