segments. (Most transmission targets are longer.) For scanners specified as having Dmax greater than 3, the charts of choice are<br />

the Stouffer Transmission Step Wedge T4110 or Danes-Picta TS28D, which are too large to be tested easily with<br />

35mm film scanners.<br />

Follow the remainder of instructions in Photographing the chart and running Stepchart, above. Be sure to select the<br />

correct chart type from the Stepchart input dialog box (right).<br />

The <strong>Imatest</strong> algorithm for finding dynamic range is remarkably accurate. <strong>Imatest</strong> detects chart zones using the smallest density step<br />

that results in uniformly spaced detected zones (see Algorithm). For smaller steps, noise can be mistaken for zone boundaries. For<br />

larger steps, fewer zones are detected.<br />

The dynamic range is the difference in density between the zone where the pixel level is 98% of its maximum value (250 for 24-bit<br />

color, where the maximum is 255), estimated by interpolation, and the darkest zone that meets the measurement criterion. The<br />

repeatability of this measurement is better than 1/4 f-stop.<br />

Here is a result for the Canon EOS-10D at ISO 400, converted from RAW format with Capture One LE.<br />

EOS 10D, ISO 400, Converted with Capture One LE<br />

276 of 451