<strong>Paroscientific</strong>, <strong>Inc</strong>. Document No. 8360-002 Rev. K. 8/1/20028.5 Relative Humidity SensorThe humidity sensor uses a polymer capacitance element. It was selected for itsaccuracy, stability, and resistance to contamination. Its output is corrected fortemperature effects over a wide temperature range.The recovery time of the aspirated RH sensor is typically less than 2 minutes. Verylarge steps in humidity can take longer to reach full equilibrium. A step change of 75%RH settles within 5% of final value in 2 minutes and within 1% of final value in 5minutes. In addition, there can be hysteresis effects of a few percent after prolongedexposure to humidity above 90% RH.8.6 Relative Humidity CalibrationThe MET3A humidity sensor is a standard humidity probe characterized by themanufacturer. All MET3A units are conformance tested against a transfer standard toensure proper performance.MET3A Installation, Operation & Maintenance Guide Page 25