OPERATOR'S MANUAL FOR THE SYSTEM<br />

3.6.3 LAN/Wireless Connection<br />

<strong>Best</strong> <strong>Medical</strong> <strong>Canada</strong> Ltd also provides an optional solution for establishing <strong>wireless</strong><br />

connections in additional treatment rooms via the LAN and Bluetooth Wireless<br />

Transceiver Package, TN-RD-38/40. This accessory allows the mobileMOSFET<br />

Reader Module to be shared amongst multiple rooms by connecting to a PC via a<br />

LAN (Local Area Network) connection.<br />

Figure 3-6: LAN Wireless Connection<br />

The Bluetooth Wireless transceiver supplied is a serial device, however a LAN<br />

connection can be implemented if a Serial Server is connected to the LAN port.<br />

Virtual communications port (COM Port) software (supplied with the TN-RD-38/40)<br />

allows the PC to connect to this COM Port anywhere on the LAN. <strong>Best</strong> <strong>Medical</strong><br />

<strong>Canada</strong> Ltd has tested and supplies a Vlinx Model ESP901 Ethernet Serial Server.<br />

(Please note that installation may require a LAN administrator to configure the end<br />

user's PC software IP settings.) Note: With virtual COM Port software, some users<br />

experience a delay while the PC searches for COM Ports on the LAN.<br />

Mount the BMC-configured Bluetooth <strong>wireless</strong> transceiver on an interior wall of the<br />

treatment room. Attach the power supply, then connect the transceiver to the Serial<br />

Server. Next, connect the Serial Server to the LAN and attach the LAN/Wireless<br />

Transceiver's power adapter (See Figure 3-6: LAN Wireless Connection).<br />

The distance between the transceiver and mobileMOSFET as well as the placement<br />

of the transceiver and mobileMOSFET must be as specified in Section 3.6.1.<br />

Figure 3-7: Multiple Systems and Treatment Rooms and Figure 3-8: Single<br />

System/Multiple Treatment Rooms provide examples of possible configurations:<br />

Page 22 of 62