Fluid Warmer Display Panel<br />

The Fluid Warmer Display Panel provides continuous information<br />

about the operation of the Fluid Warmer. A liquid crystal display (LCD)<br />

indicates recirculating solution temperature. Just below the LCD, four<br />

light-emitting diodes (LEDs) indicate operation modes for the device.<br />

For identification purposes, the diodes are shown illuminated.<br />

1 Recirculating Solution Temperature<br />

The temperature of the recirculating solution is displayed in the<br />

LCD panel. The temperature is displayed in degrees Celsius.<br />

Note: This is NOT the temperature of fluid delivered to the<br />

patient—the display reflects the temperature of the recirculating<br />

solution.<br />

2 Automatic Operation LED<br />

This green LED indicator illuminates when the power is ON and<br />

the Disposable Set has been properly installed. When lit, this<br />

indicates the Fluid Warmer is operating.<br />

3 Check Disposables LED<br />

This yellow LED indicator illuminates and an audible attention<br />

signal beeps when the Disposable Set is not properly installed. See<br />

“Interlocks” on page 17 for directions on clearing the Check<br />

Disposables alarm.<br />

4 Add Recirculating Solution LED<br />

This yellow LED indicator illuminates and an audible attention<br />

signal beeps when reservoir is low. Additional recirculating<br />

solution must be added to the reservoir. Maximum capacity for<br />

the reservoir is 1.4 liters.<br />

5 Over Temperature LED<br />

This red LED indicator illuminates and an audible warning signal<br />

beeps when the recirculating solution is over the acceptable<br />

temperature for safe use.<br />

Reservoir Level Display<br />

The Reservoir Level Display has a clear window for viewing the amount<br />

of recirculating solution present in the reservoir. Check the reservoir to<br />

ensure the solution level is near the maximum level indicator (a). If the<br />

recirculating solution level is too low, the Add Recirculating Solution<br />

LED on the Display Panel illuminates and an audible attention signal<br />

beeps.<br />

S E C T I O N 7 • C o n t r o l s a n d D i s p l a y s<br />

16 Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer | Operator’s Manual<br />

LEVEL<br />

SYSTEM 1000<br />

MAX 1<br />

MIN<br />