KEYPAD AND DISPLAY DESCRIPTIONSIn itialization Display The Initialization display enables you to set specific parameters for the survey toselect an instrument self test. This display is accessible from the Main Menu.HH:MM:SS <strong>12</strong>:00:00YY/MM/DD 91/06/24JOB #: 1024 COLE-COLE:OPERATOR: R. MILLER OMIT# SLICES: 0SERIAL #: 9105002TIMING: 2TEST: YES UNITS: METREHEATER: NODURATION: 10SYNC. CHANL: 1AUTO. REJ.: NO VAR. SLICES: NOFULL SAT. DET. YES SLICE WIDTHS: ->Mx 340-520Change ?BAT1: bbbHH:MM:SS - It is the time of day in a 24 hour format. The actual time that thesynchronization phase begins, is recorded with the data of each measurement.YY/MM/DD - It is the date which is recorded in the header of the data dump. Anew date generates a new header.JOB # - Is the number of the current survey project. It accepts up to 6 digits.OPERATOR - The operator's name or number. This parameter accepts up to 9characters.SERIAL # - It is the instrument serial number or any other identification numberwith a maximum of 9 digits.TEST - Enables or disables the internal test generator. See the Specificationssection for the nominal values.HEATER - Enables or disables the display heater however, if the temperature isabove -15°C, enabling this parameter has no effect.DURATION - Is the number of IP cyclesto be measured before the Autotermination(maximum 999 cycles).3-7