OPERATOR'S MANUAL FOR THE SYSTEM<br />

4.2.2 Dosimeter Selection<br />

The next step in the Configuration window is to designate the type of dosimeter that<br />

will be used in the treatment session, e.g. the standard single MOSFET, (one<br />

dosimeter on a flex), or a Linear 5ive Array dosimeter (five dosimeters on one flex).<br />

This feature is most important to those who wish to use the Linear Array, for reasons<br />

outlined in “Section 5.3.2 Linear 5ive MOSFET Array Dosimeter”.<br />

This feature is found at the bottom of the Configuration window (see Figure 4-4:<br />

Dosimeter Selection Menu below) with the following message: “Please select the<br />

type of dosimeter used by Reader #X”.<br />

The software provides the option of choosing from: Single MOSFET, TN-252LA5<br />

Array, TN-502LA5 Array, or TN-1002LA5 Array.<br />

Figure 4-4: Dosimeter Selection Menu<br />

4.2.3 Calibration Parameters (CFs and CRs):<br />

These fields are used to show Calibration Parameters, including Calibration Factors<br />

(CFs) and Correction Factors (CRs). A CF translates the MOSFET’s voltage reading<br />

into a dose unit (i.e. Gy, cGy, R), and a CR is a multiplicative factor that can be used<br />

to accommodate certain external factors (such as extrapolating partial dose depth<br />

data to Dmax).<br />

Page 28 of 62