<strong>Paroscientific</strong>, <strong>Inc</strong>. Document No. 8360-002 Rev. K. 8/1/2002Command: M2Description: Read or set span adjustment coefficient for calculated humidity value.Command: *0100M2Response: *0001M2= ######Note: Changing this parameter effects the temperature calibration!Note that the adjusted output is expressed as H a = M2*(T + Z2); changing M2 willaffect the instrument’s adjusted zero offset if M2 is not 1.Command:E1,E2,F1,F2,G1,G2,H1,H2,K1,K2Description: Read or set temperature and humidity calibration coefficients.Channel 1: E1, F1, G1, H1, K1 (temperature)Channel 2: E2, F2, G2, H2, K2 (humidity)Example:Command: *0100E1Response: *0001E1=######Note: Changing these parameters effect the temperature and/or humidity calibration!MET3A Installation, Operation & Maintenance Guide Page 13