OPERATOR'S MANUAL FOR THE SYSTEM<br />
Page 44 of 62<br />
NOTE<br />
MOSFETS AND A MOSFET LINEAR ARRAY CANNOT BE READ<br />
SIMULTANEOUSLY. IF YOU WISH TO READ THE MOSFET LINEAR<br />
ARRAY, ENSURE NO SINGLE MOSFETS ARE CONNECTED TO THE<br />
READER. CONVERSELY, IF YOU WISH TO READ SINGLE MOSFETS,<br />
ENSURE THE MOSFET LINEAR ARRAY IS NOT CONNECTED TO THE<br />
READER.<br />
5.2 RECOMMENDED DOSIMETER AND BIAS SENSITIVITY COMBINATIONS FOR<br />
APPLICATIONS IN RADIOTHERAPY AND RADIOLOGY<br />
In order to determine which components best meet your needs, please review the<br />
following chart:<br />
RADIOTHERAPY APPLICATION<br />
TYPICAL<br />
APPLICATION<br />
Scatter Dose/<br />
Research<br />
RADIOTHERAPY<br />
DOSE<br />
1 to 20 cGy<br />
RECOMMENDED<br />
DOSIMETER<br />
High Sensitivity<br />
Dosimeter<br />
RECOMMENDED<br />
BIAS SENSITIVITY<br />
High Setting<br />
Central/ Scatter<br />
Dose 20 to 100 cGy Standard Dosimeter High Setting<br />
TBI/ Central Dose >100 cGy Standard Dosimeter Standard Setting<br />
TYPICAL<br />
APPLICATION<br />
RADIOLOGY APPLICATION<br />
RADIOLOGY SKIN<br />
ENTRANCE DOSE<br />
Radiology 0.1 to 5 R<br />
Normal<br />
Fluoroscopy<br />
High Dose<br />
Fluoroscopy<br />
RECOMMENDED<br />
DOSIMETER<br />
High Sensitivity<br />
Dosimeter<br />
RECOMMENDED<br />
BIAS SENSITIVITY<br />
High Setting<br />
5 to 20 R Standard Dosimeter High Setting<br />
20 to > 100 R Standard Dosimeter Standard Setting