OPERATOR'S MANUAL FOR THE SYSTEM<br />
3.5 CONNECT SYSTEM<br />
1) Connect the <strong>system</strong> together with all cables (see Figure 3-3: Equipment Set-up for<br />
a Wireless Connection, below).<br />
NOTE: If you are using the <strong>wireless</strong> connection, DO NOT connect the READER-PC<br />
CABLE.<br />
2) Prior to first use, place the Reader on Fast Charge. Turn the mobileMOSFET<br />
OFF and connect the universal power adapter for the Reader to the “POWER IN”<br />
jack on the back of the Reader, then plug the country specific line cord into the AC<br />
outlet. The red Fast Charge LED will go on. When the Fast Charge LED goes off<br />
and the green Trickle Charge LED goes on, the battery is fully charged. This will<br />
take a maximum of three hours.<br />
3) Power on the <strong>system</strong> and allow the electronics to warm up for at least fifteen (15)<br />
minutes before proceeding. If the green Power LED is not flashing, the power<br />
switch must be switched up to the momentary position to reconnect the battery<br />
power.<br />
3.6 CONNECT PC<br />
3.6.1 Wireless Connection<br />
Mount the TN-configured Bluetooth <strong>wireless</strong> transceiver on an interior wall of the<br />
treatment room. Attach the 5V DC power supply, and connect the transceiver to the<br />
PC’s COM port using the RS-232 cable (See Figure 3-3: Equipment Set-up for a<br />
Wireless Connection).<br />
Figure 3-3: Equipment Set-up for a Wireless Connection<br />
Page 20 of 62