General Catalogue Chemistry - BECO Internacional LTDA.
General Catalogue Chemistry - BECO Internacional LTDA.
General Catalogue Chemistry - BECO Internacional LTDA.
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
667 349<br />
photometry<br />
Determination of the absorption characteristic<br />
of an untreated chlorophyll solution<br />
By simple means it is possible to produce an<br />
alcohol extract from crushed leaves. This untreated<br />
chlorophyll solution may then be subjected to<br />
photometric analysis.<br />
With the aid of the digital spectrophotometer<br />
(> 667 349) it is possible to record a continuous<br />
absorption spectrum; in spite of the presence<br />
of contamination the existence of absorption<br />
peaks in the blue and red ranges can be<br />
detected.<br />
Equipment list<br />
60 LEYBOLD DIDACTIC · GENERAL CATALOGUE CHEMISTRY<br />
667 352<br />
Determination of the absorption characteristic<br />
of an untreated chlorophyll solution<br />
Quantity Description Cat. No.<br />
1 Digital spectral photometer . . .667 349<br />
1 Computer connection cable<br />
with software . . . . . . . . . . . . . .667 352<br />
1 Personal computer<br />
667 349<br />
Digital spectral photometer<br />
The digital spectral photometer is an economical,<br />
easy-to-operate single beam photometer.<br />
The digital display allows the simultaneous indication<br />
of transmission and absorption or concentration<br />
and factor.<br />
The wavelength can also be indicated in both<br />
cases.The filter wheel contains 4 filters each of<br />
which covers a different range of wavelengths.<br />
Specific wavelengths are selected manually<br />
after the appropriate filter has been positioned<br />
in the beam.<br />
The integrated interface allows connection to<br />
the RS 232 interface of a computer. The software<br />
which is accompanying the computer<br />
connection cable (> 667 352 ) allows the<br />
recording of the absorption as a factor of the<br />
concentration, time or wavelength.<br />
The function of a large digital display can be<br />
called up on the monitor.<br />
Wavelength range: 340...900 nm<br />
Resolution: 2.5 nm<br />
Wavelength accuracy: ± 1.0 nm<br />
Photometric stability: ± 0.1 %T<br />
Photometric linearity: ± 0.1 %T<br />
Measurement ranges:<br />
Transmission: 0...100 %<br />
Absorption: 0...1.999<br />
Concentration: 0...199<br />
Factor: 0.1...999<br />
Display:<br />
LED, 4-digit, 7 mm high<br />
Cells:<br />
Round cells,<br />
15 mm external diameter,<br />
100 mm high<br />
Cell holder: 15 mm<br />
Data output: Serial, 9-pin<br />
Voltage supply:<br />
115/230V, 50/60 Hz, switchable<br />
Dimensions: 410 x 210 x 220 mm<br />
Weight: 5.8 kg<br />
Scope of delivery<br />
1 Digital spectral photometer,<br />
including RS 232 interface<br />
2 Round cells<br />
667 352<br />
Computer connection cable<br />
This cable allows the digital spectral photometer<br />
(> 667 349 ) to be connected directly to<br />
the serial interface of a computer. DOS software<br />
is also included in the scope of supply