25.02.2013 Views

operator's manual wireless dosimetry system - Best Medical Canada

operator's manual wireless dosimetry system - Best Medical Canada

operator's manual wireless dosimetry system - Best Medical Canada

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

OPERATOR'S MANUAL<br />

FOR THE<br />

WIRELESS DOSIMETRY SYSTEM<br />

For use with mobileMOSFET Software Versions 2.2 and above<br />

FDA LISTING: mobileMOSFET Wireless Dosimetry System listed as<br />

Patient Dose Verification System, 510(k) #K041557<br />

HEALTH CANADA: <strong>Medical</strong> Device License No. 68061


OPERATOR'S MANUAL FOR THE SYSTEM<br />

WARRANTY<br />

<strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd. warrants the mobileMOSFET Model TN-RD-70-W, dosimeters and<br />

cables to be free from defects in material and workmanship for a period of one year from the date of<br />

shipment.<br />

This warranty does not apply to defects resulting from unauthorized modification or misuse of any<br />

product or part, or to battery replacement or damage caused by battery leakage. Dosimeters<br />

damaged due to excessive mishandling (examples include but are not limited to the flex being<br />

strongly kinked or cut, or the epoxy bulb being cracked) will not be covered under warranty. Please<br />

see Section 5.4 for more information on proper handling of dosimeters.<br />

Please see the mobileMOSFET Software License Agreement for software support provided.<br />

During the warranty period <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd will, at our option, either repair or replace<br />

within 30 days any product that proves to be defective. Repaired or replaced products are<br />

warranted for the balance of the original warranty period or at least 90 days.<br />

To exercise the warranty or obtain service, write or call the local <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd<br />

representative, or contact <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd headquarters at the address and phone<br />

number shown below:<br />

<strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd.<br />

25B Northside Road<br />

Ottawa, ON K2H 8S1 <strong>Canada</strong><br />

Tel: (613) 596-4563 Fax: (613) 596-5243<br />

Toll Free (North America only) 1-877-668-6636<br />

www.Team<strong>Best</strong>.com<br />

bmcinfo@Team<strong>Best</strong>.com<br />

* In no event shall <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd. be liable for any damages whatsoever arising<br />

out of the use of or inability to use this product.<br />

PRODUCT PATENTS<br />

U.S. Patent Numbers 6,614,025<br />

5,444,254<br />

4,678,916<br />

5,117,113<br />

6,650,930<br />

European (U.K. & DE) Patent Numbers EP 0575800<br />

EP 0471957<br />

Japanese Patent Number 3012364<br />

Page 2 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

mobileMOSFET Software END USER LICENSE AGREEMENT<br />

PLEASE READ CAREFULLY – The mobileMOSFET Software that comes with this<br />

<strong>system</strong> is a product of <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd. and may not be reproduced. You<br />

may only install and use one copy of the product. If you have need for multiple<br />

copies, you must enter into a specific MULTIPLE END USER LICENSE<br />

AGREEMENT with <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd.<br />

<strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd. reserves the right to make changes and improvements to<br />

its products without providing notice.<br />

CONNECTBLUE AB OEM BLUETOOTH*<br />

cB-0090-01 and cB-0701-01<br />

FCC COMPLIANCE STATEMENT<br />

This device complies with Part 15 of the FCC Rules. Operation is subject to the<br />

following two conditions: (1) this device may not cause harmful interference and (2)<br />

this device must accept any interference received, including interference that may<br />

cause undesired operation.<br />

*© connectBlue AB 2000-2003. BLUETOOTH is a trademark owned by the Bluetooth SIG Inc.<br />

MANUFACTURER’S EU REPRESENTATIVE<br />

Emergo Europe<br />

Molenstraat 15, 2513 BH<br />

The Hague<br />

The Netherlands<br />

NOTE: As of January 1, 2007, the name of the manufacturer "Thomson & Nielsen<br />

Electronics Ltd." changed to "<strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd." Any future reference to Thomson<br />

& Nielsen Electronics Ltd. in product literature, artwork or instrumentation is to be deemed<br />

as being <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd.<br />

Document #101735.08 March 2007<br />

Page 3 of 62<br />

0120


OPERATOR'S MANUAL FOR THE SYSTEM<br />

EXCLUSIVE USER WEB-SITE<br />

<strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd welcomes you to an exclusive group of<br />

professionals who believe that MOSFET dosimeters are the best <strong>dosimetry</strong><br />

device for every application. In an effort to help you easily access all of the<br />

available support tools, we offer a web-site that is open to MOSFET<br />

Dosimeter Users only. It contains many helpful features, including Technical<br />

Support, Lists of Publications and Technical Notes, the latest Operator<br />

Manual updates and News Updates, such as User Group Meeting<br />

announcements. We encourage you to contact us for your password to enter<br />

this special section of our web-site.<br />

www.Team<strong>Best</strong>.com<br />

e-mail: bmcinfo@Team<strong>Best</strong>.com<br />

Toll Free: 1-877-668-6636 (N. America)<br />

Ph: (613) 596-4563<br />

Page 4 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

CONTENTS<br />

1 GENERAL INFORMATION .....................................................................................................8<br />

1.1 THIS MANUAL ............................................................................................................8<br />

1.2 INTRODUCTION.........................................................................................................8<br />

1.3 mobileMOSFET FEATURES.......................................................................................9<br />

1.4 SAFETY PRECAUTIONS............................................................................................9<br />

1.5 REFERENCES............................................................................................................9<br />

2 SYSTEM DESCRIPTION......................................................................................................10<br />

2.1 GENERAL .................................................................................................................10<br />

2.2 mobileMOSFET COMPONENTS ..............................................................................11<br />

2.3 OVERVIEW OF mobileMOSFET SOFTWARE FEATURES .....................................12<br />

3 SYSTEM SET-UP .................................................................................................................17<br />

3.1 SYSTEM REQUIREMENTS AND SETUP ................................................................17<br />

3.2 BATTERY OPERATION............................................................................................18<br />

3.3 CONFIGURE BIAS SENSITIVITY.............................................................................18<br />

3.4 CONNECT DOSIMETERS ........................................................................................19<br />

3.5 CONNECT SYSTEM .................................................................................................20<br />

3.6 CONNECT PC...........................................................................................................20<br />

3.6.1 Wireless Connection......................................................................................20<br />

3.6.2 Cable Connection ..........................................................................................21<br />

3.6.3 LAN/Wireless Connection..............................................................................22<br />

3.7 INSTALLING mobileMOSFET SOFTWARE..............................................................24<br />

4 OPERATING PROCEDURE .................................................................................................25<br />

4.1 EQUIPMENT SET-UP (Screen 1) .............................................................................25<br />

4.2 EQUIPMENT SETUP: CONFIGURATION ................................................................26<br />

4.2.1 Connect To Reader Module...........................................................................26<br />

4.2.2 Dosimeter Selection.......................................................................................28<br />

4.2.3 Calibration Parameters (CFs and CRs): ........................................................28<br />

4.3 Treatment information (Screen 2)..............................................................................30<br />

4.3.1 Image Selection.............................................................................................31<br />

4.3.2 Using Patients' Photographs During Measurement.......................................32<br />

4.3.3 Templates – A Time Saving Feature .............................................................32<br />

4.3.4 Dose Report Files: A Quick Template Alternative .........................................33<br />

4.4 DOSE MEASUREMENT (Screen 3)..........................................................................33<br />

4.5 LINEAR COMPENSATION FOR ARRAYS ...............................................................34<br />

4.6 READING DOSIMETERS INDIVIDUALLY................................................................35<br />

4.7 REMOVING DOSIMETERS ......................................................................................35<br />

4.8 DOSIMETRY REPORT (Screen 4) ...........................................................................35<br />

4.9 SEQUENTIAL READING ..........................................................................................37<br />

Page 5 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

4.10 SIMULTANEOUS TREATMENT SESSIONS............................................................ 39<br />

4.11 Message History Txt Files ......................................................................................... 40<br />

4.12 Message Windows .................................................................................................... 40<br />

5 DESCRIPTION OF SYSTEM COMPONENTS ..................................................................... 41<br />

5.1 mobileMOSFET READER MODULE......................................................................... 41<br />

5.1.1 Reader Side Panel Layout............................................................................. 42<br />

5.1.2 Switches and LEDs ....................................................................................... 42<br />

5.1.3 Reader Back Panel Layout............................................................................ 43<br />

5.2 RECOMMENDED DOSIMETER AND BIAS SENSITIVITY COMBINATIONS FOR<br />

APPLICATIONS IN RADIOTHERAPY AND RADIOLOGY ....................................... 44<br />

5.3 DOSIMETERS........................................................................................................... 45<br />

5.3.1 Single MOSFET Dosimeters ......................................................................... 45<br />

5.3.2 Linear 5ive MOSFET Array Dosimeter .......................................................... 46<br />

5.3.3 Linear 5ive ARRAY: CLINICAL IMPLEMENTATION AND DOSE SAMPLING47<br />

5.4 WIRELESS TRANSCEIVER ..................................................................................... 50<br />

6 mobileMOSFET SYSTEM MESSAGES................................................................................ 51<br />

7 CALIBRATION PROCEDURE .............................................................................................. 52<br />

7.1 CALIBRATION METHODS FOR RADIOTHERAPY USING THE mobileMOSFET<br />

SOFTWARE AIDED CALIBRATION TOOL .............................................................. 53<br />

7.2 CALIBRATION OF MOSFET AND LINEAR 5ive ARRAY MOSFET DOSIMETERS:<br />

GUIDELINES FOR DIFFERENT ENERGIES (e.g. BRACHYTHERAPY) ....................... 55<br />

7.3 BUILD-UP & PARTIAL BUILD-UP ............................................................................ 57<br />

7.4 CALIBRATION FOR DIAGNOSTIC X-RAY ENERGIES ........................................... 57<br />

7.5 TESTING A DOSIMETER FOR FADE (CHANGE IN PERFORMANCE).................. 58<br />

8 MAINTENANCE ....................................................................................................................59<br />

8.1 mobileMOSFET BATTERY MAINTENANCE AND REPLACEMENT........................ 59<br />

8.2 ORDERING REPLACEMENT PARTS ...................................................................... 60<br />

8.3 TECHNICAL SPECIFICATIONS ............................................................................... 61<br />

Page 6 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

LIST OF FIGURES<br />

FIGURE 3-1: MOBILEMOSFET READER MODULE STATUS PANEL............................................18<br />

FIGURE 3-2: SYSTEM CONFIGURATION ......................................................................................19<br />

FIGURE 3-3: EQUIPMENT SET-UP FOR A WIRELESS CONNECTION ........................................20<br />

FIGURE 3-4: MULTIPLE READERS CONNECTED VIA WIRELESS TO A PC...............................21<br />

FIGURE 3-5: VARIOUS WIRED SCENARIOS FOR CONNECTING READERS TO A PC..............21<br />

FIGURE 3-6: LAN WIRELESS CONNECTION.................................................................................22<br />

FIGURE 3-7: MULTIPLE SYSTEMS AND TREATMENT ROOMS...................................................23<br />

FIGURE 3-8: SINGLE SYSTEM/MULTIPLE TREATMENT ROOMS ...............................................23<br />

FIGURE 3-9: WINDOWS “RUN” POP-UP DIALOG BOX .................................................................24<br />

FIGURE 4-1: EQUIPMENT SET-UP SCREEN.................................................................................25<br />

FIGURE 4-2: READER CONFIGURATION WINDOW .....................................................................27<br />

FIGURE 4-3: CONNECTING TO A MOBILEMOSFET READER .....................................................27<br />

FIGURE 4-4: DOSIMETER SELECTION MENU ..............................................................................28<br />

FIGURE 4-5: TREATMENT INFORMATION SCREEN ....................................................................30<br />

FIGURE 4-6: PATIENT IMAGE WINDOW........................................................................................31<br />

FIGURE 4-7: DOSE MEASUREMENT SCREEN .............................................................................34<br />

FIGURE 4-8: DOSIMETRY REPORT SCREEN ...............................................................................37<br />

FIGURE 4-9: SEQUENTIAL READING SCREEN ............................................................................38<br />

FIGURE 4-10: SEQUENTIAL READING SETTINGS .......................................................................39<br />

FIGURE 4-11: MULTIPLE VERIFICATION SESSIONS ...................................................................40<br />

FIGURE 5-1: MODEL TN-RD-70-W TOP PANEL ............................................................................41<br />

FIGURE 5-2: MODEL TN-RD-70-W SIDE PANEL ...........................................................................42<br />

FIGURE 5-3: MODEL TN-RD-70-W BACK PANEL ..........................................................................43<br />

FIGURE 5-4: SYSTEM CONNECTIONS ..........................................................................................43<br />

FIGURE 5-5: MOSFET LINEAR 5IVE ARRAY .................................................................................46<br />

FIGURE 5-6: MOSFET LINEAR 5IVE ARRAY BEFORE INSERTION IN THE 3-WAY CATHETER48<br />

FIGURE 5-7: WIRELESS TRANSCEIVER .......................................................................................50<br />

FIGURE 5-8: EXTENSION OF TRANSCEIVER-READER CABLE ..................................................51<br />

FIGURE 7-1: CALIBRATION WINDOW............................................................................................53<br />

FIGURE 8-1: BATTERY PACK PLACEMENT ..................................................................................59<br />

Page 7 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

1 GENERAL INFORMATION<br />

1.1 THIS MANUAL<br />

The information in this <strong>manual</strong> is provided in eight sections and is intended for use<br />

with mobileMOSFET Software Versions 2.2 and above. After a general introduction,<br />

the instructions are laid out in the order of operation, with SYSTEM DESCRIPTION<br />

and SYSTEM SET-UP instructions followed by OPERATING PROCEDURE. This is<br />

followed by reference sections describing the System Components, Calibration and<br />

other key information.<br />

1.2 LABELING<br />

The following table defines the labeling symbols that may be found on certain<br />

mobileMOSFET <strong>system</strong> components and/or documentation.<br />

Manufacturer<br />

Date of manufacture (with date format of YYYY-MM)<br />

Authorized Representative for the European Union<br />

Consult Instructions<br />

SN Serial Number<br />

1.3 INTRODUCTION<br />

mobileMOSFET, model TN-RD-70-W, is <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd's mobile, battery<br />

operated dose verification <strong>system</strong> intended for use in applications such as External<br />

Beam and Brachytherapy radiotherapy to verify treatment planning, and Radiology to<br />

measure dose uniformity and radiation exposure to the patient.<br />

The System consists of PC-based user-interface software and one or more<br />

mobileMOSFET Reader Modules. The Reader Module is small, lightweight, and<br />

battery-operated. It hosts the MOSFET dosimeters and realizes the measurement<br />

circuitry. The Software provides a console on screen for the operator to perform all<br />

required actions in dose measurements, such as reading, displaying, saving and<br />

printing results.<br />

The mobileMOSFET Remote Monitoring Dose Verification Software can be used to<br />

perform five treatment sessions simultaneously and each session can read up to eight<br />

readers. Note: Throughout this Manual, the term "Reader" refers to the<br />

mobileMOSFET Reader Module.<br />

Page 8 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

The advantages of this <strong>system</strong> include:<br />

• Reader portability through battery operation.<br />

• Readers can be connected to a controlling PC through a <strong>wireless</strong> connection or by an<br />

RS-232 cable.<br />

• Each Reader can read up to 5 dosimeters or a single MOSFET Linear Array. Eight<br />

Readers can be simultaneously used during a single procedure if required, allowing for<br />

a total of 40 dosimeters.<br />

• If LAN-to-Serial adapters are employed together with the <strong>wireless</strong> transceivers, the<br />

distance between the PC and the Readers is greatly extended.<br />

1.4 mobileMOSFET FEATURES<br />

• Battery Operated – The NiMH power pack will allow for more than 20 hours of<br />

continuous use, and allows for considerable mobility in treatment rooms.<br />

• PC Connection – The Reader Module is connected to a PC to output the data from the<br />

mobileMOSFET Reader Module via the mobileMOSFET Software program model TN-<br />

RD-75, which allows the user to generate a report for the patient’s record.<br />

• Software Controlled Operation - using mobileMOSFET Remote Monitoring Dose<br />

Verification Software<br />

• Direct Reading – Software readout in dosimeter voltage (mV)<br />

• High Sensitivity - Depending on the Bias Sensitivity setting, the dosimeter and the<br />

energy of interest used, dosimeter response ranges up to ≅35mV/cGy<br />

1.5 SAFETY PRECAUTIONS<br />

• Dosimeters supplied are intended for external application to intact skin or insertion into a<br />

catheter suitable for the use intended. Contact of dosimeters or other parts of the <strong>system</strong><br />

with breached or compromised skin, bodily fluids or mucous membranes, is not<br />

recommended. Should you have any questions, please contact <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd.<br />

• Material in the dosimeter may, on contact with skin, cause mild sensitization in some<br />

persons.<br />

• This System is intended for use by qualified personnel who recognize radiation hazards<br />

associated with the use of radiation therapy and radiology equipment, and who exercise<br />

appropriate safety measures.<br />

• The <strong>system</strong> should only be used with the components listed in 2.2 below.<br />

• The System contains no operator serviceable parts except for the rechargeable batteries in<br />

the unit.<br />

1.6 REFERENCES<br />

Publications and technical notes describing the principles of operation of MOSFET dosimeters<br />

and their applications are available on request from <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd or visit our<br />

website at www.Team<strong>Best</strong>.com<br />

.<br />

Page 9 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

2 SYSTEM DESCRIPTION<br />

2.1 GENERAL<br />

The System is compact and easily transportable within a typical Radiotherapy or<br />

Radiology treatment clinic or other laboratory. The dosimeters are placed on the<br />

patient and measured by the Reader before and after irradiation. All data is output via<br />

the mobileMOSFET Remote Monitoring Dose Verification Software, provided with<br />

the <strong>system</strong> to control the Reader and generate a report for the patient’s record.<br />

In operation, the mobileMOSFET Reader is set up in the treatment room and<br />

connected to the required number of Dosimeters. The maximum number of MOSFET<br />

dosimeters is five per Reader. Alternatively, one can connect a single MOSFET<br />

Linear Array to a mobileMOSFET Reader. The mobileMOSFET Software can support<br />

up to eight Readers simultaneously during a single procedure.<br />

The measurement procedure is controlled by a PC using the mobileMOSFET<br />

Software through the RS-232C cable or <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd provided <strong>wireless</strong><br />

transceiver (BMC Part No. TN-RD-38). MOSFET dosimeters or the MOSFET Linear<br />

Array can be taped to the patient or placed in a phantom. The dose measurements<br />

are carried out by clicking the “START” and “READ” buttons on the Dose<br />

Measurement Screen.<br />

In applications, the mobileMOSFET Software can:<br />

• Handle up to five (5) verification sessions simultaneously. Operations in each<br />

session are totally independent.<br />

• For each verification session, up to eight (8) Reader modules (or up to 40<br />

dosimeters/8 Linear Arrays) can be used. Operations on each Reader<br />

module (i.e. Start or Read) can be performed collectively or individually.<br />

• For each Reader module, all dosimeters will be measured at the same time,<br />

however, measurement on individual dosimeters can also be easily<br />

performed (see Section 4.6).<br />

The following section contains a detailed description of the System components.<br />

For setting up, operating and calibrating the System, see Sections 3.0, 4.0 and 7.0,<br />

respectively.<br />

Page 10 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

2.2 mobileMOSFET COMPONENTS<br />

TN-RD-70-W, mobileMOSFET Wireless Dosimetry System includes the following:<br />

TN-RD-16 - Reader Module<br />

TN-RD-38 - Bluetooth Wireless Transceiver, + 6ft./2m Cable<br />

TN-RD-48 - 50 ft./15m RS-232 Cable (PC to Transceiver connection)<br />

TN-RD-75 - mobileMOSFET Remote Monitoring Dose Verification Software,<br />

One (1) License<br />

TN-RD-77 - <strong>Medical</strong> Universal Power Adapter for mobileMOSFET<br />

TN-502RD - MOSFET Dosimeters (Standard Sensitivity)<br />

Optional:<br />

TN-502RD - MOSFET Dosimeters (Standard Sensitivity)<br />

TN-502RDM - microMOSFET Dosimeters (Standard Sensitivity)<br />

TN-502RD-H - MOSFET Dosimeters (Standard Sensitivity) with Heat Shrink<br />

Reinforcement<br />

TN-502RD-HRO - MOSFET Dosimeters (Standard Sensitivity) with Radio-Opaque<br />

Marker<br />

TN-1002RD - High Sensitivity MOSFET Dosimeters<br />

TN-1002RDM - High Sensitivity microMOSFET Dosimeters<br />

TN-1002RD-H - High Sensitivity MOSFET Dosimeters with Heat Shrink<br />

Reinforcement<br />

TN-1002RD-HRO - High Sensitivity MOSFET Dosimeters with<br />

Radio-Opaque Marker<br />

TN-252LA5 - Standard Sensitivity Linear 5ive MOSFET Array<br />

TN-502LA5 - High Sensitivity Linear 5ive MOSFET Array<br />

TN-1002LA5 Extreme Sensitivity Linear 5ive MOSFET Array<br />

TN-RD-56-0.63 - 0.63 cm Brass Hemispherical Build-up Cap<br />

TN-RD-55-1.5 - 1.5 cm Hemispherical Build-up Cap<br />

TN-RD-38/40 - LAN (Local Area Network) and Bluetooth Wireless<br />

Transceiver Package<br />

TN-RD-38 - Bluetooth Wireless Transceiver, 6 ft./2m RS-232 Connection<br />

Cable<br />

To set-up the <strong>system</strong>, see Section 3.0 (Please note, you will need the Reader’s<br />

Serial Number.)<br />

Page 11 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

2.3 OVERVIEW OF mobileMOSFET SOFTWARE FEATURES<br />

This program catalogs its functions into five groups:<br />

1. Equipment Setup<br />

2. Treatment Information<br />

3. Dose Measurement<br />

4. Dosimetry Report<br />

5. Calibration Tool<br />

Five display/control screens let users access these function groups.<br />

The following is an overview of these screens.<br />

1. Equipment Setup<br />

• Select Reader Count<br />

• Assign Readers<br />

• Change Calibration Factors (CFs) and Correction Factors (CRs)<br />

• Check Status<br />

(See Section 4.2 for a more detailed description of equipment set-up.)<br />

Page 12 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

2. Treatment Information<br />

• Input patient ID, treatment settings, and equipment information<br />

• Assign dosimeters to patient(s), set target dose and indicate dosimeter positions<br />

on the chosen image<br />

• Use the program’s standard images or import your own (BITMAP, JPEG or GIF)<br />

• Use templates to save time<br />

Page 13 of 62


3. Dose Measurement<br />

OPERATOR'S MANUAL FOR THE SYSTEM<br />

• Click on the START button to initialize before IRRADIATION<br />

• After irradiation, click on the READ button to obtain dose measurements<br />

• Dose Data will be recorded and displayed on-screen & percentage deviation<br />

from Target Dose will be calculated automatically<br />

Page 14 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

4. Dosimetry Report<br />

In the last step of the measurement procedure, Dosimetry Report, you can:<br />

• Review your information, add in comments & save them<br />

• Export your data as Microsoft Word® Excel® (tab-delimited files) documents or<br />

Plain Text Files<br />

• Edit saved reports<br />

Page 15 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

5. Calibration Tool<br />

• Will guide you through the Calibration Process<br />

• Practice with the Calibration Simulation feature<br />

• Save multiple Calibration Factors (CFs)<br />

• Save multiple Correction Factors (CRs)<br />

Page 16 of 62


3 SYSTEM SET-UP<br />

OPERATOR'S MANUAL FOR THE SYSTEM<br />

3.1 SYSTEM REQUIREMENTS AND SETUP<br />

Computer:<br />

• PC: with Windows®* Operating Systems 98/2000/XP.<br />

• Mouse: PC Compatible.<br />

• Wireless Transceiver or Cable: A Bluetooth serial port adapter or RS-232 cable<br />

can be used to connect the PC to the mobileMOSFET Reader<br />

• Monitor: any color monitor. This program can be used with screen resolution of 800<br />

x 600 or higher. For the best effect, we strongly recommend that the screen<br />

resolution be set to 1024 x 768.<br />

• Printer: any laser printer or inkjet printer. The recommended paper sizes are<br />

"Letter" (8.5” x 11”) or "A4". If the paper is too narrow, the reports may not print out<br />

clearly.<br />

• One (1) Standard Nine (9) Pin Male RS-232 ComPort<br />

Reader Module:<br />

• mobileMOSFET Patient Dose Verification System Reader (Model TN-RD-70-W)<br />

• 100 � 240vAC Power Adapter (Note: The Reader is DC)<br />

• Wireless Bluetooth, or RS-232 Cable to COM port<br />

Dosimeters:<br />

Must be one of:<br />

• TN-502RD Standard Sensitivity MOSFET Dosimeters<br />

• TN-502RDM Standard Sensitivity microMOSFET Dosimeters<br />

• TN-1002RD High Sensitivity MOSFET Dosimeters<br />

• TN-1002RDM High Sensitivity microMOSFET Dosimeters<br />

• TN-252LA5 Standard Sensitivity Linear 5ive MOSFET Array<br />

• TN-502LA5 High Sensitivity Linear 5ive MOSFET Array<br />

• TN-1002LA5 Extreme Sensitivity Linear 5ive MOSFET Array<br />

Build-up Caps:<br />

• TN-RD-56-0.63 0.63 cm Brass Hemispherical Build-up Cap<br />

• TN-RD-55-1.5 1.5 cm Hemispherical Build-up Cap<br />

For more specifications on the MOSFETs, contact <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd directly for<br />

Technical Note #7, "MOSFET Dosimeter Specifications" or visit our website at<br />

www.Team<strong>Best</strong>.com<br />

* Windows® Operating Systems 98/2000/XP, Windows XP are registered trademarks of MICROSOFT<br />

Corporation.<br />

Page 17 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

3.2 BATTERY OPERATION<br />

The mobileMOSFET is intended to operate on battery power. The battery level for the<br />

mobileMOSFET Reader Module is indicated on the Equipment Setup and Dose<br />

Measurement screens, as well as windows that will pop-up advising you if your<br />

battery level is getting low. There are two methods to recharge the batteries, Fast<br />

Charge and Trickle Charge. The charging status is visible by the LEDs on the side of<br />

the Reader. A green LED indicates that the unit is in Trickle Charge, while a red LED<br />

indicates the Reader is in a Fast Charge.<br />

To Fast Charge the mobileMOSFET Reader Module, ensure the unit is turned OFF<br />

prior to connecting the unit to a power source. A Fast Charge will completely charge a<br />

mobileMOSFET with a drained battery in approximately three hours. Once the<br />

batteries are charged, the red LED will go out and the green LED will come on,<br />

indicating Trickle Charge.<br />

Trickle Charge supplies the battery with power at a continuous low rate thereby<br />

keeping it charged. Trickle Charge allows the user to continue treatment sessions<br />

when the Reader's battery is running low. Instead of having to be concerned with the<br />

battery losing power, the user can just continue on with the session. To use Trickle<br />

Charge, the mobileMOSFET must be ON and the power supply connected to the<br />

Reader.<br />

Prior to first use, place the Reader on Fast Charge (mobileMOSFET turned OFF<br />

and power connected) until the red (Fast Charge) LED goes OFF and the green<br />

(Trickle Charge) LED comes on. This will take a maximum of three hours.<br />

Figure 3-1: mobileMOSFET Reader Module Status Panel<br />

3.3 CONFIGURE BIAS SENSITIVITY<br />

The Bias located inside the Reader provides regulated voltage to the MOSFET<br />

dosimeters. This ensures reproducibility and linearity over the lifetime of the<br />

MOSFET. The mobileMOSFET Reader provides a choice of two Bias sensitivities in<br />

order to cover a wide range of doses with optimum reproducibility. The <strong>system</strong> allows<br />

you to choose between two options: Standard (STD) for routine use, or High (HI)<br />

Sensitivity for low doses.<br />

Page 18 of 62<br />

ATTENTION<br />

BEFORE USING THE mobileMOSFET, IT IS RECOMMENDED THAT YOU CHOOSE ONE<br />

SENSITIVITY SETTING AND LEAVE IT THERE FOR THE DURATION OF THE LIFE OF THE<br />

DOSIMETERS TO BE USED.


OPERATOR'S MANUAL FOR THE SYSTEM<br />

To determine which sensitivity best meets your needs, please review the chart in<br />

Section 5.3, or contact <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd for a copy of our Technical Note #3,<br />

“Reproducibility Using Model TN-RD-22 Dual Sensitivity Bias Supply”.<br />

A Sensitivity Switch is located on the front of the Reader adjacent to the five dosimeter<br />

ports on the left-hand side (refer to Figure 5-4: System Connections). The switch is<br />

recessed so that it cannot be accidentally switched to the wrong setting. In order to<br />

change the setting, use a fine tipped object, e.g. a pen, to move the switch to the<br />

appropriate setting.<br />

3.4 CONNECT DOSIMETERS<br />

MOSFET dosimeters require a permanent connection to a Reader for optimal<br />

accuracy readings. If your dosimeters are to be used for the first time, connect them<br />

to the Reader for at least one to twenty-four hours before use. Only connect and<br />

calibrate those dosimeters that are to be used (up to five (5) individual MOSFETs or<br />

one (1) Linear Array), and follow this procedure when ordering replacement<br />

dosimeters. You must designate the type of dosimeter, single or array that you are<br />

using, i.e. TN-252LA5, when configuring the equipment. (For Calibration, see Section<br />

7)<br />

NOTE: Single MOSFETs and the MOSFET Linear Array CANNOT be read<br />

simultaneously on a Reader. If you wish to read the MOSFET Linear<br />

Array, ensure that no MOSFETs are connected to the Reader and then<br />

designate which type of Linear Array you will be using. To do this, go<br />

to the Equipment Setup window in the mobileMOSFET Software, click<br />

on the Configuration button and follow the instructions provided.<br />

Conversely, if you wish to read single MOSFETs, ensure the MOSFET<br />

Linear Array is not connected to the Reader.<br />

Figure 3-2: System Configuration<br />

Page 19 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

3.5 CONNECT SYSTEM<br />

1) Connect the <strong>system</strong> together with all cables (see Figure 3-3: Equipment Set-up for<br />

a Wireless Connection, below).<br />

NOTE: If you are using the <strong>wireless</strong> connection, DO NOT connect the READER-PC<br />

CABLE.<br />

2) Prior to first use, place the Reader on Fast Charge. Turn the mobileMOSFET<br />

OFF and connect the universal power adapter for the Reader to the “POWER IN”<br />

jack on the back of the Reader, then plug the country specific line cord into the AC<br />

outlet. The red Fast Charge LED will go on. When the Fast Charge LED goes off<br />

and the green Trickle Charge LED goes on, the battery is fully charged. This will<br />

take a maximum of three hours.<br />

3) Power on the <strong>system</strong> and allow the electronics to warm up for at least fifteen (15)<br />

minutes before proceeding. If the green Power LED is not flashing, the power<br />

switch must be switched up to the momentary position to reconnect the battery<br />

power.<br />

3.6 CONNECT PC<br />

3.6.1 Wireless Connection<br />

Mount the TN-configured Bluetooth <strong>wireless</strong> transceiver on an interior wall of the<br />

treatment room. Attach the 5V DC power supply, and connect the transceiver to the<br />

PC’s COM port using the RS-232 cable (See Figure 3-3: Equipment Set-up for a<br />

Wireless Connection).<br />

Figure 3-3: Equipment Set-up for a Wireless Connection<br />

Page 20 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

The distance between the transceiver and mobileMOSFET should be less than ten (10)<br />

metres (33 feet) (see Section 5.5). For quality communication, it is suggested that the<br />

transceiver be installed at a height of two (2) metres (6 feet) or more, and if possible,<br />

on a surface without any obstacles between it and the intended location of the<br />

mobileMOSFET.<br />

The fifteen (15) metre (50 foot) Reader-PC cable can be used as an extension if the<br />

RS-232 cable between the computer and the transceiver is not long enough.<br />

More than one mobileMOSFET can be connected to a PC through a <strong>wireless</strong><br />

connection as depicted in Figure 3-2: System Configuration and Figure 3-4: Multiple<br />

Readers Connected via Wireless to a PC.<br />

PC<br />

Figure 3-4: Multiple Readers Connected via Wireless to a PC<br />

3.6.2 Cable Connection<br />

A Reader-PC Cable is also supplied with the mobileMOSFET Software. Attach one<br />

end to the PC and the other to the Serial Data Port on the back of the<br />

mobileMOSFET Unit.<br />

More than one mobileMOSFET can be connected to a PC as depicted in Figure 3-5:<br />

Various Wired Scenarios for Connecting Readers to a PC.<br />

PC<br />

PC<br />

COM 1<br />

COM 1<br />

COM 2<br />

USB<br />

USB TO<br />

SERIAL<br />

PORT<br />

HUB<br />

Figure 3-5: Various Wired Scenarios for Connecting Readers to a PC<br />

Page 21 of 62<br />

COM 5<br />

COM 6<br />

COM 7<br />

mobileMOSFET 1<br />

mobileMOSFET 2<br />

mobileMOSFET 3<br />

mobileMOSFET 1<br />

mobileMOSFET 2<br />

mobileMOSFET 1<br />

mobileMOSFET 2<br />

mobileMOSFET 3


OPERATOR'S MANUAL FOR THE SYSTEM<br />

3.6.3 LAN/Wireless Connection<br />

<strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd also provides an optional solution for establishing <strong>wireless</strong><br />

connections in additional treatment rooms via the LAN and Bluetooth Wireless<br />

Transceiver Package, TN-RD-38/40. This accessory allows the mobileMOSFET<br />

Reader Module to be shared amongst multiple rooms by connecting to a PC via a<br />

LAN (Local Area Network) connection.<br />

Figure 3-6: LAN Wireless Connection<br />

The Bluetooth Wireless transceiver supplied is a serial device, however a LAN<br />

connection can be implemented if a Serial Server is connected to the LAN port.<br />

Virtual communications port (COM Port) software (supplied with the TN-RD-38/40)<br />

allows the PC to connect to this COM Port anywhere on the LAN. <strong>Best</strong> <strong>Medical</strong><br />

<strong>Canada</strong> Ltd has tested and supplies a Vlinx Model ESP901 Ethernet Serial Server.<br />

(Please note that installation may require a LAN administrator to configure the end<br />

user's PC software IP settings.) Note: With virtual COM Port software, some users<br />

experience a delay while the PC searches for COM Ports on the LAN.<br />

Mount the BMC-configured Bluetooth <strong>wireless</strong> transceiver on an interior wall of the<br />

treatment room. Attach the power supply, then connect the transceiver to the Serial<br />

Server. Next, connect the Serial Server to the LAN and attach the LAN/Wireless<br />

Transceiver's power adapter (See Figure 3-6: LAN Wireless Connection).<br />

The distance between the transceiver and mobileMOSFET as well as the placement<br />

of the transceiver and mobileMOSFET must be as specified in Section 3.6.1.<br />

Figure 3-7: Multiple Systems and Treatment Rooms and Figure 3-8: Single<br />

System/Multiple Treatment Rooms provide examples of possible configurations:<br />

Page 22 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

Figure 3-7: Multiple Systems and Treatment Rooms<br />

Figure 3-8: Single System/Multiple Treatment Rooms<br />

Page 23 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

3.7 INSTALLING mobileMOSFET SOFTWARE<br />

The entire software program is contained on one CD-ROM.<br />

1) Before proceeding, we recommend that you close any applications you may have<br />

running as Setup cannot install <strong>system</strong> files or update shared files if they are in<br />

use.<br />

2) Insert the CD in the computer.<br />

3) From the Start menu, select Run<br />

4) Click the Browse button. From the pop-up dialog box “Browse”, go to the<br />

CD-ROM drive under “My Computer”. Click on “setup.exe”.<br />

5) Open this and then click OK. This will then take you back to the “Run” pop-up<br />

dialog box. Click on OK.<br />

Figure 3-9: Windows “Run” Pop-up Dialog Box<br />

6) Follow the instructions shown on your screen. (DO NOT change the default<br />

installation folder.)<br />

When the installation is finished, you will find that mobileMOSFET Software has been<br />

added to the PC’s “Start | Programs” list.<br />

Page 24 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

4 OPERATING PROCEDURE<br />

Set-up the <strong>system</strong> as described in Section 3, SYSTEM SET-UP. Click on the<br />

mobileMOSFET Software entry of your computer’s “Start | Programs” menu to run<br />

the program. The program will open to the Equipment Setup Screen as shown in<br />

Figure 4-1: Equipment Set-up Screen.<br />

4.1 EQUIPMENT SET-UP (SCREEN 1)<br />

Through this first screen, the User accesses many functions of the mobileMOSFET<br />

Software. The number of Readers that will be used for each treatment session are<br />

selected and assigned, the calibration parameters (CFs and CRs) can be modified,<br />

dosimeter status can be verified and the overall <strong>system</strong> status of connected<br />

components can be monitored throughout the treatment session. (See Section 7.1 for<br />

more information regarding CFs and CRs.)<br />

Figure 4-1: Equipment Set-up Screen<br />

The first step in this screen is to set the number of mobileMOSFET Reader Modules<br />

that will be used in the Treatment Session and then connect to them through the<br />

CONFIGURE button. By clicking on this button a Configuration screen will pop-up<br />

through which you can perform the steps noted above.<br />

In the Reader Assignment section of the screen, the number of Readers being used<br />

during each session, each Reader's Serial Number (S/N), Nickname (the User may<br />

assign a "name" up to 5 characters and/or a number to identify it) and the port that the<br />

Reader is connected to are displayed.<br />

Page 25 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

MOSFET CF/CR – Each Reader can read up to five (5) dose points. One Standard<br />

Dosimeter = one Dose Point. A Linear Array has five Dose Points and will display<br />

each position on the Flex graphically on the right-hand side of the screen. The<br />

dosimeters numbers are listed next to their corresponding Reader number, e.g. #1-4.<br />

#1 = the Reader Module and –4 = the Dosimeter #4, or Dose Point #4.<br />

Calibration Factor (CFs) units can be mV/cGy/, mV/Gy, or mV/R. The valid range is<br />

0.01 to 9999.99. If you leave any of the CF fields blank, the CF unit will be in "mV<br />

mode" therefore no CF, i.e. the dosimeter mV reading has not been altered. CRs are<br />

optional Correction Factors. When used, CFs and CRs calculate the dose according to<br />

the formula "DOSE = CR (Optional) * [Voltage/CF]". (See Section 7.1 for more<br />

information regarding CFs and CRs.)<br />

The Reader Status section of the screen shows the last time that the Reader status<br />

was verified and is updated each time it is checked. The MOSFET Voltage portion<br />

indicates the total voltage for each dosimeter's position. It will also indicate if a<br />

dosimeter is absent or saturated (exceeded the Reader's dose range) as well as the<br />

Bias setting for each Reader (Standard or High).<br />

The type of MOSFET dosimeter being used as well as the dosimeter position is<br />

displayed graphically. For example, for Standard MOSFETS a picture of a<br />

mobileMOSFET indicating each dosimeter's position will be displayed. If a Linear 5ive<br />

MOSFET Array (which has five positions) is being used, it will show that this is the type<br />

of MOSFET connected to the Reader.<br />

The Battery column provides a graphic indication of the approximate charge state of<br />

each Reader's battery.<br />

4.2 EQUIPMENT SETUP: CONFIGURATION<br />

4.2.1 Connect To Reader Module<br />

mobileMOSFET Software allows you to connect to the Reader Module using either a<br />

cable or <strong>wireless</strong> transceiver, and automatically determines the type of connection<br />

(wired or <strong>wireless</strong>) being used. To connect to the Reader, click the Configure button<br />

from the Equipment Setup window (see Figure 4-1: Equipment Set-up Screen).<br />

A separate Reader Configuration window will open where you select the number of<br />

Readers you wish to use (as shown in Figure 4-2: Reader Configuration Window).<br />

Page 26 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

Click Assign to search for and assign Reader(s). You are also given the option of<br />

assigning a name (e.g. "One") to identify your Readers. You are allowed a<br />

maximum of 5 alpha/numeric characters for each of these nicknames.<br />

Figure 4-2: Reader Configuration Window<br />

If a <strong>wireless</strong> transceiver is found, the program will search for Readers in the<br />

vicinity as shown in Figure 4-3: Connecting to a mobileMOSFET Reader.<br />

To select the Reader you wish to assign, double-click the icon of that Reader.<br />

Figure 4-3: Connecting to a mobileMOSFET Reader<br />

If a cable connection is found, double-click the icon of the Reader you wish to configure<br />

as with the <strong>wireless</strong> transceiver.<br />

Page 27 of 62<br />

0043


OPERATOR'S MANUAL FOR THE SYSTEM<br />

4.2.2 Dosimeter Selection<br />

The next step in the Configuration window is to designate the type of dosimeter that<br />

will be used in the treatment session, e.g. the standard single MOSFET, (one<br />

dosimeter on a flex), or a Linear 5ive Array dosimeter (five dosimeters on one flex).<br />

This feature is most important to those who wish to use the Linear Array, for reasons<br />

outlined in “Section 5.3.2 Linear 5ive MOSFET Array Dosimeter”.<br />

This feature is found at the bottom of the Configuration window (see Figure 4-4:<br />

Dosimeter Selection Menu below) with the following message: “Please select the<br />

type of dosimeter used by Reader #X”.<br />

The software provides the option of choosing from: Single MOSFET, TN-252LA5<br />

Array, TN-502LA5 Array, or TN-1002LA5 Array.<br />

Figure 4-4: Dosimeter Selection Menu<br />

4.2.3 Calibration Parameters (CFs and CRs):<br />

These fields are used to show Calibration Parameters, including Calibration Factors<br />

(CFs) and Correction Factors (CRs). A CF translates the MOSFET’s voltage reading<br />

into a dose unit (i.e. Gy, cGy, R), and a CR is a multiplicative factor that can be used<br />

to accommodate certain external factors (such as extrapolating partial dose depth<br />

data to Dmax).<br />

Page 28 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

The Reader will only read the MOSFET voltage in mV. The program transforms the<br />

obtained voltage into dose according to User-defined Calibration Factors (CFs). The<br />

program also enables the user to specify optional Correction Factors (CRs) for the<br />

dose calculation.<br />

The Calibration parameters (CFs and CRs) must be set, because they will be used to<br />

calculate the doses according to the formula "Dose = CR (optional) * [Voltage / CF]".<br />

The units of a CF can be in mV/cGy, mV/R or mV/Gy. The valid CF value is 0.01 to<br />

9999.99. If you leave any of the CF fields blank, the CF units will be in "mV Mode". In<br />

this case, the dose will be represented as a mV and it will be a direct voltage reading.<br />

The allowed CR range is 0.01 to 99.99.<br />

The software allows you to use different CFs/CRs for different treatments:<br />

1) When you have determined the CF/CFs, save them to a file by clicking on "Save" in<br />

the Configuration window. Give the files specific names to differentiate them.<br />

2) The next time you wish to use a particular set of CF/CRs, open the Configuration<br />

window and click on "Load" and choose from your predetermined list of CF/CR files.<br />

Click on the file that you wish to use and the new CF/CRs will populate the<br />

configuration window which can then be used for ensuing treatments.<br />

You can set the CFs in mobileMOSFET Software by selecting Calibrate MOSFET from<br />

the Tools drop-down menu of the Equipment Setup screen (See Figure 4-1).<br />

Note: It is recommended that new MOSFETs be calibrated prior to the first time they are<br />

used clinically.<br />

See Section 7.1 for more information regarding CFs and CRs.<br />

Page 29 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

4.3 TREATMENT INFORMATION (SCREEN 2)<br />

Figure 4-5: Treatment Information Screen<br />

When you have completed the Equipment Setup, you can then move to the next step by<br />

simply clicking on the tab "Treatment Info". Treatment Information is input here, including<br />

the target dose, as well as assigning the dosimeters to be used to the patient or phantom.<br />

As the dosimeter locations are displayed both in a table as well as graphically, a<br />

description of the location is entered in the Position column.<br />

Once this information has been completed, it can be saved for use in future sessions<br />

without having to re-input patient information or reassigning the dosimeters (see Section<br />

4.3.3, for the use of templates).<br />

ATTENTION<br />

• You will NOT be able to read dose data if you do not select one or more<br />

dosimeters. Select the dosimeters that you wish to read by clicking the MOSFET<br />

Use column corresponding to the dosimeter(s) that you wish to use. A checkmark<br />

will appear indicating which dosimeter(s) have been selected.<br />

• When assigning dosimeters, ensure that you select the dosimeters that directly<br />

correspond to the dosimeters you are actually using on the patient.<br />

Page 30 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

4.3.1 Image Selection<br />

The mobileMOSFET Software allows you to change the image displayed to reflect<br />

the treatment session. To change an image, click on the Operations button shown<br />

in Figure 4-5: Treatment Information Screen. Using the drop-down menu, select<br />

"Change Patient Image", which will open a separate window, as shown in Figure 4-6:<br />

Patient Image Window.<br />

Figure 4-6: Patient Image Window<br />

This program provides six optional images, called “Standard Images”.<br />

They are:<br />

� Unisex Body<br />

� Female Chest<br />

� Male Head<br />

� Female Head<br />

� Female Body<br />

� IMRT Phantom<br />

You can also import custom images from a digital camera or scanned photo. Any<br />

BITMAP (*.bmp), JPEG (*.jpeg, *.jpg) or GIF ( *.gif) image may be used. There is no<br />

size limitation for images.<br />

Please note that in order to maximize the amount of space used for saved reports,<br />

the images will be compressed to 40% of their original size. The images may<br />

therefore not be as clear as the original file for this reason.<br />

To use mobileMOSFET standard library images, click on the images to preview<br />

them. To select one from the list, click on it and then the "Apply" button. To use a<br />

Custom Image, refer to the "Custom Image Used" list in the same manner.<br />

Page 31 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

If the image you want to use is not in the list, but is already stored in a file, click the<br />

"Other" button to find and open it. After you apply this image, it will be automatically<br />

added to the "Custom Image Used" list.<br />

After the assignment of dosimeters, the corresponding Pointer/Dosimeter label will<br />

appear on the image area. To indicate the specific location, drag the pointer to the<br />

appropriate position (using the Treatment screen shown in Figure 4-5: Treatment<br />

Information Screen).<br />

4.3.2 Using Patients' Photographs During Measurement<br />

Using custom images to indicate dosimeter sites was discussed briefly, above.<br />

These images, however, may not exist before you run the program.<br />

If you have a digital camera, you can use a photo of the patient or phantom taken<br />

during treatment as follows:<br />

A. In Step 2, input the patient's treatment information, but DO NOT select an image.<br />

B. In Step 3, after placing dosimeters on the patient's body, take a photo. Then do<br />

the measurement.<br />

C. Minimize the mobileMOSFET Software window. Input the photo to your<br />

computer, saving it on the hard disk as a JPEG file. Restore the "Step 3" Screen.<br />

D. Return to the "Treatment Information" screen. Add the saved JPEG file to<br />

"Custom Image Used" and select it. Open the picture.<br />

E. Go to "Dosimetry Report" screen to view the report.<br />

Remember: You can change the treatment information and dosimeterassignment<br />

at any time BEFORE exiting the program. If you<br />

change this information AFTER the corresponding report has<br />

been saved, however, you should save the changed report under<br />

a new name. (Alternatively, you can delete the previous report.)<br />

4.3.3 Templates – A Time Saving Feature<br />

Treatment Information templates may be created in advance of the Treatment<br />

Session in order to save time, and/or to outline the target dose plan for a patient<br />

over the course of his/her treatments. Templates are created to guard information<br />

that a physicist or therapist will not want to repeatedly enter, e.g. Target Doses,<br />

Patient Name, Dosimeter Positions, etc. Multiple templates may be created.<br />

To use this feature:<br />

1) Enter the information into the Treatment Information window, as described in<br />

Section 4.3.<br />

2) To save the data, click on the Operations button and select the “Save Template”<br />

option in the drop-down menu.<br />

3) The data entered in the Treatment Information window will be saved for future<br />

use.<br />

Page 32 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

To open a template:<br />

1) Click on the Operations button in the Treatment Information window and select<br />

the “Open Template File” option in the drop-down menu.<br />

2) If necessary, enter additional information or change the data.<br />

3) Proceed to the Dose Measurement window.<br />

4.3.4 Dose Report Files: A Quick Template Alternative<br />

Instead of creating the abovementioned Template files, one may also use a Dose<br />

Report File as a Template in order to avoid re-entering data, such as Patient Name<br />

& Target Doses.<br />

To use this feature:<br />

1) Click on the Operations button in the Treatment Information window and select<br />

the “Open Saved Report File” option in the drop-down menu.<br />

2) If necessary, enter additional information or change the data.<br />

3) Proceed to the Dose Measurement screen.<br />

4.4 DOSE MEASUREMENT (SCREEN 3)<br />

A. Place MOSFETs and required build-up on the patient. (We suggest that you print out<br />

the dosimeter-site diagram from the "Treatment Information" screen as a reference.)<br />

If required, a low-level disinfecting of the patient contact area of the dosimeter may be<br />

accomplished with a swab of 70% isopropyl alcohol prior to use. Using a sterile<br />

gauze pad moistened with alcohol, wipe the entire length of the dosimeter a minimum<br />

of three times and allow to air dry.<br />

B. The dosimeters can now be attached to the patient using commonly available surgical<br />

or paper tape. Care should be taken in tape selection due to the fragility of the<br />

dosimeter tip. As an example, <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd recommends the use of 3M<br />

Micropore tape with low adhesion, but not 3M Transpore tape with very high<br />

adhesion. Use care when removing the tape, and be sure to remove the tape prior to<br />

removing the dosimeters from the patient or phantom. Please contact <strong>Best</strong> <strong>Medical</strong><br />

<strong>Canada</strong> Ltd for more information.<br />

IMPORTANT<br />

WHILE THE MOSFET DOSIMETER IS DESIGNED FOR EVERYDAY<br />

CLINICAL USE, ITS LIFE CAN BE SHORTENED THROUGH<br />

MISHANDLING. PLEASE ENSURE THAT PHYSICAL STRESS TO THE TIP<br />

OF THE DOSIMETER IS MINIMIZED DURING PROCEDURES.<br />

Reference Section 5.4 on Dosimeter Care<br />

C. To retain calibration, dosimeter orientation, beam and field size, the amount of buildup<br />

should be the same as that used during calibration. (See Section 7, CALIBRATION.)<br />

D. If build-up is required, refer to Section 7.1, Calibration for Radiotherapy.<br />

Page 33 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

E. INITIALIZE the MOSFETs by clicking on the START button on the screen.<br />

F. Irradiate.<br />

G. AFTER irradiation, read the MOSFETs ----- click on the READ button to obtain<br />

dose information.<br />

Figure 4-7: Dose Measurement Screen<br />

4.5 LINEAR COMPENSATION FOR ARRAYS<br />

On the Dose Measurement screen there is a column entitled "LC". This column is<br />

used to display the "Linear Compensation" factors for TN's Linear Array dosimeters.<br />

If a single MOSFET is being used (i.e. TN-502RD), no compensation is required and<br />

"N/A" will be displayed in the LC column.<br />

A Linear Array, however, requires compensation. If a voltage difference is detected<br />

during the Start-Read period, the mobileMOSFET Software will calculate the<br />

corresponding LC number based on the Array in use and the detected difference.<br />

If the type of Array has not been designated, a question mark, "?", will appear in the<br />

LC column to indicate that the Array has not been set. To do this, return to the<br />

Configuration Window in the Equipment Setup screen and select the type of array you<br />

are using. The compensation will automatically be applied to the final doses.<br />

Page 34 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

4.6 READING DOSIMETERS INDIVIDUALLY<br />

The mobileMOSFET Dose Reporter Software gives the user the ability to read the<br />

MOSFET dosimeters individually during measurement. This allows the user to take<br />

readings without having to stop during the treatment session to move the dosimeters<br />

to another position on the patient and the results output in a single Dose Report.<br />

To use this application, go to the "Treatment Info" Screen, select which dosimeter(s)<br />

to be read first, proceed to the Dose Measurement window, read them, then repeat<br />

this process as often as necessary until all dosimeters have been read. By using this<br />

method, the software only Initializes or Reads the dosimeters identified as "in-use"<br />

when the Start or Read button is clicked.<br />

For example, the patient has two areas (fields) that are to be irradiated, the face and<br />

the neck. Dosimeters A1 and A2 are placed on the patient's face, while dosimeters<br />

A4 and A5 are placed on the neck.<br />

First, prior to irradiating the face, go to the Treatment Info tab and under MOSFET<br />

Use, select A1 and A2 (ensuring that the remaining dosimeters are not selected,<br />

therefore not "in use"). Next, proceed to the Dose Measurement Screen, click on the<br />

Start button before the beam is turned on, and then click on the Read button after the<br />

beam has been turned off.<br />

Before the next field (neck) is irradiated, go back to the Treatment Info window and<br />

remove MOSFETs A1 and A2 from the "in use" column, and then select A4 and A5.<br />

Again go to the Dose Measurement window and repeat the process of clicking on the<br />

Start button before the beam is turned on and the Read button after it has been<br />

turned off.<br />

To output the results on one Dosimetry Report, simply go back to the Treatment<br />

Information screen and select MOSFETs A1 and A2 so that A1, A2, A4 and A5 are all<br />

"in use". The report can be printed out from the Dosimetry Report window, where the<br />

measurement results from all 4 dosimeters will be combined in a single report.<br />

4.7 REMOVING DOSIMETERS<br />

Read the dosimeters within 15 minutes of irradiation and then carefully remove the<br />

dosimeters from the patient.<br />

(Note – See Section 5.4 regarding the importance of dosimeter care.)<br />

4.8 DOSIMETRY REPORT (SCREEN 4)<br />

This last step allows you to review the information you've obtained in several ways.<br />

You can type in any comments/notes you wish to add, save the data and then print it<br />

in TN's report format. Saved reports can be opened and edited if necessary. If<br />

changes are made to a saved report, however, it is recommended that the revised<br />

report be saved under a new name. In addition, you can easily export the data to<br />

both Microsoft® Office Excel, Microsoft® Office Word, and Plain Text file formats.<br />

Page 35 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

All mobileMOSFET Software report files have a filename with the extension of ".mrpt".<br />

The default file name is "Patient's First Name + Patient's Last Name + Date<br />

(YYYMMDD) + Time (when the latest dose data was received) + .mrpt". For example,<br />

if John Smith was treated on August 18, 2005 and the last dose data was received at<br />

1:21 p.m., the default file name would be:<br />

John Smith_20050818_1321.mrpt<br />

Please note that if you wish to use the data collected in a future session, you must<br />

save it in the ".mrpt" format. The mobileMOSFET Software gives you the option of<br />

saving reports with or without password protection. Should you wish to protect your<br />

report, you will be asked to enter a password (maximum of 20 characters, no spaces<br />

are allowed) and then confirm it.<br />

To export the information to Microsoft® Office Excel Microsoft® Office Word or as a<br />

Plain Text File, simply select a format from the Operations drop-down menu as shown<br />

in Figure 4-8: Dosimetry Report Screen, and the mobileMOSFET Software will<br />

automatically export the data.<br />

If you export to Microsoft® Office Excel, a window will automatically open and<br />

populate "Sheet 1" with the raw dose data and "Sheet 2" will contain the patient image<br />

and dosimeter locations. When you export to Microsoft® Office Word, a Word<br />

document will be automatically populated with the entire dose data information. This<br />

document can then be saved and imported in Record and Verify programs that allow<br />

for the use of Word documents. Finally, if you want to retrieve the final dose data<br />

from another software program, e.g. a dose calculation software program, then you<br />

can export and save the data to a Plain Text file.<br />

Once the data has been exported, you may save your data as an Excel, Word or Plain<br />

Text file, however, remember that it must also be saved as a mobileMOSFET<br />

Software report if you wish to access the data in the future. Also, as the data in the<br />

Excel, Word or Plain Text documents can be altered, please ensure that the<br />

mobileMOSFET Software Dosimetry Report is the final report for your legal/patient<br />

files.<br />

Page 36 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

Figure 4-8: Dosimetry Report Screen<br />

4.9 SEQUENTIAL READING<br />

This feature enables you to read dose data periodically or at pre-determined time<br />

intervals. This is a useful tool for research and development activities or lengthy<br />

procedures such as Brachytherapy or Fluoroscopy.<br />

To access this feature, run the mobileMOSFET Software and go to the Dose<br />

Measurement screen. Click on Tools | Sequential Reading. (As the Sequential<br />

Readings rely on uninterrupted communication with the <strong>wireless</strong> transceiver, only one<br />

verification session can be running at a time in order to prevent any delays.)<br />

An example of the results from a Sequential Reading are shown in Figure 4-9:<br />

Sequential Reading Screen.<br />

Page 37 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

Figure 4-9: Sequential Reading Screen<br />

Page 38 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

To alter the settings of this screen, click on the CHANGE button in the Settings area of the<br />

screen. The pop-up screen offers two areas that can be customized, "What to Read" and<br />

"How to Read".<br />

Figure 4-10: Sequential Reading Settings<br />

1) What to Read - Choose to read in Dosimeter Total Voltage (mV) and you will see<br />

total voltage accumulating with radiation exposure. If you choose Incremental Dose<br />

After Initialization, incremental dose (based on previously input CFs) will be<br />

displayed from a 0mV reference.<br />

2) How to Read:<br />

a) Automatically (Default): The software will continuously read the doses at<br />

prescribed intervals during irradiation. You do not need to continually press the<br />

READ button, as this will be performed automatically from the software.<br />

The Interval function is used to read dose data at specified time intervals. The<br />

interval between readings can be set to a wide area of values ranging from<br />

10 seconds, 15 seconds, up to 30 minutes.<br />

b) Manually: This allows you to click on the READ button to measure the<br />

dosimeter doses on demand. Initialization is only required once.<br />

4.10 SIMULTANEOUS TREATMENT SESSIONS<br />

The mobileMOSFET Software allows the user to simultaneously conduct up to five<br />

treatment sessions. Each session has a dedicated verification window, and the<br />

operations of each are totally independent.<br />

To add a new session, click on the File | Add Verification Window in any of the<br />

screens. Each session is assigned a "Session Number" from 1 to 5. This number is<br />

displayed on the window's title bar as well as in the top right-hand corner of the<br />

window along with the Patient's name.<br />

Page 39 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

To move from one session to another you can minimize the current session. All<br />

open sessions will be shown at the bottom of the screen. You can also click on<br />

"File" and select the Verification Session you wish to switch to from the drop-down<br />

menu.<br />

Figure 4-11: Multiple Verification Sessions<br />

4.11 MESSAGE HISTORY TXT FILES<br />

All messages sent by the Reader to the Software are recorded in a Message History<br />

File. The Message History Files are saved in "C:\TN-Dosimetry\mobileMsgAchieve",<br />

which can be viewed and printed using "Notepad". One can cut and paste this<br />

information into other electronic documents.<br />

A Message History File contains the connection and dose history of each connected<br />

Reader, including records of connection activities, battery status and dosimeter<br />

readings.<br />

Each time a Message History File is opened, the mobileMOSFET will check its size.<br />

It if is larger than 50KB, it will be archived to "C:\TN-Dosimetry\mobileMsgArchive".<br />

4.12 MESSAGE WINDOWS<br />

Pop-up windows are used to display Reader Status messages from the Reader.<br />

View the Reader Status records at any time on each screen by selecting<br />

Connection Log or Reader Status Log from the Tools | View menu.<br />

Page 40 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

5 DESCRIPTION OF SYSTEM COMPONENTS<br />

5.1 mobileMOSFET READER MODULE<br />

Figure 5-1: Model TN-RD-70-W Top Panel<br />

The operations of the Reader are controlled by the mobileMOSFET Software.<br />

The Reader carries out the following functions:<br />

• Measures the dosimeter before and after irradiation.<br />

• Provides power and data distribution.<br />

Page 41 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

5.1.1 Reader Side Panel Layout<br />

The side panel contains the status LEDs. The following paragraphs describe these<br />

items in detail.<br />

Figure 5-2: Model TN-RD-70-W Side Panel<br />

5.1.2 Switches and LEDs<br />

The “POWER” SWITCH –has three states (OFF, ON, INITIALIZE<br />

BATTERY/MOMENTARY). The INITIALIZE position is a temporary position used to<br />

reconnect the battery if the unit battery has been completely drained. To INITIALIZE<br />

the Reader, depress the switch to the ON position.<br />

When using the unit, ensure that the mobileMOSFET is in the ON state and the<br />

green LED is flashing. The ON position is with the switch depressed to I, and the<br />

OFF position is with the switch depressed to O (please see Figure 5-3: Model TN-<br />

RD-70-W Back Panel).<br />

1<br />

2<br />

3<br />

4<br />

1<br />

2<br />

The “WIRELESS” LED – is a blue indicator that signals if the<br />

mobileMOSFET Reader is currently connected to a PC via a <strong>wireless</strong><br />

transmitter.<br />

The “POWER” LED – is also a green indicator that signals that the<br />

unit is turned on.<br />

The “TRICKLE” LED – is a green indicator that the unit is currently in<br />

trickle charge.<br />

The “FAST” LED – is a red indicator that signals the unit is currently in<br />

a fast charge.<br />

Page 42 of 62<br />

3<br />

4


OPERATOR'S MANUAL FOR THE SYSTEM<br />

5.1.3 Reader Back Panel Layout<br />

The following paragraphs describe each of the connectors and their specific<br />

purpose:<br />

POWER SWITCH<br />

Figure 5-3: Model TN-RD-70-W Back Panel<br />

POWER IN: Input power jack for the supplied universal power adapter.<br />

ON – OFF: Power switch (also reference Section 5.1.2)<br />

RS-232 I/O: RS-232C standard serial port used by the Reader to transmit data to<br />

and from the PC via a <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd supplied cable.<br />

WIRELESS CONNECTION: The remote connection to the transceiver is embedded<br />

inside the mobileMOSFET Reader Module<br />

BIAS SELECTION<br />

POWER JACK<br />

MOSFET DOSIMETER PORTS<br />

Figure 5-4: System Connections<br />

BIAS SELECT - Sets the sensitivity of the mobileMOSFET. Refer to Section 3.3<br />

for additional information regarding this setting.<br />

MOSFETs - Up to 5 MOSFETs can be connected to a mobileMOSFET<br />

Reader.<br />

ARRAY - A single Linear MOSFET Array can be connected to the<br />

mobileMOSFET.<br />

Page 43 of 62<br />

STANDARD SERIAL PORT<br />

LINEAR ARRAY PORT


OPERATOR'S MANUAL FOR THE SYSTEM<br />

Page 44 of 62<br />

NOTE<br />

MOSFETS AND A MOSFET LINEAR ARRAY CANNOT BE READ<br />

SIMULTANEOUSLY. IF YOU WISH TO READ THE MOSFET LINEAR<br />

ARRAY, ENSURE NO SINGLE MOSFETS ARE CONNECTED TO THE<br />

READER. CONVERSELY, IF YOU WISH TO READ SINGLE MOSFETS,<br />

ENSURE THE MOSFET LINEAR ARRAY IS NOT CONNECTED TO THE<br />

READER.<br />

5.2 RECOMMENDED DOSIMETER AND BIAS SENSITIVITY COMBINATIONS FOR<br />

APPLICATIONS IN RADIOTHERAPY AND RADIOLOGY<br />

In order to determine which components best meet your needs, please review the<br />

following chart:<br />

RADIOTHERAPY APPLICATION<br />

TYPICAL<br />

APPLICATION<br />

Scatter Dose/<br />

Research<br />

RADIOTHERAPY<br />

DOSE<br />

1 to 20 cGy<br />

RECOMMENDED<br />

DOSIMETER<br />

High Sensitivity<br />

Dosimeter<br />

RECOMMENDED<br />

BIAS SENSITIVITY<br />

High Setting<br />

Central/ Scatter<br />

Dose 20 to 100 cGy Standard Dosimeter High Setting<br />

TBI/ Central Dose >100 cGy Standard Dosimeter Standard Setting<br />

TYPICAL<br />

APPLICATION<br />

RADIOLOGY APPLICATION<br />

RADIOLOGY SKIN<br />

ENTRANCE DOSE<br />

Radiology 0.1 to 5 R<br />

Normal<br />

Fluoroscopy<br />

High Dose<br />

Fluoroscopy<br />

RECOMMENDED<br />

DOSIMETER<br />

High Sensitivity<br />

Dosimeter<br />

RECOMMENDED<br />

BIAS SENSITIVITY<br />

High Setting<br />

5 to 20 R Standard Dosimeter High Setting<br />

20 to > 100 R Standard Dosimeter Standard Setting


OPERATOR'S MANUAL FOR THE SYSTEM<br />

5.3 DOSIMETERS<br />

The dosimeter is composed of a 1.4 meter cable attached to a length of thin, semiopaque<br />

polyimide laminate, also known as "the flex", with a silicon detector mounted<br />

on the end<br />

5.3.1 Single MOSFET Dosimeters<br />

On Single MOSFETs, the silicon detector itself is mounted on the end of the polyimide<br />

material while on the Linear 5ive Array (see Section 5.3.2) they are mounted at 2cm<br />

inter-spacing, each MOSFET under a 1mm layer of black epoxy.<br />

Individual calibration factors (CFs) of between 0.01 and 9999.99 for up to 20 dosimeters can<br />

be operator entered into the mobileMOSFET Software. The CFs are used to convert the<br />

dosimeter voltage to cGy, R or Gy when selected.<br />

The following two tables show the dosimeter sensitivities.<br />

STANDARD SENSITIVITY DOSIMETER MODELS TN-502RD & TN-502RDM<br />

DOSIMETER<br />

BIAS<br />

SENSITIVITY<br />

SETTING<br />

Page 45 of 62<br />

NOMINAL<br />

SENSITIVITY<br />

ENERGY SOURCE<br />

Standard Standard 1 mV/ cGy Co 60 with buildup<br />

Standard High 3 mV/ cGy Co 60 with buildup<br />

Standard Standard 3 mV/ R Diagnostic X-ray<br />

Standard High 9 mV/ R Diagnostic X-ray<br />

HIGH SENSITIVITY DOSIMETER MODELS TN-1002RD & TN-1002RDM<br />

DOSIMETER<br />

BIAS<br />

SENSITIVITY<br />

SETTING<br />

NOMINAL<br />

SENSITIVITY<br />

ENERGY SOURCE<br />

High Sensitivity Standard 3 mV/ cGy Co 60 with buildup<br />

High Sensitivity High 9 mV/ cGy Co 60 with buildup<br />

High Sensitivity Standard 10 mV/ R Diagnostic X-ray<br />

High Sensitivity High 30 mV/ R Diagnostic X-ray


OPERATOR'S MANUAL FOR THE SYSTEM<br />

5.3.2 Linear 5ive MOSFET Array Dosimeter<br />

For additional information, see Technical Note #8.<br />

Real-time evaluation of the quality of the implant or HDR dose delivery can be provided by<br />

the in-vivo use of a MOSFET 5ive Linear Array Dosimeter (models TN-252LA5, TN-502LA5,<br />

or TN-1002LA5). These measurements will allow immediate assessment of post-implant<br />

base and apex dose coverage, as well as the dose to organs at risk such as the urethra,<br />

rectum or bladder.<br />

Standard MOSFET Dosimeters, which have one dosimeter located at the tip of the flex, are<br />

linear with dose, with a 5% maximum variation of the calibration factor throughout the entire<br />

20,000mV reading range of the MOSFET. The initial calibration factor can be used for all<br />

measurements, without a need for further calibration or correction.<br />

In the case of the Linear 5ive Array, linearity corrections (LC) are implemented automatically<br />

in the mobileMOSFET Dose Reporter Software at doses above 5000cGy. This is run ‘in the<br />

background’ by the software and is visible to the user in the LC column. These linearity<br />

corrections will remain accurate throughout the 20,000 mV reading range of the dosimeter.<br />

For this reason it is very important that the user designate the correct type of Linear Array<br />

being used as well as the Bias selection (Standard or High) to allow the software to properly<br />

calculate and correct the CFs.<br />

Figure 5-5: MOSFET Linear 5ive Array<br />

Dimensions: 1.5mm wide, 46cm long, 1.3mm thick, 2cm Inter-MOSFET Spacing.<br />

The Radio-Opaque Marker located at tip of flex, 7mm away from<br />

the center of the first MOSFET on the array.<br />

Page 46 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

Radiation Characteristics:<br />

• 0.04 mm 2 active detection area<br />

• Suitable for photon and electron modalities<br />

• Isotropic response (±3% for 360 ο )<br />

• Dose rate independent<br />

• Constant sensitivity under full build-up for all energies<br />

• Visible under CT or Fluoroscopy<br />

Sensitivity under full build-up ( 60 Co)<br />

ARRAY STANDARD SENSITIVITY HIGH SENSITIVITY<br />

TN-252LA5<br />

TN-502LA5<br />

TN-1002LA5<br />

Page 47 of 62<br />

0.81 mV/cGy<br />

2.76 mV/cGy<br />

6.43 mV/cGy<br />

Dose reproducibility at standard sensitivity bias at 1σ<br />

1.11 mV/cGy<br />

3.8 mV/cGy<br />

9.27 mV/cGy<br />

ARRAY 20 cGY 200 cGY<br />

TN-252LA5<br />

TN-502LA5<br />

TN-1002LA5<br />

< 2 %<br />

< 2 %<br />

< 2 %<br />

Three array types for all your applications:<br />

< 1 %<br />

< 1.5 %<br />

< 1.5 %<br />

1. TN-252LA5 for moderate to high dose (20~2000 cGy) applications:<br />

(HDR Brachytherapy 192 Ir, IMRT QA)<br />

2. TN-502LA5 for low to moderate dose (1~100 cGy) applications:<br />

(LDR Brachytherapy, Diagnostic X-rays)<br />

3. TN-1002LA5, for low and very low dose (1~20cGy) applications<br />

(LDR Brachytherapy, Diagnostic X-rays).<br />

5.3.3 LINEAR 5ive ARRAY: CLINICAL IMPLEMENTATION AND DOSE SAMPLING<br />

Pre-Implant Procedure:<br />

1. Prior to the in-vivo measurements the array is sterilized in Cidex solution.<br />

2. The array is then attached to a Bias Supply and connected to a MOSFET 20<br />

Reader for initialization.


OPERATOR'S MANUAL FOR THE SYSTEM<br />

3. The detector is inserted into the central branch of an approved three way sterile<br />

catheter via a ‘Y’ connector (see Figure 5-6: MOSFET Linear 5ive Array Before<br />

Insertion in the 3-Way Catheter), which then is inserted in the urethra, rectum or<br />

bladder. The radio-opaque marker, located on the tip of the array, helps visualize,<br />

under fluoroscopy or CT, the 5 MOSFET positions relative to the internal bladder<br />

wall (at the lower edge of the catheter balloon). The array size, its waterproof cable,<br />

and the use of a Y type connector make body fluid evacuation clean and easy,<br />

helping to make the patient more comfortable.<br />

During the LDR Implant or HDR Procedure:<br />

1. During the implant or HDR procedure, 5 dose points can be read directly from the<br />

array, or remotely from a computer, displaying the real urethral dose profile<br />

throughout the prostate location, compared to the planned profile.<br />

2. From urethra dose readings, one can infer the maximum initial dose rate (IDRmax)<br />

to the urethra and compare it to the recommended dose limit (1.5-2 mPD:<br />

minimal Peripheral Dose). The measured dose versus distance curve along the<br />

prostatic urethra contains the information about two factors related to the quality<br />

of the implant: the dose rate at the base of the prostate and the length of urethra<br />

that receives the mPD, compared to the length of the urethra measured with<br />

ultrasound.<br />

End of Implant Procedure:<br />

1. After the implant, real time <strong>dosimetry</strong> using the Linear Array MOSFET Dosimeter<br />

allows for correction of lack of the proper dose coverage at the prostate base.<br />

However there is no practical way to reduce an excessive urethral dose, except<br />

that a patient may be followed closely during post treatment phase and<br />

considered to be at increased risk for urethral complications.<br />

Figure 5-6: MOSFET Linear 5ive Array Before Insertion in the 3-Way Catheter<br />

Page 48 of 62


5.4 DOSIMETER CARE<br />

OPERATOR'S MANUAL FOR THE SYSTEM<br />

IMPORTANT<br />

WHILE THE MOSFET DOSIMETER IS DESIGNED FOR EVERYDAY CLINICAL<br />

USE, ITS LIFE MAY BE SHORTENED THROUGH MISHANDLING. PLEASE<br />

ENSURE THAT THE IMPORTANT GUIDELINES PROVIDED BELOW ARE USED<br />

TO CARE FOR DOSIMETERS. ALSO PLEASE REFERENCE THE WARRANTY<br />

SECTION OF THIS DOCUMENT.<br />

5.4.1 STORAGE<br />

New dosimeters are packaged in a protective sleeve. Always store the dosimeter in this<br />

sleeve when not in use.<br />

5.4.2 USE OF TAPE<br />

Care should be taken in tape selection. Always use low adhesive surgical tape to attach<br />

MOSFETs to a surface. <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd recommends the use of 3M Micropore<br />

tape with low adhesion, but not 3M Transpore tape with very high adhesion. Use care when<br />

removing the tape, and be sure to remove the tape first, prior to removing the dosimeters<br />

from the patient or phantom. If required, dosimeters can be ordered with additional heatshrink<br />

for extra strength. Contact <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> for information.<br />

5.4.3 CARE DURING USE<br />

A. While the dosimeter is designed to flex during clinical use, avoid kinking the flex as this<br />

can shorten it's life, as well as affect the product warranty.<br />

B. Always use sterile lubricating gel to ease insertion as this helps to prevent kinking.<br />

C. Avoid tugging on the flex to remove.<br />

D. Use bolus to protect the dosimeter when using phantoms or build-up material.<br />

E. Avoid touching the dosimeter connector to avoid possible ESD (electrostatic discharge)<br />

shocks.<br />

F. Do not modify the dosimeter by removing any of the protective coverings that are<br />

designed to protect the dosimeter during use.<br />

Page 49 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

5.5 WIRELESS TRANSCEIVER<br />

MOUNTING HOLES, 03.3x23 mm<br />

5V DC INPUT<br />

Figure 5-7: Wireless Transceiver<br />

The transceiver is used to control the Reader and receive dose data <strong>wireless</strong>ly. It<br />

has been configured so that it is dedicated to <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd’s<br />

mobileMOSFET Wireless Dosimetry System.<br />

Before using the transceiver, you must:<br />

• Mount the transceiver in the treatment room. For reliable communication, we<br />

suggest that you place the transceiver in a location (on a wall) at a height of 2<br />

meters/6.5 feet or higher from the floor, and without obstacles (i.e. brick or steel<br />

reinforced walls) between the transceiver and the Reader. Do not place the<br />

transceiver close to large metal objects.<br />

• Provide power to the transceiver using the 5V DC Power Supply<br />

• Connect the transceiver to your computer’s COM Port using the DTE-to-DTE<br />

RS-232 cable<br />

NOTE: THIS TRANSCEIVER IS INTENDED FOR SHORT-RANGE (10 METERS/30<br />

FEET) COMMUNICATION ONLY. IF THE READER IS PLACED TOO FAR FROM<br />

THE TRANSCEIVER, COMMUNICATION MAY NOT BE ESTABLISHED.<br />

The <strong>wireless</strong> transceiver emits four different colors of light to indicate which mode it is in.<br />

They are:<br />

Green - The transmitter is idle.<br />

Red - It is communicating with the PC.<br />

Blue - It is communicating with the Reader.<br />

Purple - It is attempting to communicate with the Reader.<br />

Page 50 of 62<br />

CABLE CONNECTOR


OPERATOR'S MANUAL FOR THE SYSTEM<br />

If the RS-232 cable provided is not long enough to connect the transceiver to your<br />

computer, you can extend it by using the Reader-to-PC cable (15 meters/50 feet long) as<br />

shown in Figure 3-3: Equipment Set-up for a Wireless Connection. For LAN Connection,<br />

please see Section 3.6.3.<br />

TRANSCEIVER<br />

TRANSCEIVER CABLE<br />

Figure 5-8: Extension of Transceiver-Reader Cable<br />

6 MOBILEMOSFET SYSTEM MESSAGES<br />

SATURATED – Will appear on the PC after an operation if the dosimeter<br />

has acquired an accumulated dose that exceeds the<br />

reading capability of this instrument. The dosimeter must<br />

be replaced, and the Reader will not display dose<br />

information for the depleted dosimeter.<br />

ABSENT - Will appear on an operation if no dosimeter is connected to<br />

the Reader on the respective channel.<br />

UN-INITIALIZED – Will appear during a read operation when a dosimeter has<br />

failed to be initialized.<br />

BATTERY ICON – A battery icon is displayed on the PC to indicate the power<br />

status of the Reader's battery.<br />

DOSIMETER CONFLICT – A pop-up window will be displayed if MOSFETs and a<br />

MOSFET Linear Array are BOTH connected to the<br />

mobileMOSFET Reader.<br />

Page 51 of 62<br />

COMPUTER<br />

READER-TO-PC CABLE<br />

(15 METERS/50 FEET)


OPERATOR'S MANUAL FOR THE SYSTEM<br />

7 CALIBRATION PROCEDURE<br />

To obtain the maximum measurement accuracy, dosimeters should be calibrated prior<br />

to the first measurement, and also after any lengthy period of not being used.<br />

Calibration is performed by measuring the radiation sensitivity of the dosimeter under<br />

known conditions. The dosimeter is placed under a radiation beam at a specified dose<br />

level, and the output voltage of the dosimeter is then compared with the set level. The<br />

ratio of the measured voltage value to the actual value of the radiation dose determines<br />

the calibration factor. To improve the accuracy of the calibration procedure, the<br />

measurement should be performed several times and the average value of the<br />

calibration factor calculated. Because the radiation level varies sharply close to the<br />

air/medium interface, it is also recommended that “build-up caps” are employed during<br />

the calibration procedure to obtain maximum accuracy and repeatability.<br />

This process will result in calibration factors which are almost independent of modality<br />

or energy (>1MV or >1MeV). This energy independence also holds true when the<br />

dosimeters are used on patients (see the list of publications available from <strong>Best</strong><br />

<strong>Medical</strong> <strong>Canada</strong> Ltd at www.Team<strong>Best</strong>.com)<br />

Once calibrated, the dosimeter may be used to measure doses at specific sites, e.g.<br />

dose under bolus, under an immobilization device, or scatter dose to the eye. The<br />

calibration factor determined for full build-up is used under these circumstances.<br />

There are situations where other types of dose information are required and a different<br />

calibration approach may be needed. One example is the determination of entrance<br />

and exit doses; another is calibrating for Brachytherapy. <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd<br />

maintains an updated record of published information on this and other applications<br />

which may be of use to the operator.<br />

It is very important that the user designate the correct type of Linear Array being used<br />

as well as the Bias selection (Standard or High) to allow the software to properly<br />

calculate and correct the CFs.<br />

Note: Brachytherapy/Linear Array – For additional calibration information on calibrating<br />

for brachytherapy, especially using the Linear Array, please see Section 7.2.<br />

Page 52 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

7.1 CALIBRATION METHODS FOR RADIOTHERAPY USING THE MOBILEMOSFET<br />

SOFTWARE AIDED CALIBRATION TOOL<br />

The mobileMOSFET Software provides a facility to simplify the calibration of MOSFET<br />

dosimeters.<br />

Note 1: To use more than one mobileMOSFET Reader at the same time, simply<br />

click on the menu File/Add Calibration Window.<br />

Note 2: As a means of becoming familiar with the Calibration Procedure, you may<br />

also select Simulation Mode from the Setup dialog in the right hand corner<br />

of the screen.<br />

Figure 7-1: Calibration Window<br />

From the Equipment Setup Screen click the drop-down menu item. A “Calibration”<br />

window will pop up as shown in Figure 7-1: Calibration Window. This is the user<br />

interface for the software tool to calibrate the MOSFET dosimeters. This tool helps<br />

you record dose data, calculate the CFs for each dosimeter, get averaged CF values<br />

after multiple readings, calculate deviations to determine whether to accept the<br />

obtained CF values, and save, print and apply the results. It is important that the<br />

correct type of MOSFET and Bias Setting are selected when configuring the<br />

Calibration Setup screen.<br />

Page 53 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

The Dose can be delivered to dosimeters by two methods:<br />

Uniform: This choice allows you to apply one expected dose, e.g.<br />

200cGy, and it will apply to all dosimeters.<br />

Designate: Apply different expected dose values to each dosimeter.<br />

The calibration procedure is as follows:<br />

A. Set up the <strong>system</strong> as described in Section 3, SYSTEM SET-UP. Ensure that the<br />

Bias Sensitivity switch on the Reader Module is set to the desired setting as per<br />

Section 3.3 and that the dosimeter(s) to be used has been identified.<br />

B. Run the mobileMOSFET Software. Turn on the Reader. Connect and check that<br />

the Software CFs are set to the default of 1.00 mV/cGy and that the CRs are set<br />

to 1.00. Next, confirm that the dosimeters have been connected as outlined in<br />

Section 3.4.<br />

C. Using a low-adhesive tape, (e.g. 3M paper tape sample supplied with unit) attach<br />

the dosimeters to a block of backscatter material. If build-up is to be used, place<br />

the required amount of build-up material on top of the flat face of the dosimeter.<br />

Although the dosimeters are isotropic, there is a ±2% difference in response from<br />

the front to the back. After calibration, the dosimeters should be applied to the<br />

patient in the same orientation to the beam as when calibrated. See Section 7.3<br />

below for further information on Build-up.<br />

D. Place the dosimeters near the center of a radiation beam of desired energy, with<br />

the build-up material and the flat side of the dosimeters facing the beam for high<br />

energy radiotherapy doses (>1MV). An ionization chamber or a LINAC’s<br />

internal monitor chamber may be used to measure a known dose and the<br />

beam field size should be recorded for future reference.<br />

E. Click on Start button to initialize MOSFETs before irradiation.<br />

F. Irradiate dosimeters.<br />

G. After irradiation click the READ button to measure results.<br />

H. Enter the obtained dose value.<br />

I. Click “Accept” to store the obtained CFs to “Raw CF Pool”.<br />

J. Repeat steps E – I at least 3 times, to optimize reproducibility by averaging.<br />

K. Review the Final CFs (averaged from the Raw CF Pool). If they are acceptable,<br />

click on the boxes in the CF Acceptance. You may then apply Save and/or Print<br />

the calibration results.<br />

CF Calculation is based on the following:<br />

MOSFET mV Reading (mV)<br />

CF = KNOWN RADIATION VALUE (e.g. cGy/R/Gy)<br />

Page 54 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

7.2 CALIBRATION OF MOSFET AND LINEAR 5ive ARRAY MOSFET DOSIMETERS:<br />

GUIDELINES FOR DIFFERENT ENERGIES<br />

(e.g. BRACHYTHERAPY)<br />

Disclaimer: All of the information and methods described in Section 7.2 are for<br />

reference only. For proper calibration and to account for the different radiation<br />

processes at different radiation energies and modalities, it is strongly advised to refer<br />

to detailed <strong>dosimetry</strong> methods as described in literature, (e.g. Introduction to<br />

Radiological Physics and Radiation Dosimetry by Frank H. Attix, John Wiley & Sons).<br />

CALIBRATION AT DIFFERENT ENERGIES:<br />

A) High energy beams (1 MeV~18 MeV):<br />

The MOSFETs do not show energy variations of CFs for high energy beams<br />

(1MeV -18MeV) for electron and photon modalities. Indeed this variation is within 3<br />

% at Dmax for this energy spectrum. Effects linked to different electron<br />

contamination at high energy beams, Linac Make dependent, may induce<br />

variations in CFs . It is then advised to calibrate MOSFETs in every Linac and for<br />

the energies of interest. Data sheets of typical CFs, measured for different Single<br />

and Linear Array MOSFETs at Co-60 beams, under 6mm build-up, are available<br />

for reference.<br />

B) Cesium sources (0.662 MeV):<br />

The Cesium-137, with a main energy spectrum centered around 0.662 MeV<br />

provides a response close to the one obtained using Co-60. Only few percent of<br />

CF variation could be observed at this energy. To calibrate the MOSFETs, a small<br />

amount of build-up is needed, in the range of 3-6mm, to account for the electronic<br />

equilibrium.<br />

C) Iridium sources (380 keV):<br />

Iridium-192 is widely used in High Dose Rate (HDR) Brachytherapy. Its average<br />

energy is around 380 keV. This spectrum contains low energy components below<br />

100 keV, which increase the MOSFET sensitivity as a result of enhanced<br />

photoelectric effect (~ Z 4 ) on the MOSFET active area made of SiO2. As for the<br />

MOSFET response with Ir-192, it increases by a factor of approximately 1.2 from<br />

the CF values obtained with Co-60.<br />

For calibration of the MOSFETs, it is advised to put the source catheter in the<br />

middle of a wide solid water block material (30cm wide, 5cm thick), and lay the<br />

MOSFET parallel to the source catheter at a distance of 5-10cm (to avoid steep<br />

gradient variation of dose at the vicinity of the source). Then protective bolus and<br />

solid water bloc is used to cover the MOSFETs and source catheter to account for<br />

dose scattering contribution.<br />

Note: Caution must be used to ensure that the dosimeter(s) do not get crushed.<br />

If a heavy piece of build-up must be used, please ensure that it contains a<br />

customized groove for the MOSFET or a soft piece of Bolus is placed on top of the<br />

Page 55 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

dosimeter(s) in order to cushion it and prevent breakage.*. Reference Section 5.4<br />

on dosimeter care.<br />

Using treatment planning software, and the actual source activity, source position,<br />

and dwell time, dose can be estimated within the vicinity of the source, particularly<br />

at the position of the MOSFET of interest. This will enable the calibration of any<br />

MOSFET or Linear Array dosimeter. For the individual calibration of the 5<br />

MOSFETs of the Linear Array, it is advised to move the source in its catheter by<br />

steps of 2cm, and ensure that it is vertical to every single array MOSFET, to avoid<br />

source anisotropy effects. In general, 3 short exposures of 30-50 cGy can be<br />

applied to every MOSFET to obtain its CF.<br />

Another variation for calibration is to use a water tank with 2 parallel jigs at a fixed<br />

distance; one jig holding the Linear Array, and the other the source catheter. In<br />

this case a calibrated ionization chamber can be also placed at the same distance<br />

from the source to allow for absolute dose estimation.<br />

D) Iodine source (30 keV):<br />

Iodine-125 seeds are widely used in Low Dose Rate (LDR) Brachytherapy. The<br />

energy spectrum is centered around 30 keV. Due to the photoelectric effect on<br />

Silicon, the MOSFET response can increase by a factor of 3-4 from its response at<br />

Cobalt energy.<br />

Two approaches can be used for calibration:<br />

- An orthovoltage X-ray machine can be used to simulate the 30 keV energy<br />

response. As the average energy of an X-ray spectrum is a third of the tube's<br />

maximum electron energy (given by the kilovolt setting), a setting of 90-100 kV<br />

of the X-ray tube can be used to obtain the MOSFET response at 30 keV. A flat<br />

ionization chamber should be used to estimate the absolute dose under 1-2cm<br />

build-up (to account for photon scattering). The tube filtering and spectrum<br />

modification should be checked before performing this operation, as filtered<br />

beams tend to provide lower average energies than expected (1/3).<br />

- An Iodine seed can also be used for absolute and fine calibration. This requires<br />

a special phantom*, in which the seed is inserted with the MOSFET (in a<br />

machined groove) underneath the seed by 2-4cm. As seed activity is weak,<br />

several minutes (15-30 minutes) will be required to record a significant dose<br />

reading in mV on the MOSFETs. Using the activity of the seed given by the<br />

respective manufacturer and following the guidelines of dose calculation from<br />

TG-43 protocol (or suitable software), dose can be estimated at the MOSFET<br />

position, allowing exact CF estimation. An ionization chamber can be also used<br />

for dose estimation at the MOSFET position; however its large volume may not<br />

provide the spatial resolution needed.<br />

*Contact <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd or your phantom manufacturer for MOSFET<br />

groove dimensions and tolerances.<br />

Page 56 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

7.3 BUILD-UP & PARTIAL BUILD-UP<br />

Because the interface between skin and air, or any backscatter medium and air,<br />

provides a region of rapidly varying dose levels, it is often appropriate to use build-up<br />

on the MOSFET. (The MOSFET has 0.8 mm inherent build-up.) This build-up is<br />

usually a hemispherical device which fits over the MOSFET and allows the simulation<br />

of the performance of the MOSFET at a defined depth (2 cm x 2 cm² bolus is also<br />

used). The use of build-up allows a more accurate measurement of the radiation<br />

level. <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd supplies a brass hemispherical build-up cap that<br />

covers all Photon energies as well as 15 and 18 MeV Electrons. A 1.5 cm water<br />

equivalent build-up cap is available for the lower Electron energies, should build-up be<br />

a requirement. (Go to www.Team<strong>Best</strong>.com for more details).<br />

There are situations where it is intended to verify the Dmax (maximum level on the<br />

dose-depth curve) at the prescription point on a patient. Full build-up would, in<br />

principle, be used on the dosimeter. However, it is not always recommended that full<br />

build-up be used on dosimeters placed on a patient. The reason for this is that the<br />

additional build-up material would perturb the depth-dose on the patient. Physicists<br />

using the MOSFET <strong>system</strong> for patient QA <strong>dosimetry</strong> frequently use partial build-up for<br />

these circumstances. For example, 1 cm of bolus is used at 6 MV, 2 cm at all higher<br />

photon energies (see the list of publications available from <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd).<br />

In these situations of partial build-up, correction factors should be derived to<br />

determine the relationship between the dose at partial build-up and at the Dmax.<br />

7.4 CALIBRATION FOR DIAGNOSTIC X-RAY ENERGIES<br />

Calibrations should be carried out as described above for Radiotherapy EXCEPT that<br />

the black bulb side of the dosimeters should face the beam for low energy<br />

radiography doses (


OPERATOR'S MANUAL FOR THE SYSTEM<br />

7.5 TESTING A DOSIMETER FOR FADE (CHANGE IN PERFORMANCE)<br />

The dosimeters will generally maintain their calibration for 12 months. However, due<br />

to lack of use, the fade component may increase to >3%. To verify if your dosimeters<br />

are within the specification, perform the following procedure:<br />

A. Connect the dosimeter(s) to a fully charged mobileMOSFET Reader and allow it to<br />

remain connected for at least one (1) hour before performing the test. Connect the<br />

<strong>system</strong> as described in Section 3.0, SYSTEM SET-UP. Turn the Reader “ON” and allow<br />

the <strong>system</strong> to warm-up for fifteen (15) minutes.<br />

B. Press the START switch to initialize the dosimeters, then irradiate the dosimeters as<br />

described in Section 7.0 CALIBRATION PROCEDURE, to a dose of approximately<br />

200mV. START the dosimeter within one minute of irradiation and record the result from<br />

the PC (Reader’s messages) for the sensor in question (X0).<br />

C. Wait fifteen (15) minutes then press “START” and record the dosimeter result from the<br />

PC (Reader’s messages) for the sensor in question (X15) again. Compare the initial<br />

reading of each dosimeter (X0) to its final reading (X15), and use the following<br />

calculation to determine the current fade of the dosimeter:<br />

| X15 mV – X0 mV| * 100%<br />

|F| = D mV<br />

The operator can make a determination of use for any dosimeter showing fade<br />

greater than 3%. If there is any doubt, please contact the manufacturer or the<br />

representative in your area.<br />

Page 58 of 62


8 MAINTENANCE<br />

OPERATOR'S MANUAL FOR THE SYSTEM<br />

8.1 mobileMOSFET BATTERY MAINTENANCE AND REPLACEMENT<br />

The mobileMOSFET uses long life rechargeable NiMH batteries. NiMH should retain<br />

their characteristics for a period of 5 years (battery performance may degrade after<br />

this time depending on use). If the battery performance degrades, it is recommended<br />

that they be replaced.<br />

Contact <strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd to order a replacement battery pack for the Reader,<br />

and follow this procedure for battery replacement:<br />

A. Ensure the unit is powered off and the power adaptor is not connected to the<br />

mobileMOSFET.<br />

B. Remove the four screws from the bottom of the Reader, and remove the Reader’s case<br />

to access the battery pack.<br />

C. Note the battery pack's orientation in the Reader, as well as the placement of the cables.<br />

Figure 8-1: Battery Pack Placement<br />

D. Disconnect the battery power connector and remove the old battery pack (note that it is<br />

attached with Velcro, so there will be some resistance).<br />

E. Connect the new battery pack, and reassemble the unit.<br />

F. Prior to first use, switch the power switch all the way ON (momentary position). Ensure<br />

that the green LED is flashing (indicates that the power is on).<br />

G. Switch the power OFF, then connect the external power supply. The red LED indicates<br />

that the unit is in Fast Charge (reference. Section 3.2). When the red LED goes off and<br />

the green LED illuminates (Trickle Charge), the unit is fully charge (maximum of three<br />

hours).<br />

Page 59 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

8.2 ORDERING REPLACEMENT PARTS<br />

<strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd maintains a complete inventory of all normal replaceable<br />

parts.<br />

To place an order or obtain information concerning parts, contact your <strong>Best</strong> <strong>Medical</strong><br />

<strong>Canada</strong> Ltd representative or the manufacturer directly.<br />

When ordering parts, include the following information:<br />

1. Instrument Model Number<br />

2. Instrument Serial Number<br />

3. Part Description<br />

4. Part Number<br />

Page 60 of 62


OPERATOR'S MANUAL FOR THE SYSTEM<br />

8.3 TECHNICAL SPECIFICATIONS<br />

Technical specifications for the mobileMOSFET System using the TN-502RD series<br />

dosimeter. (Higher sensitivity dosimeters (TN-1002RD series) as well as Linear<br />

Array dosimeters (TN-LA5 series) are available on request.).)<br />

Accuracy<br />

Accuracy relative to an absolute source<br />

is determined during calibration by the<br />

user. Dosimeter accuracy over its dose<br />

life is approximately four percent.<br />

Recalibration every 7,000 millivolts, or<br />

equivalent accumulated exposure, will<br />

maintain this accuracy.<br />

Sensitivity for Standard<br />

TN-502RD Dosimeters (typical values)<br />

Under full buildup: (energies ≥ 1MeV)<br />

1 mV/cGy for Standard Bias Setting.<br />

2.7 mV/cGy for High Sensitivity Bias<br />

Setting.<br />

Diagnostic X-ray energies:<br />

9 mV/R for High Sensitivity Bias Setting.<br />

Resolution and Capacity<br />

The System records MOSFET voltage up<br />

to 20,000 mV with 0.01 mV resolution.<br />

The display is autoranging to keep<br />

appropriate precision.<br />

Dose Range<br />

20,000 cGy for Standard Bias Setting.<br />

7,000 cGy for High Sensitivity Bias<br />

Setting.<br />

Page 61 of 62<br />

Reader Long Term Stability<br />

Reader stability error is less than one<br />

percent of maximum accumulated dose,<br />

±1 mV, over one year.<br />

Reader Linearity<br />

Linearity error is less than 1% of the total<br />

dose reading, ±1 mV, over the full dose<br />

range.<br />

System Dose to Dose Reproducibility<br />

(at 68% confidence using 60 Co).<br />

DOSE<br />

Angular Dependence<br />

The error resulting from irradiating the<br />

dosimeter under full build-up with beam<br />

energies of 1 MeV to 20 MeV at any<br />

angle is ±2% through 360 degrees of<br />

rotation.<br />

Dosimeter Shelf Life<br />

BIAS SUPPLY<br />

SENSITIVITY<br />

HIGH STANDAR<br />

D<br />

200 cGy


OPERATOR'S MANUAL FOR THE SYSTEM<br />

Fade<br />

20 hours with readings<br />

every 5 minutes.<br />

Battery Re-charge Time:

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!