SECTION 10<br />

Troubleshooting<br />

Only qualified personnel authorized by the institution should perform any<br />

routine maintenance and repairs to the Level 1 ® NORMOFLO ® H-<strong>1100</strong><br />

Irrigation Fluid Warmer.<br />

The following two tables feature general troubleshooting information<br />

along with slow flow rate troubleshooting.<br />

General Troubleshooting Guide<br />

Problem Check the following<br />

No Power Check to see if unit is plugged in and power is turned ON.<br />

Be sure the unit is plugged into a working MAINS receptacle.<br />

Disposables<br />

Alarm activates<br />

Add<br />

Recirculating<br />

Solution Alarm<br />

activates<br />

Over<br />

Temperature<br />

Alarm activates<br />

Check to see that the Heat Exchanger and Disposable Set<br />

are properly installed.<br />

Fill reservoir to the maximum reservoir level with recirculating<br />

solution.<br />

Press OFF button to clear alarm. Then, press ON button to<br />

power on. If the Fluid Warmer continues to alarm,<br />

discontinue use of the medical device and remove from<br />

service. Contact <strong>Smiths</strong> <strong>Medical</strong> or an authorized<br />

representative for service.<br />

Hot Cabinet Check the air inlet on the bottom of the unit. Remove any<br />

blockage or dust to insure adequate air flow.<br />

Heat Exchanger<br />

hard to install<br />

Green Automatic<br />

Operating LED<br />

doesn't light up<br />

during setup<br />

Lubricate O-Rings in #1 and #2 block heat exchanger<br />

sockets with silicone lubricant. Silicone lubricant part<br />

# 80-04-002.<br />

Verify unit is plugged into MAINS.<br />

If still no LED illumination, discontinue use of the medical<br />

device and remove from service.<br />