Fluid Warmer Display Panel<br />
The Fluid Warmer Display Panel provides continuous information<br />
about the operation of the Fluid Warmer. A liquid crystal display (LCD)<br />
indicates recirculating solution temperature. Just below the LCD, four<br />
light-emitting diodes (LEDs) indicate operation modes for the device.<br />
For identification purposes, the diodes are shown illuminated.<br />
1 Recirculating Solution Temperature<br />
The temperature of the recirculating solution is displayed in the<br />
LCD panel. The temperature is displayed in degrees Celsius.<br />
Note: This is NOT the temperature of fluid delivered to the<br />
patient—the display reflects the temperature of the recirculating<br />
solution.<br />
2 Automatic Operation LED<br />
This green LED indicator illuminates when the power is ON and<br />
the Disposable Set has been properly installed. When lit, this<br />
indicates the Fluid Warmer is operating.<br />
3 Check Disposables LED<br />
This yellow LED indicator illuminates and an audible attention<br />
signal beeps when the Disposable Set is not properly installed. See<br />
“Interlocks” on page 17 for directions on clearing the Check<br />
Disposables alarm.<br />
4 Add Recirculating Solution LED<br />
This yellow LED indicator illuminates and an audible attention<br />
signal beeps when reservoir is low. Additional recirculating<br />
solution must be added to the reservoir. Maximum capacity for<br />
the reservoir is 1.4 liters.<br />
5 Over Temperature LED<br />
This red LED indicator illuminates and an audible warning signal<br />
beeps when the recirculating solution is over the acceptable<br />
temperature for safe use.<br />
Reservoir Level Display<br />
The Reservoir Level Display has a clear window for viewing the amount<br />
of recirculating solution present in the reservoir. Check the reservoir to<br />
ensure the solution level is near the maximum level indicator (a). If the<br />
recirculating solution level is too low, the Add Recirculating Solution<br />
LED on the Display Panel illuminates and an audible attention signal<br />
beeps.<br />
S E C T I O N 7 • C o n t r o l s a n d D i s p l a y s<br />
16 Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer | Operator’s Manual<br />
LEVEL<br />
SYSTEM 1000<br />
MAX 1<br />
MIN<br />
a