SECTION 4<br />
Important Safety Information<br />
This section covers information for prescribers and guidelines for safe use<br />
of the Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer (Fluid<br />
Warmer).<br />
WARNINGS<br />
Death or serious injury may occur to the patient or user if these<br />
warnings are not followed:<br />
S E C T I O N 4 • I m p o r t a n t S a f e t y I n f o r m a t i o n<br />
• Read and follow all instructions, labeling, and accompanying<br />
documents supplied with this medical device. Failure to follow<br />
instructions, including all warnings and cautions, could result in<br />
death or serious injury to the patient or user.<br />
• Do not use the Fluid Warmer in high-energy fields such as: MRI,<br />
X-RAY, portable and mobile RF communications equipment, and<br />
other such devices. The Fluid Warmer may act as a projectile in<br />
a strong magnetic field, cause image artifacts, or not function<br />
as intended.<br />
• Do not bend the heat exchanger. Bending may damage the<br />
heat exchanger allowing communication between the<br />
recirculating solution and irrigation fluid path, resulting in the<br />
delivery of inappropriate fluids.<br />
• Exposed conductor on MAINS power cord can cause an<br />
electrocution hazard. Remove device from service if MAINS<br />
power cord has exposed wires.<br />
• Activation of the Over Temperature warning signal indicates<br />
that warming has stopped and immediate operator intervention<br />
is required. Failure to clear the over temperature condition or to<br />
take the device out of service may result in death or serious<br />
injury to the patient.<br />
Continued<br />
4 Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer | Operator’s Manual