S E C T I O N 5 • O u t o f t h e B o x - A s s e m b l y<br />
Step 5 - Preparation for Use<br />
1 Remove the fill-port plug (a) on the front of the warming unit and<br />
fill the reservoir to the maximum level with 1.4 liters of one of the<br />
following solutions:<br />
• 0.3% Hydrogen Peroxide/Sterile Water Solution: Mix 140 ml of<br />
3% hydrogen peroxide and 1,260 ml of sterile water.<br />
Note: If this option is selected, always use a 0.3% hydrogen<br />
peroxide/sterile water solution when adding solution to the<br />
reservoir.<br />
• Distilled Water<br />
Note: Change the recirculating solution every 30 days.<br />
2 Replace the fill-port plug.<br />
3 Lubricate O-Rings in #1 Block (b) and #2 Block (c). Place a small<br />
amount of silicone lubricant, provided in the supplied O-Ring Kit,<br />
on a cotton swab and apply all around the inside of each O-Ring.<br />
Step 6 - Perform Electrical Safety Tests<br />
Perform all applicable electrical safety tests as required per institutional<br />
procedure. These include but are not limited to:<br />
• Leakage current<br />
• Hypot<br />
• Ground bond test<br />
WARNING<br />
• Grounding reliability can only be achieved when MAINS power<br />
cords are connected to a properly grounded receptacle. Risk of<br />
electrical shock exists if the equipment is not connected to a<br />
properly grounded receptacle, resulting in death or serious<br />
injury to the patient or user.<br />
The Electrical Safety Check must be performed by qualified personnel<br />
authorized by the institution to perform such testing. The Safety Check<br />
must be performed and documented at least once per year, or according to<br />
institutional policy.<br />
Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer | Operator’s Manual 11