H-1100 - Smiths Medical
H-1100 - Smiths Medical
H-1100 - Smiths Medical
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Electromagnetic Environmental Recommendations<br />
Level 1 ® NORMOFLO ® Irrigation Warming Sets<br />
(Disposable Sets)<br />
Disposable Set Accessories<br />
46 Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer | Operator’s Manual<br />
S E C T I O N 1 5 • S p e c i f i c a t i o n s<br />
Recommended separation distances between portable and mobile RF communications equipment and<br />
the Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer<br />
The Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer is intended for use in an electromagnetic environment in<br />
which radiated RF disturbances are controlled. The customer or the user of the Fluid Warmer can help prevent<br />
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications<br />
equipment (transmitters) and the Fluid Warmer as recommended below, according to the maximum output power of<br />
the communications equipment.<br />
Rated maximum<br />
Separation distance according to frequency of transmitter m<br />
output power of 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz<br />
transmitter<br />
d=[3.5/V1]√P<br />
d=[3.5/E1]√P<br />
d=[7/E1]√P<br />
W<br />
0.01 0.116 0.116 0.233<br />
0.1 0.368 0.368 0.737<br />
1 1.16 1.16 2.33<br />
10 3.69 3.69 7.38<br />
100 11.66 11.66 23.33<br />
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres<br />
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output<br />
power rating of the transmitter in watts (W) according to the transmitter manufacturer.<br />
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.<br />
NOTE 1 The guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and<br />
reflection from structures, objects and people.<br />
IR-40<br />
Product Description<br />
Normothermic Irrigating Set<br />
IR-500 Normothermic Irrigating Set<br />
IR-600 Normothermic Irrigating Set<br />
IRI-600 Normothermic Irrigating Set<br />
IRI-600B Normothermic Irrigating Set<br />
IR-700 Normothermic Irrigating Set with vented spikes for use with<br />
rigid plastic containers<br />
Product Description<br />
IR-6 2-spike adapter to convert IR/IRI series disposables to a<br />
4-spike configuration<br />
IR-7 2-spike adapter to convert IR/IRI series disposables to a<br />
4-spike configuration - for use with rigid plastic containers