SETTING UP THE <strong>IPR</strong>-<strong>12</strong>34. Toggle until No appears.Selecting The 35. Move the cursor to FULL SAT. DET.Full SaturationDetection36. Toggle until Yes appears.Setting 37. Move the cursor to TIMING.The Timing38. Select 2.Setting Mx 39. Move the cursor to Mx.40. Enter 340.41.Press the → key.42. Enter 520.Setting Cole- 43. Move the cursor to OMIT # SLICES.C ole Slices44. Enter 0.Setting 45. Move the cursor to UNIT.The Unit46. Toggle until Metre appears.The following is a typical Initialization Display:InitializationDisplayHH:MM:SS <strong>12</strong>:00:00YY/MM/DD 91/06/24JOB #:OPERATOR:1024R. MILLERCOLE-COLE:OMIT # SLICES: 0SERIAL #: 9105002TIMING: 2TEST: YES UNITS: MetreHEATER:DURATION:NO10S YNC . CHANL.:AUTO REJ.:1NO VAR. SLICES: NoFULL SAT. DET. YES SLICE WIDTHS: -->Mx 340 - 520Custom Slice WidthSetup DisplayChange ?47. Move the cursor to Variable SlicesBAT: bbb4-5