IPR-12 Manual - Scintrex
IPR-12 Manual - Scintrex
IPR-12 Manual - Scintrex
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
KEYPAD AND DISPLAY DESCRIPTIONSCLEAR MEMORY - This parameter is selected to clear the memory andrequires a (Y)es or (N)o answer. Once you clear the memory by selecting 'Y' allof the data is lost, except for the setup parameters.Signal-NoiseThe Signal-Noise Monitor display shows the signal of a selected dipoleM onitor Displayin order to help you troubleshoot if you encounter a problem. This display isavailable from the Main Menu.Figure 5: Example of Signal-Noise DisplayNotes:• The input signals are sampled at 50ms intervals and the data value of theselected dipole displayed as a pixel on the display. This enables you to observesignal, noise and SP drift for up to approximately <strong>12</strong> seconds at a time. The fullscale value can be changed in 1-2-5 steps. The SP can also be removed in orderto observe the small noise on top of a large SP.DIP. # - This is the number of the selected dipole for which the display applies.Full Scale - The full scale value (+xx) is the - vertical value of the box outline.You can adjust the value to between +1mV to +14V.3-11