OPERATOR'S MANUAL FOR THE SYSTEM<br />

E. INITIALIZE the MOSFETs by clicking on the START button on the screen.<br />

F. Irradiate.<br />

G. AFTER irradiation, read the MOSFETs ----- click on the READ button to obtain<br />

dose information.<br />

Figure 4-7: Dose Measurement Screen<br />

4.5 LINEAR COMPENSATION FOR ARRAYS<br />

On the Dose Measurement screen there is a column entitled "LC". This column is<br />

used to display the "Linear Compensation" factors for TN's Linear Array dosimeters.<br />

If a single MOSFET is being used (i.e. TN-502RD), no compensation is required and<br />

"N/A" will be displayed in the LC column.<br />

A Linear Array, however, requires compensation. If a voltage difference is detected<br />

during the Start-Read period, the mobileMOSFET Software will calculate the<br />

corresponding LC number based on the Array in use and the detected difference.<br />

If the type of Array has not been designated, a question mark, "?", will appear in the<br />

LC column to indicate that the Array has not been set. To do this, return to the<br />

Configuration Window in the Equipment Setup screen and select the type of array you<br />

are using. The compensation will automatically be applied to the final doses.<br />