Periodic Electrical Testing<br />
Earth Leakage<br />
The Fluid Warmer must be tested in accordance with EN 60601-1. The<br />
earth leakage current test should be performed with the immersion<br />
heater circuit in the full ON condition; for this reason the leakage<br />
current test should be performed on units which have room<br />
temperature recirculating solution in the reservoir. Units not meeting<br />
this standard should be returned to <strong>Smiths</strong> <strong>Medical</strong> or an authorized<br />
representative for service.<br />
Ground Continuity<br />
The Fluid Warmer must be tested in accordance with EN 60601-1.<br />
WARNING<br />
• Grounding reliability can only be achieved when MAINS power<br />
cords are connected to a properly grounded receptacle. Risk of<br />
electrical shock exists if the equipment is not connected to a<br />
properly grounded receptacle, resulting in death or serious<br />
injury to the patient or user.<br />
36 Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer | Operator’s Manual<br />
S E C T I O N 1 1 • T e s t i n g