OPERATOR'S MANUAL FOR THE SYSTEM<br />

Page 44 of 62<br />

NOTE<br />

MOSFETS AND A MOSFET LINEAR ARRAY CANNOT BE READ<br />

SIMULTANEOUSLY. IF YOU WISH TO READ THE MOSFET LINEAR<br />

ARRAY, ENSURE NO SINGLE MOSFETS ARE CONNECTED TO THE<br />

READER. CONVERSELY, IF YOU WISH TO READ SINGLE MOSFETS,<br />

ENSURE THE MOSFET LINEAR ARRAY IS NOT CONNECTED TO THE<br />

READER.<br />

5.2 RECOMMENDED DOSIMETER AND BIAS SENSITIVITY COMBINATIONS FOR<br />

APPLICATIONS IN RADIOTHERAPY AND RADIOLOGY<br />

In order to determine which components best meet your needs, please review the<br />

following chart:<br />

RADIOTHERAPY APPLICATION<br />

TYPICAL<br />

APPLICATION<br />

Scatter Dose/<br />

Research<br />

RADIOTHERAPY<br />

DOSE<br />

1 to 20 cGy<br />

RECOMMENDED<br />

DOSIMETER<br />

High Sensitivity<br />

Dosimeter<br />

RECOMMENDED<br />

BIAS SENSITIVITY<br />

High Setting<br />

Central/ Scatter<br />

Dose 20 to 100 cGy Standard Dosimeter High Setting<br />

TBI/ Central Dose >100 cGy Standard Dosimeter Standard Setting<br />

TYPICAL<br />

APPLICATION<br />

RADIOLOGY APPLICATION<br />

RADIOLOGY SKIN<br />

ENTRANCE DOSE<br />

Radiology 0.1 to 5 R<br />

Normal<br />

Fluoroscopy<br />

High Dose<br />

Fluoroscopy<br />

RECOMMENDED<br />

DOSIMETER<br />

High Sensitivity<br />

Dosimeter<br />

RECOMMENDED<br />

BIAS SENSITIVITY<br />

High Setting<br />

5 to 20 R Standard Dosimeter High Setting<br />

20 to > 100 R Standard Dosimeter Standard Setting