OPERATOR'S MANUAL FOR THE SYSTEM<br />

5.3.2 Linear 5ive MOSFET Array Dosimeter<br />

For additional information, see Technical Note #8.<br />

Real-time evaluation of the quality of the implant or HDR dose delivery can be provided by<br />

the in-vivo use of a MOSFET 5ive Linear Array Dosimeter (models TN-252LA5, TN-502LA5,<br />

or TN-1002LA5). These measurements will allow immediate assessment of post-implant<br />

base and apex dose coverage, as well as the dose to organs at risk such as the urethra,<br />

rectum or bladder.<br />

Standard MOSFET Dosimeters, which have one dosimeter located at the tip of the flex, are<br />

linear with dose, with a 5% maximum variation of the calibration factor throughout the entire<br />

20,000mV reading range of the MOSFET. The initial calibration factor can be used for all<br />

measurements, without a need for further calibration or correction.<br />

In the case of the Linear 5ive Array, linearity corrections (LC) are implemented automatically<br />

in the mobileMOSFET Dose Reporter Software at doses above 5000cGy. This is run ‘in the<br />

background’ by the software and is visible to the user in the LC column. These linearity<br />

corrections will remain accurate throughout the 20,000 mV reading range of the dosimeter.<br />

For this reason it is very important that the user designate the correct type of Linear Array<br />

being used as well as the Bias selection (Standard or High) to allow the software to properly<br />

calculate and correct the CFs.<br />

Figure 5-5: MOSFET Linear 5ive Array<br />

Dimensions: 1.5mm wide, 46cm long, 1.3mm thick, 2cm Inter-MOSFET Spacing.<br />

The Radio-Opaque Marker located at tip of flex, 7mm away from<br />

the center of the first MOSFET on the array.<br />

Page 46 of 62